BD BBL™ Inhibitory Mold Agar Deep Fill 298191Recalls & reported problems
Becton, Dickinson And Company · BD BBL™ Inhibitory Mold Agar Deep Fill · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
These BD BBL Inhibitory Mold Agar Deep Fill plates are culture media. Reports mostly describe visible fungal growth (Penicillium) contaminating plates and preventing usable patient results, with no health impact mentioned. There are 40 reports, no deaths, no injuries, and no recalls naming this device. Reports rose from 1 to 39 in the past 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 40
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 10382902981913), received between Jan 3, 2025 and Feb 4, 2026 — classified as 40 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (39 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings39
- Device Markings/Labelling Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions40
Recent adverse event reports
The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 12 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257733 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 22 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257732 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 25 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257731 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 29 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257730 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 17 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257729 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 11 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257728 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 10 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257726 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 3 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257725 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 4 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257724 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 8 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257723 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 2 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257722 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 31 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257572 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 20 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257570 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 30 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257569 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 33 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257567 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 14 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257566 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 21 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257565 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 32 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257564 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 34 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257563 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 24 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257562 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 23 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257559 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 26 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257558 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 16 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257557 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 15 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257556 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED AFTER USING ONE (1) BD BBL¿ INHIBITORY MOLD AGAR DEEP FILL PLATE, THERE WAS VISIBLE FUNGAL GROWTH ON THE PLATE. AS THE FUNGAL CONTAMINATION WAS IMMEDIATELY NOTICED AFTER INCUBATION, NO PATIENT RESULTS WERE REPORTED. THE FUNGAL GROWTH WAS LATER IDENTIFIED AS PENICILLIUM SPECIES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 18 OF 34.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257555 ↗
Other CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL devices
- BD BBL™ Phenylethyl Alcohol Agar with 5% Sheep Blood 221277Becton, Dickinson And Company · CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL303 reports
- REMEL R01272Remel, Inc. · Culture Media, Selective And Non-Differential1 recall0 reports
- REMEL R01144Remel, Inc. · Culture Media, Selective And Non-Differential1 recall0 reports
- BD BBL™ Columbia CNA Agar with 5% Sheep Blood 221353Becton, Dickinson And Company · CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL9 reports
- BD BBL™ XLD Agar 221284Becton, Dickinson And Company · CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL9 reports
- BD BBL™ Lowenstein-Jensen Medium, Gruft, Slants 297653Becton, Dickinson And Company · CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL5 reports
- BD BBL™ Middlebrook and Cohn 7H10 Agar Slants 220958Becton, Dickinson And Company · CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL2 reports
- BD BBL™ Columbia CNA Agar with 5% SB and BD BBL MacConkey II Agar BD I Plate™ 221601Becton, Dickinson And Company · CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page