PINNACLE GRIPTION 1217-32-050Recalls & reported problems
Depuy Orthopaedics, Inc. · PINNACLE GRIPTION ACETABULAR SHELL SECTOR 50mm OD · Class 2 · Implant
Quick summary
This DePuy Pinnacle Gription hip socket is an implant used in hip replacement surgery. The 182 reports mention the cup or plastic liner moving out of place, with patients reporting pain and instability that led to repeat surgery. The reports include 172 injuries, 10 malfunctions and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 184
- Deaths reported
- 0
- Injuries reported
- 174
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 184 adverse event reports that list this exact device (DI 10603295010296), received between Jul 14, 2017 and Sep 15, 2026 — classified as 174 injuries, 10 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (35 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem59
- Device Dislodged or Dislocated35
- Loss of Osseointegration25
- Osseointegration Problem17
- Use of Device Problem15
- Malposition of Device14
- Insufficient Device Problem Information11
- Noise, Audible6
Patient problems
- Pain54
- No Code Available49
- Inadequate Osseointegration39
- Unspecified Infection32
- Insufficient Information21
- Foreign Body Reaction17
- Discomfort16
- Joint Laxity14
Recent adverse event reports
The 25 most recent of 184 reports, in the reporter's own words as published by the FDA.
InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem
It was reported that patient was unstableDOI: (b)(6) 2021.DOR: (b)(6) 2026.Affected side: Unknown.
Patient problems: Joint Laxity
MAUDE report 50770964 ↗InjuryJul 28, 2026Device Dislodged or Dislocated
It was reported that the patient underwent hip arthroplasty surgery in 2022. Components implanted were Corail femoral stem, 32mm +1 Articuleze metal head, 32/50 +4 10 degree Pinnacle Marathon acetabular liner, and 50mm Pinnacle Sector Gription shell. Surgery was completed successfully at the time. In (b)(6) 2026, patient notified the surgeon that they ¿had felt a pop in their hip region and hip felt odd¿. X-rays showed that revision surgery was necessary. During the hip revision surgery, the surgeon discovered that the acetabular liner was partially dissociated. The liner was partially dissociated in the posterior aspect. The surgeon then removed the partially dissociated liner and the previous femoral head. The liner was replaced with an ALtrex polyethylene one of identical dimensions. Surgeon also replaced the femoral head with a Biolox Delta Ceramic head of identical dimensions. The hip was then reduced and surgery finished successfully according to plan. No additional information is available.
Patient problems: Insufficient Information
MAUDE report 50059959 ↗InjuryJun 22, 2026Device Dislodged or Dislocated
IT WAS REPORTED THAT LINER WAS REVISED DUE TO SUSPECTED DISASSOCIATION. LINER WAS DISCOVERED TO BE BROKEN INTO 3 PIECES. ALL PIECES WERE RETRIEVED. FEMORAL HEAD WAS ALSO REPLACED. THERE'S NO ALLEGATION FOR THE FEMORAL HEAD, IT WAS EXCHANGED BECAUSE IT WAS REMOVED TO ACCESS THE LINER. THE PATIENT WAS EXPERIENCING PAIN AND "CLUNKING" NOISE WHEN WALKING AND LOSS OF STABILITY. DOI: (B)(6) 2024. DOR: (B)(6) 2026. AFFECTED SIDE: RIGHT HIP.
Patient problems: Joint Laxity; Pain
MAUDE report 25578364 ↗InjuryFeb 5, 2026Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO INSTABILITY. DOI: UNKNOWN. DOR: (B)(6) 2026. AFFECTED SIDE: UNKNOWN HIP.
Patient problems: Joint Laxity
MAUDE report 24266069 ↗InjuryJan 23, 2026Use of Device Problem
PATIENT WAS REVISED DUE TO CUP POSITION. DOI: UNKNOWN. DOR: (B)(6) 2026. AFFECTED SIDE: RIGHT HIP.
