DELTA XTEND 130790024Recalls & reported problems
Depuy (Ireland) · DELTA Xtend Locking Metaglene Screw Dia 4.5 Lg 24mm · Class 2 · Implant
All models of DELTA XTEND (194)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy Delta Xtend shoulder implant has 118 adverse event reports, with 104 injuries and no deaths. Many describe the shoulder base plate or screws loosening, and some describe fractures, infections or pain that led to revision surgery. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 122
- Deaths reported
- 0
- Injuries reported
- 108
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 122 adverse event reports that list this exact device (DI 10603295027836), received between Aug 31, 2017 and Sep 28, 2026 — classified as 108 injuries, 14 malfunctions.
The most frequently reported specific device problem is “Osseointegration Problem” (36 reports).
10 other recalls cover devices cleared under the same FDA submission (K062250, K081620, K120174, K192855); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem50
- Osseointegration Problem36
- Fracture13
- Loss of Osseointegration13
- Insufficient Device Problem Information3
- Loss of or Failure to Bond3
- Difficult to Insert2
- Failure to Osseointegrate2
Patient problems
- Inadequate Osseointegration48
- Unspecified Infection44
- No Code Available26
- