Skip to content

PINNACLE GRIPTION 1217-30-062Recalls & reported problems

Depuy Orthopaedics, Inc. · PINNACLE GRIPTION ACETABULAR SHELL MULTI-HOLE 62mm OD · Class 2 · Implant

All models of PINNACLE GRIPTION (57)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10603295010036), received between Oct 5, 2018 and Jan 3, 2024 — classified as 9 injuries.

The most frequently reported specific device problem is “Loss of Osseointegration” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • Loss of Osseointegration3
  • Osseointegration Problem2
  • No Apparent Adverse Event1
  • Use of Device Problem1

Patient problems

  • No Code Available4
  • Inadequate Osseointegration3
  • Pain3
  • Unspecified Infection2
  • Weakness2
  • Abscess1
  • Ambulation or Postural Difficulties1
  • Anxiety1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT WAS INFECTED. WASH OUT WAS DONE. DOI: UNKNOWN. DOR: (B)(3) 2023. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 18435118 ↗
  • InjuryMar 15, 2022Use of Device Problem; Use of Device Problem

    MEDICAL RECORD RECEIVED. PATIENT WAS REVISED TO ADDRESS FEMORAL COMPONENT MALPOSITION AND MULTIPLE DISLOCATIONS. THERE WAS GROSSLY MALPOSITIONED ACETABULAR COMPONENT THAT HAS BECOME WELL FIXATED TO BONE WITH SUPERIOR ACETABULAR WALL LOSS AND POSTEROLATERAL WALL INSUFFICIENCY. DOI: (B)(6) 2010 - DOR: (B)(6) 2010 ( LEFT HIP).

    Patient problems: Unspecified Tissue Injury

    MAUDE report 13770011 ↗
  • InjuryDec 10, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED TO ADDRESS PAIN. DOI: UNKNOWN. DOR: (B)(6) 2021. LEFT HIP.

    Patient problems: Pain

    MAUDE report 12977643 ↗
  • InjuryApr 28, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED. IT WAS INDICATED THAT THE PATIENT HAS HAD INFECTION EVENTS AFTER REVISION ( (B)(6) 2016, (B)(6) 2016) WHICH WAS TREATED WITH IV ANTIBIOTICS. NO REVISION SURGERY WAS REPORTED. DOI: (B)(6) 2009 (STEM) DOI: (B)(6) 2016 DOR: NONE REPORTED LEFT HIP.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 11738766 ↗
  • InjuryJul 15, 2020Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event

    AFTER REVIEW OF MEDICAL RECORDS, VISIT NOTES REPORTED THAT PATIENT HAD DIFFICULTY IN WALKING.

    Patient problems: Abscess; Ambulation or Postural Difficulties; Anxiety; Calcium Deposits/Calcification; Discomfort; Edema; Fibrosis; Inflammation; No Code Available; Pain; Unspecified Infection; Weakness

    MAUDE report 10276114 ↗
  • InjuryNov 14, 2019Loss of Osseointegration; Osseointegration Problem; Loss of Osseointegration; Osseointegration Problem

    PATIENT WAS REVISED TO ADDRESS LOOSENING OF THE CUP AT BONE TO IMPLANT INTERFACE. ORIGINAL IMPLANT DATE UNKNOWN. DOI: UNKNOWN. DOR: (B)(6) 2019. LEFT HIP.

    Patient problems: Inadequate Osseointegration; No Code Available

    MAUDE report 9323519 ↗
  • InjuryMay 10, 2019Loss of Osseointegration; Osseointegration Problem; Loss of Osseointegration; Osseointegration Problem

    LITIGATION ALLEGES THAT THE PATIENT SUFFERS FROM PAIN, DISCOMFORT, INFLAMMATION, AND LARGE AMOUNTS OF TOXIC COBALT CHROMIUM METAL ION PARTICLES TO BE RELEASED INTO THE BLOOD, TISSUE, AND BONE. UPDATE REC¿D 1/28/2014 ¿ PFS AND MEDICAL RECORDS RECEIVED. PART/LOT WAS PROVIDED. ORIGINAL LITIGATION MENTIONED ISSUES FROM METAL ON METAL; HOWEVER, THE STICKER SHEET INDICATES THAT THE PATIENT WAS ACTUALLY IMPLANTED WITH POLY. UPDATE AD 25 JUN 2018: (B)(4) HAS BEEN RE-OPENED UNDER (B)(4) DUE TO THE RECEIPT OF PPF AND MEDICAL RECORDS. IN ADDITION TO WHAT WERE PREVIOUSLY ALLEGED, PPF ALLEGES CUP LOOSENING.

    Patient problems: Inadequate Osseointegration; No Code Available

    MAUDE report 8599274 ↗
  • InjuryApr 24, 2019Loss of Osseointegration; Loss of Osseointegration

    CLAIM AND MEDICAL RECORDS RECEIVED. AFTER REVIEW OF CLAIM AND MEDICAL RECORDS, IT WAS STATED THAT THE PATIENT WAS REVISED TO ADDRESS PAIN AND ASEPTIC LOOSENING OF CUP COMPONENT. STIFFNESS, LIMPING, CLICKING, POPPING, WEAKNESS AND NUMBNESS WERE ALSO NOTED. OPERATIVE NOTES REPORTED SLIGHTLY BROWN TENTED AND TRANSLUCENT FLUID, THE TISSUE IS STAINED DARK GRAY CONSISTENT OF METALLOSIS, TWO SUPERIOR CYSTS WERE CURETTED AND THERE WAS NO ACETABULAR BONY INGROWTH NOTED. WHEN REMOVING THE ACETABULAR SHELL, THREE SCREWS BROKE BUT ONLY ONE WAS FULLY EXTRACTED AND THE TWO SCREWS WERE RECESSED SINCE THEY WERE EMBEDDED INTO THE SCLEROTIC BONE. MRI REPORTED TROCHANTERIC BURSITIS. X-RAY REPORTED ASEPTIC LOOSENING OF ACETABULAR COMPONENT AND HAS METAL WEAR DEBRIS REACTION. DOI: (B)(6) 2010; DOR: (B)(6) 2018; LEFT HIP.

    Patient problems: Cyst(s); Hypoesthesia; Inadequate Osseointegration; Limited Mobility Of The Implanted Joint; No Code Available; Pain; Tissue Damage; Weakness

    MAUDE report 8544302 ↗
  • InjuryOct 5, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INTRA-OPERATIVE NON-DISPLACED FRACTURE OF POSTERIOR WALL OF ACETABULUM DURING 2ND STAGE OF REVISION NO COMPONENTS REMOVED AT THIS TIME. DOI: (B)(6) 2018 (LEFT HIP).

    Patient problems: Bone Fracture(s)

    MAUDE report 7937830 ↗

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →

Sources and updates