Patient problems: Insufficient Information
MAUDE report 24165179 ↗InjuryJan 21, 2026Device Dislodged or Dislocated
IT WAS REPORTED THAT PINN SECTOR W/GRIPTION 50MM, ALTRX NEUT 32IDX50OD, AND DLT TS CER HD 12/14 32MM +5 WERE INVOLVED IN A REPORTED EVENT. THE EVENT INVOLVED DISSOCIATION OF THE POLY LINER FROM THE ACETABULAR SHELL, WHICH MAY HAVE BEEN INFLUENCED BY STRESSES ON THE CUP-LINER INTERFACE FOLLOWING A SPINAL FUSION PROCEDURE. THE SHELL WAS REMOVED DUE TO DAMAGE FROM THE FEMORAL HEAD ARTICULATING AGAINST THE INNER SHELL, RESULTING IN DAMAGE TO THE LOCKING MECHANISM, AND THE SYSTEM WAS CONVERTED TO DUAL MOBILITY. THE AFFECTED IMPLANTS WERE EXPLANTED.
Patient problems: Insufficient Information
MAUDE report 24133944 ↗InjuryJan 14, 2026Migration
IT WAS REPORTED THAT THE PATIENT DISLOCATED. CUP WAS REMOVED AND REVISED TO NEW CUP WITH DUAL MOBILITY. PER SURGEON THE CUP POSITION WAS OVERLY RETROVERTED. PATIENT WAS DETERMINED TO BE STABLE, AND THERE WAS NO PATIENT HARM. THERE WAS NO DELAY. DOI: (B)(6) 2023, DOR: (B)(6) 2025, AFFECTED SIDE: LEFT HIP.
Patient problems: Joint Laxity
MAUDE report 24073620 ↗InjuryDec 12, 2025Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED THAT THE PATIENT DISLOCATED A CONSTRAINED LINER. THERE WAS LOOSENING OF THE CUP AT THE NON-CEMENTED INTERFACE. THERE WERE NO DELAY, OR NO HARM TO PATIENT. NO ADDITIONAL INFORMATION IS AVAILABLE FOR REPORTING. DOI: (B)(6) 2025, DOR: (B)(6) 2025, AFFECTED SIDE: RIGHT HIP.
Patient problems: Inadequate Osseointegration
MAUDE report 23795198 ↗InjuryDec 10, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT PATIENT WAS REVISED FOR UNSTABLE CUP. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.
Patient problems: Joint Laxity
MAUDE report 23770307 ↗InjuryNov 17, 2025Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED THAT THE CUP HAD MOVED. SURGEON TOOK OUT CUP, LINER, SCREW, HIP BALL AND PATIENT WAS IMPLANTED WITH A LARGER CUP NEW POLY AND LONGER HIP BALL. DOI (B)(6) 2022, DOR (B)(6) 2025, AFFECTED SIDE- LEFT HIP.
Patient problems: Inadequate Osseointegration
MAUDE report 23575962 ↗InjuryNov 4, 2025Naturally Worn; Naturally Worn
(B)(6) SCHOOL OF ENGINEERING AT (B)(6) FIRST REPORT OF RETRIEVALS FOR STATEMENT OF WORK 10 FOR Q3 OF 2025. (B)(6) IS THE RESEARCH LOCATION, AND NOT THE FACILITY OF PLACEMENT. A DEPUY SYNTHES IMPLANT WAS REVISED AND REVIEWED FOR ANALYSIS. REASON FOR REVISION: REVISION DUE TO POLY WEAR AND PAIN. AT REVISION THE ACETABULAR CUP SHOWED AN AREA OF METAL-ON-METAL CONTACT BURNISHING, SCRATCHING, AND ABRASION. THE ACETABULAR LINER SHOWED CRACKING.
Patient problems: Pain
MAUDE report 23465771 ↗InjuryOct 27, 2025Use of Device Problem; Use of Device Problem
PATIENT WAS REVISED AS THE CUP WAS IN POOR POSITIONING. DOI: (B)(6) 2015. DOR: (B)(6) 2025 . AFFECTED SIDE: LEFT HIP.
