BD BBL™ Phenylethyl Alcohol Agar with 5% Sheep Blood 221277Recalls & reported problems

Becton, Dickinson And Company · BD BBL™ Phenylethyl Alcohol Agar with 5% Sheep Blood, Carton of 100 plates · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BD BBL Phenylethyl Alcohol Agar sheep blood plate is a lab culture plate. Most of the 303 reports describe plates that did not block the growth of certain bacteria, which can give inaccurate results. The reports include no injuries and no deaths, and no recalls name this device. Reports fell from 299 in the prior 12 months to none in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
303
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 303 adverse event reports that list this exact device (DI 10382902212772), received between Feb 13, 2023 and Aug 6, 2025 — classified as 303 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (301 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 137 reportsAug 2025: 162 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports202420252026162
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings301
  • Device Ingredient or Reagent Problem1
  • Device Markings/Labelling Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions303

Recent adverse event reports

The 25 most recent of 303 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22724834 ↗
  • MalfunctionAug 5, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22712971 ↗
  • MalfunctionAug 5, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22712970 ↗
  • MalfunctionAug 5, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22712195 ↗
  • MalfunctionAug 5, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22712193 ↗
  • MalfunctionAug 5, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22712190 ↗
  • MalfunctionAug 5, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22712188 ↗
  • MalfunctionAug 5, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22712187 ↗
  • MalfunctionAug 5, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22712186 ↗
  • MalfunctionAug 5, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22712033 ↗
  • MalfunctionAug 5, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22712030 ↗
  • MalfunctionAug 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22698302 ↗
  • MalfunctionAug 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22698301 ↗
  • MalfunctionAug 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22698194 ↗
  • MalfunctionAug 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22698192 ↗
  • MalfunctionAug 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22698191 ↗
  • MalfunctionAug 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22698190 ↗
  • MalfunctionAug 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22698189 ↗
  • MalfunctionAug 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22698188 ↗
  • MalfunctionAug 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22698187 ↗
  • MalfunctionAug 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22698186 ↗
  • MalfunctionAug 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22698185 ↗
  • MalfunctionAug 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22698184 ↗
  • MalfunctionAug 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22698183 ↗
  • MalfunctionAug 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ PHENYLETHYL ALCOHOL AGAR WITH 5% SHEEP BLOOD, A PLATE DID NOT SUPPRESS GROWTH OF GRAM NEGATIVES. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22698182 ↗

Other CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL devices

All CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL devices →