Skip to content

S-ROM MARATHON 1220-70-128Recalls & reported problems

Depuy Orthopaedics, Inc. · S-ROM MARATHON ACETABULAR LINER 28mm 10 DEGREES M SERIES · Class 3 · Implant

All models of S-ROM MARATHON (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
12
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10603295015840), received between Feb 23, 2018 and Apr 6, 2026 — classified as 12 injuries.

The most frequently reported specific device problem is “Naturally Worn” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Naturally Worn6
  • Adverse Event Without Identified Device or Use Problem3
  • Device Dislodged or Dislocated2
  • Insufficient Device Problem Information2

Patient problems

  • Pain5
  • No Code Available4
  • Joint Dislocation2
  • Fall1
  • Insufficient Information1
  • No Consequences Or Impact To Patient1
  • Not Applicable1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 6, 2026Device Dislodged or Dislocated; Naturally Worn

    IT WAS REPORTED THAT PATIENT HAD A RIGHT TOTAL HIP DONE WITH A SROM STEM AND SROM CUP PLACED AROUND THIRTY YEARS AGO. PATIENT DEVELOPED POLY WEAR, SO PATIENT HAD A HEAD AND LINER EXCHANGE DONE ON (B)(6) 2019. WAS DOING GREAT UNTIL RECENTLY; THE PATIENT FELL AND DISLOCATED. THE SURGEON WANTED TO REMOVE THE CUP. DID A COMPLETE ACETABULAR REVISION UTILIZING DUAL MOBILITY KEEPING THE SROM HEAD IN PLACE. DOI: (B)(6) 2019. DOR: (B)(6) 2026. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Fall; Joint Dislocation

    MAUDE report 24790755 ↗
  • InjuryJul 29, 2025Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE PATIENT HAD OLD SROM CUP WITH SROM STEM. THERE WAS POLY WEAR AND THE PATIENT NEEDED HEAD AND LINER EXCHANGE. ALL PIECES WERE RETRIEVED. AFFECTED SIDE: LEFT HIP.

    Patient problems: Insufficient Information

    MAUDE report 22642105 ↗
  • InjuryDec 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO PAIN. DOI: UNKNOWN. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.

    Patient problems: Pain

    MAUDE report 15894045 ↗
  • InjuryNov 2, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION COMPLETED DUE TO PAIN. DOI: UNKNOWN, DOR: OCT 19, 2022, AFFECTED SIDE: LEFT HIP.

    Patient problems: Pain

    MAUDE report 15716633 ↗
  • InjuryApr 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO PAIN. DOI: UNKNOWN; DOR: (B)(6) 2022; AFFECTED SIDE: UNKNOWN.

    Patient problems: Pain

    MAUDE report 13977060 ↗
  • InjuryJul 8, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THIS PATIENT HAD HER HIP REPLACED 20+ YEARS AGO, THE SURGEON AND HOSPITAL WHERE THE ORIGINAL PRIMARY HIP TOOK PLACE IS UNKNOWN. DR. (B)(6) REVISED THIS PATIENT'S HIP ON (B)(6) 2021 FOR A WORN OUT ACETABULAR LINER. THE PATIENT DID FINE UNTIL RECENTLY ACCORDING TO DR. (B)(6) BUT HAS DISLOCATED SEVERAL TIMES AFTER BENDING INTO A POSITION TRYING TO CATCH AN ITEM FALLING. WHAT THE ITEM AND POSITION ARE UNKNOWN. THE SURGEON REVISED THE HIP AND EXPLANTED THE LINER AND REPLACED IT WITH A CONSTRAINED LINER AND THE SUPERCUP WAS LEFT IN VITRO. THE STEM AND HIP BALL ARE NOT A DEPUY PRODUCT. THIS IS ALL OF THE INFORMATION AVAILABLE TO ME AT THIS TIME. DOI: (B)(6) , 2021 DOR: (B)(6) 2021 AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 12137973 ↗
  • InjuryMar 5, 2020Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE PATIENT¿S HIP BECAME SLIGHTLY UNSTABLE AFTER 25+ YEARS OF STABILITY. THE LINER HAS SOME SUPERIOR WEAR SO THE SURGEON ELECTED TO RETAIN THE STEM AND CUP COMPONENTS AND REPLACE THE HEAD/LINER TO RESTORE STABILITY. ORIGINAL IMPLANT DATE UNKNOWN. DOI: 25+ YEARS. DOR: (B)(6) 2020, RIGHT SIDE.

    Patient problems: No Code Available

    MAUDE report 9793029 ↗
  • InjuryNov 26, 2019Naturally Worn; Naturally Worn

    THE PATIENT WAS REVISED TO ADDRESS POLY CHANGED DUE TO WEAR. DOI: (B)(6) 1994; DOR: (B)(6) 2019; LEFT HIP.

    Patient problems: No Code Available

    MAUDE report 9378113 ↗
  • InjurySep 19, 2019Naturally Worn; Naturally Worn

    PATIENT WAS REVISED DUE TO WORN POLY. ORIGINAL IMPLANT DATE UNKNOWN. DOI: UNKNOWN. DOR: (B)(6) 2019, RIGHT HIP.

    Patient problems: No Code Available; Pain

    MAUDE report 9087794 ↗
  • InjuryMar 7, 2019Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE PATIENT HAD A RIGHT THA PERFORMED APPROXIMATELY 20 YEARS AGO. THE SURGEON FELT FROM X-RAY THAT THE PATIENT HAD DEVELOPED POLY WEAR. AFTER THE SURGEON SENT ME A PICTURE OF THE X-RAY, IT WAS DETERMINED TO BE AN SROM STEM AND SROM CUP. DURING THE SURGERY, THE SURGEON REMOVED THE SROM LINER AND 2 LOCKING PINS. THERE WAS SIGNIFICANT POLY WEAR, BUT THE SURGEON DIDN'T THINK IT WAS EXCESSIVE CONSIDERING IT HAS BEEN IN APPROXIMATELY 20 YEARS. THE SURGEON WAS ABLE TO IDENTIFY THE POLY AS A M SERIES. HE TRIALED A 10DEG TRIAL AND LIKED THE STABILITY AND LEG LENGTH. HE IMPLANTED THE NEW 28MM 10DEG M SERIES LINER AND USED 3-5.0 LINER LOCKING PINS. HE ALSO USED A CERAMIC SROM HEAD 28MM +0. HE WAS HAPPY WITH THE OUTCOME OF THE SURGERY. DOI: (B)(6) 1999; DOR: (B)(6) 2019, RIGHT HIP.

    Patient problems: No Code Available

    MAUDE report 8400297 ↗
  • InjuryJun 19, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    REVISION HIP: UNKNOWN PRIMARY DETAILS: THIS CASE IS LOGGED TO CAPTURE 1ST REVISION: WE HAD A SURGEON REPLACE A LINER 122070128 S-ROM MARATHON LINER M 28 10 DEGREE NON-CONSTRAINED INTO AN S-ROM 51MM OBLONG SHELL.

    Patient problems: No Consequences Or Impact To Patient; Not Applicable

    MAUDE report 7613516 ↗
  • InjuryFeb 23, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    THE PATIENT WAS REVISED TO ADDRESS PAIN. IT WAS STATED THAT THE PATIENT HAD THR OVER 20 YEARS AGO. DOI: 20 YEARS AGO; DOR: (B)(6) 2018; LEFT HIP.

    Patient problems: Pain

    MAUDE report 7291825 ↗

Other Prosthesis, hip, semi-constrained (metal uncemented acetabular component) devices

All Prosthesis, hip, semi-constrained (metal uncemented acetabular component) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page