Patient problems: Insufficient Information
MAUDE report 23390679 ↗InjuryJul 31, 2025Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED THAT THE CUP HAD FIBROUS TISSUE GROWTH ON THE BACKSIDE, AND THE CUP WAS LOOSE. TWO SCREWS CAME OUT AS WELL, LENGTH OF WHICH WERE UNKNOWN. PRIMARY SURGERY WAS DONE AT DIFFERENT HOSPITAL. DOI: UNKNOWN. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT HIP.
Patient problems: Inadequate Osseointegration
MAUDE report 22677288 ↗InjuryJul 11, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE CUP WAS IN A TOO VERTICAL POSITION. DOI: (B)(6) 2023, DOR: (B)(6) 2025, AFFECTED SIDE: RIGHT HIP.
Patient problems: Insufficient Information
MAUDE report 22481718 ↗InjuryJul 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ACETABULAR REVISION WAS PERFORMED DUE TO INSTABILITY. THERE WERE NO DELAYS OR COMPLAINTS. ITEMS WERE UNAVAILABLE FOR RETURN. DOI: (B)(6) 2024. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT HIP.
Patient problems: Joint Laxity
MAUDE report 22394642 ↗InjuryJun 13, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT ALL IMPLANTS WERE EXPLANTED EXCEPT FOR THE SROM SLEEVE DUE TO MALPOSITIONING. DOI: (B)(6) 2018. DOR:(B)(6) 2018. AFFECTED SIDE: LEFT HIP.
Patient problems: Insufficient Information
MAUDE report 22205081 ↗InjuryJun 5, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PINNACLE LINER WAS DISSOCIATED. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. PROCEDURE AND PATIENT OUTCOME WERE UNKNOWN. DOI: (B)(6) 2020. DOR: (B)(6) 2025. AFFECTED AREA: RIGHT HIP.
Patient problems: Insufficient Information
MAUDE report 22155331 ↗InjuryApr 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD AN INFECTED HIP. SURGEON CONDUCTED A 1 STAGE REVISION. SURGICAL DELAY WAS UNKNOWN. DOI: (B)(6) 2024. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.
Patient problems: Unspecified Infection
MAUDE report 21880974 ↗InjuryMar 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT WAS REVISED DUE TO INFECTION. REMOVAL OF PROSTHETHESIS INSERTION OF ANTIBIOTIC SPACER. AFFECTED SIDE: RIGHT HIP.
Patient problems: Unspecified Infection
MAUDE report 21577326 ↗InjuryJan 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO PAIN. CUP WAS REVISED. DOI: (B)(6) 2023. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.
Patient problems: Pain
MAUDE report 21252165 ↗InjuryJan 10, 2025Use of Device Problem; Device Dislodged or Dislocated; Noise, Audible; Use of Device Problem; Device Dislodged or Dislocated; Noise, Audible
