LCS COMPLETE 1294-12-040Recalls & reported problems
Depuy Orthopaedics, Inc. · LCS COMPLETE RPS Flexion Femoral CEMENTED STD LEFT · Class 3 · Implant
All models of LCS COMPLETE (117)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10603295022923), received between Oct 11, 2018 and Jun 27, 2023 — classified as 4 injuries.
The most frequently reported specific device problem is “Loss of or Failure to Bond” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Loss of or Failure to Bond2
Patient problems
- Pain3
- No Code Available2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Foreign Body Reaction1
- Synovitis1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJun 27, 2023Loss of or Failure to Bond; Loss of or Failure to Bond
MEDICAL RECORDS AD (B)(6) 2023, WERE REVIEWED BY A CLINICIAN. PATIENT UNDERWENT A LEFT TOTAL KNEE ARTHROPLASTY UTILIZING DEPUY SYNTHES IMPLANTS ON (B)(6) 2013, INCLUDING PATELLA. COMPETITOR CEMENT WAS UTILIZED. ON (B)(6) 2023, THE PATIENT UNDERWENT A LEFT TOTAL KNEE REVISION FOR LOOSENING. DARK STAINING OF THE SYNOVIUM WAS PRESENT WITH EXTENSIVE METALLOSIS. THE FEMORAL COMPONENT WAS LOOSE WITH NO INTERFACE CLEARLY PROVIDED. THE TIBIAL COMPONENT WAS ALSO LOOSE AT THE IMPLANT TO CEMENT INTERFACE. THE PATELLA WAS UNDERSIZED ACCORDING TO THE SURGEON AND WAS REVISED. DOI: (B)(6) 2013. DOR: (B)(6) 2023. LEFT KNEE.
Patient problems: Foreign Body Reaction; Synovitis
MAUDE report 17208747 ↗InjuryJun 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION RECEIVED INDICATED THAT NONE OF THE IMPLANTS WERE REVISED. THE PATIENTS PATELLA WAS RESURFACED AND THE ECTOPIC BONE REMOVED. NONE OF THE KNEE COMPONENTS LISTED REQUIRED REVISION AND WERE ALL LEFT IN SITU. AS THIS WAS A REVISION PROCEDURE THERE WAS NO EXTENSION TO THE SURGICAL TIME.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Pain
MAUDE report 12041158 ↗InjuryOct 23, 2019Loss of or Failure to Bond; Loss of or Failure to Bond
THE PATIENT WAS REVISED TO ADDRESS PAIN AND LOOSENING OF THE TIBIAL AND FEMORAL COMPONENT AT THE CEMENT TO IMPLANT INTERFACE. CEMENT MANUFACTURER WAS UNKNOWN. REVISED TO AN ATTUNE REVISION RP COMPONENT WITH SLEEVES AND STEMS AND TIBIAL WEDGE 15MM LEFT MEDIAL. DOI: (B)(6) 2009; DOR: (B)(6) 2019; LEFT KNEE.
Patient problems: No Code Available; Pain
MAUDE report 9227237 ↗InjuryOct 11, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED TO ADDRESS PAIN AND LOOSENING OF THE TIBIAL AND FEMORAL COMPONENT AT THE BONE TO CEMENT INTERFACE.
Patient problems: No Code Available; Pain
MAUDE report 7955941 ↗
Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices
- ATTUNE 1506-10-005Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing442 reports
- ATTUNE 1506-10-004Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing442 reports
- ATTUNE 1518-20-038Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing3 deaths393 reports
- LCS SIGMA 1294-33-130Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing1 recall310 reports
- ATTUNE 1518-20-035Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing342 reports
- P.F.C. SIGMA 96-0102Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing1 death319 reports
- ATTUNE 1518-20-041Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing308 reports
- ATTUNE 1506-10-007Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing284 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page