Skip to content

NA 1294-35-425Recalls & reported problems

Depuy (Ireland) · TIBIAL TRAY ROTATING PLATFORM M.B.T. REVISION THICK TRAY SIZE 4 25mm CEMENTED · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10603295026051), received between Oct 31, 2019 and Apr 23, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Loss of or Failure to Bond” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Loss of or Failure to Bond1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • No Code Available1
  • Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 23, 2024Loss of or Failure to Bond; Loss of or Failure to Bond

    A DEPUY SYNTHES IMPLANT WAS REVISED AND REVIEWED FOR ANALYSIS. REASON FOR REVISION: TIBIAL TRAY LOOSE AND PAIN. AT REVISION, THE TIBIAL INSERT SHOWED MINOR YELLOWING.

    Patient problems: Pain

    MAUDE report 19168951 ↗
  • InjurySep 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT'S KNEE WAS REVISED DUE TO INFECTION AND LOOSENING OF THE FEMORAL & TIBIAL COMPONENT AT THE BONE TO CEMENT INTERFACE. UNKNOWN CEMENT WAS USED. REMOVAL OF ALL IMPLANTS DUE TO CEMENT COULDN'T SEE MOST NUMBERS. ONLY ABLE TO OBTAIN THE NUMBERS FROM FEMORAL STEM, TC3 RP INSERT AND THE THICK TRAY. ALL OTHER PARTS WERE COVERED BY CEMENT. NO FURTHER INFORMATION IS AVAILABLE. DOI: (B)(6) 2015. DOR: (B)(6) 2021. RIGHT KNEE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12517247 ↗
  • InjuryOct 31, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2017, THE PATIENT UNDERWENT A LEFT KNEE REVISION DUE TO LOOSENING, INSTABILITY, DISLOCATED POLYETHYLENE, FOOTDROP AND EXTENSOR MECHANISM COMPROMISE. THE SURGEON NOTED THE POLYETHYLENE INSERT WAS COMPLETELY DISLOCATED. THE PATIENT WAS IMPLANTED WITH DEPUY KNEE SYSTEM AND COMPETITOR¿S CEMENT. THERE WERE NO COMPLICATIONS REPORTED WITH THE PROCEDURE. DOI: (B)(6) 2015 (PATELLAR AND FEMORAL COMPONENTS). DOI: (B)(6) 2017 (TIBIAL TRAY, INSERT AND BONE CEMENT). DOR: (B)(6) 2017 (LT KNEE).

    Patient problems: No Code Available

    MAUDE report 9260700 ↗

Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices

All Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page