LITIGATION RECORDS RECEIVED. PATIENT ALLEGES MALFUNCTIONING, FAILURE, MECHANICAL LOOSENING OF THE INTERNAL PROSTHETIC JOINT, SQUEAKING SOUNDS, HIP PAIN , CLICKING, IMBALANCE, DIFFICULTY WALKING. IT BAND CORROSION AND FOUND A POLYETHYLENE FAILURE DUE TO A FRACTURE OF THE LOCKING MECHANISM. ECCENTRIC SUPERIOR POSITION OF THE FEMORAL HEAD RELATIVE TO THE ACETABULUM IS CONSISTENT WITH POLYETHYLENE WEAR. CALCIFIED UTERINE FIBROID. CONSISTENT WITH POSTERIOR IMPINGEMENT TRUNNION INTACT WITH A FEW MINOR SCRATCHES. WAS FOUND TO HAVE A DISSOCIATED ACETABULAR LINER. FEMORAL COMPONENT SHOWED SIGNIFICANT STRIPE WEAR. BLACK FILM TO THE ANTERIOR ASPECT, POSTERIOR CAPSULE AND SOME OF THE ABDUCTOR MUSCULATURE. DOI: (B)(6), 2021. DOR: (B)(6), 2023. RIGHT HIP. ON (B)(6)2021 THE PATIENT HAD A RIGHT TOTAL HIP ARTHROPLASTY TO ADDRESS OSTEOARTHRITIS. DEPUY PRODUCTS WERE IMPLANTED DURING THIS SURGERY. (B)(6)2023 RADIOGRAPHS NOTE THE PATIENT WAS STATUS POST RIGHT TOTAL HIP ARTHROPLASTY WITH THE INDICATIONS FOR RIGHT HIP JOINT PAIN. ECCENTRIC SUPERIOR POSITION OF THE FEMORAL HEAD RELATIVE TO THE ACETABULUM IS CONSISTENT WITH POLYETHYLENE WEAR. (B)(6)2023, MEDICAL RECORDS NOTE THE PATIENT HAS CATASTROPHIC FA
Patient problems: Ambulation or Postural Difficulties; Foreign Body Reaction; Joint Laxity; Loss of Range of Motion; Pain; Swelling/ Edema
MAUDE report 21124444 ↗InjuryNov 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED, THAT THERE WAS A POST-OP INFECTION OF A HIP.
Patient problems: Unspecified Infection
MAUDE report 20666148 ↗InjuryNov 6, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
ADDITIONAL INFORMATION RECEIVED STATES THAT THERE WAS NO SURGICAL DELAY.
Patient problems: Insufficient Information
MAUDE report 20622744 ↗InjuryOct 4, 2024Migration; Migration
PATIENT WAS SUBLIMING FROM ORIGINAL HIP ARTHROPLASTY, SO SURGEON PULLED ORIGINAL CUP, LINER, AND HEAD BALL. NO KNOWN INFECTION PRESENT. NO DELAY OF SURGERY. UNKNOWN ORIGINAL SURGERY DATE. DOI: UNKNOWN DOR: (B)(6) 2024 AFFECTED SIDE : LEFT HIP.
Patient problems: Insufficient Information
MAUDE report 20369524 ↗InjuryAug 9, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
A PATIENT HAD REVISION TOTAL HIP REPLACEMENT SURGERY IN (B)(6) 2024 AS A RESULT OF A DISLOCATION AND ASSOCIATED LINER DISSOCIATION. THE COMPONENTS IMPLANTED AT THE TIME OF REVISION WERE: 28/50 NEUTRAL MARATHON LINER (121928050) AND A 28MM +1.5MM ARTICULEZE METAL FEMORAL HEAD (136511000). ON SATURDAY, JULY 27TH, PATIENT WAS GETTING UP FROM LYING ON THE SAND AT THE BEACH AND FELT SOMETHING CHANGE IN THEIR HIP. PATIENT THEN EXPERIENCED PAIN AND WENT TO THE HOSPITAL. AFTER X-RAYS WERE TAKEN IT WAS DETERMINED REVISION SURGERY NEEDED TO TAKE PLACE. PATIENT HAD SURGERY AND IT WAS DETERMINED THAT THE LINER (121928050) HAD DISSOCIATED FROM THE ACETABULAR SHELL (121732050). SURGEON REMOVED THE DISSOCIATED LINER (121928050) AND REMOVED THE FEMORAL HEAD (136511000). HIP WAS REDUCED AND CLOSURE TOOK PLACE. ALL WENT ACCORDING TO PLAN AND FINISHED SUCCESSFULLY. EXPLANTED COMPONENTS WERE RETAIN AND SENT TO LOCAL IMPLANT RETRIEVAL LAB FOR ANALYSIS, AS PER REGIONAL PROTOCOLS. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Discomfort; Pain
MAUDE report 19949421 ↗
Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.