PINNACLE GRIPTION 1217-17-068Recalls & reported problems

Depuy Orthopaedics, Inc. · PINNACLE GRIPTION ACETABULAR SHELL REVISION (DPX) 68mm OD 62 LINER · Class 2 · Implant

All models of PINNACLE GRIPTION (57)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10603295009528), received between Aug 2, 2018 and Jun 10, 2025 — classified as 5 injuries.

The most frequently reported specific device problem is “Loose or Intermittent Connection” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loose or Intermittent Connection2
  • Adverse Event Without Identified Device or Use Problem1
  • Fracture1
  • Loss of Osseointegration1
  • Migration1
  • Osseointegration Problem1
  • Use of Device Problem1

Patient problems

  • Inadequate Osseointegration4
  • Joint Laxity1
  • Limited Mobility Of The Implanted Joint1
  • No Code Available1
  • Not Applicable1
  • Pain1
  • Tissue Damage1
  • Unspecified Infection1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS PATIENT HAS A LEFT TOTAL HIP THAT IS INFECTED. ALL COMPONENTS WERE REMOVED. PATIENT HAS MULTIPLE SURGERY DATES, (B)(6) 2020. (B)(6) 2022. (B)(6) 2025. DOI: UNKNOWN. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 22175987 ↗
  • InjuryDec 1, 2020Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE INSTABILITY AND LOOSED CUP AT THE BONE TO IMPLANT INTERFACE. DON¿T KNOW WHEN OR WHERE WAS IMPLANTED PRIOR, DON¿T HAVE ACCESS TO LOT #S, AND ORIGINAL IMPLANT DATE UNKNOWN. NO SURGICAL DELAY. DOI: UNKNOWN; DOR: (B)(6) 2020; RIGHT HIP.

    Patient problems: Inadequate Osseointegration; Joint Laxity

    MAUDE report 10927788 ↗
  • InjuryMar 20, 2019Fracture; Use of Device Problem; Loss of Osseointegration; Fracture; Use of Device Problem; Loss of Osseointegration

    PFS ALLEGES PAIN, INSTABILITY, AND RECURRENT ABSCESSES, AND USING CANE FOR MOBILITY. AFTER REVIEW OF MEDICAL RECORDS, PATIENT WAS REVISED TO ADDRESS ASEPTIC LOOSENING. GROSS PATHOLOGY REPORTED OF A LARGE VOLUME SYNOVIAL FLUID WITH CONSISTENT SYNOVITIS, METALLOSIS, METALLIC DEBRIS, BROKEN RIM FIT SCREW ONLY ONE OF FOUR SCREW , GROSS LOOSENING OF CUP. OPERATIVE REPORT CUP HAD A VERTICAL ALTITUDE WITH GROSS ANTEVERSION AND HAD MIGRATED. LOOSENING CONFIRMED. ONE BROKEN SCREW WERE READILY IDENTIFIED AS BROKEN INVOLVING THE LIPPED PORTION OF THE CUP AND DEMONSTRATED WITH GROSS METALLOSIS AND GRAY STAINED. DOI: (B)(6) 2012; DOR: (B)(6) 2014; (RIGHT HIP).

    Patient problems: Inadequate Osseointegration; Limited Mobility Of The Implanted Joint; No Code Available; Pain; Tissue Damage

    MAUDE report 8435669 ↗
  • InjuryAug 6, 2018Loose or Intermittent Connection; Migration; Loose or Intermittent Connection; Migration

    PATIENT WAS REVISED TO ADDRESS SUPERIOR MIGRATION OF THE CUP. IT WAS ALSO INDICATED THAT THE CUP WAS LOOSE AT BONE TO IMPLANT INTERFACE. THE CONSTRUCT WAS REPLACED WITH A TF AUGMENT AND MULTIHOLE GRIPTION CUP. DOI: 2017; DOR: (B)(6) 2018; LEFT HIP.

    Patient problems: Inadequate Osseointegration; Not Applicable

    MAUDE report 7751671 ↗
  • InjuryAug 2, 2018Loose or Intermittent Connection; Loose or Intermittent Connection

    PINNACLE PPF AND STICKER SHEETS RECEIVED. PPF ALLEGES LOOSENING OF CUP AND COMPONENT FRACTURE . DOI: (B)(6) 2008; DOR: NOT REPORTED (RIGHT HIP).

    Patient problems: Inadequate Osseointegration

    MAUDE report 7742177 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →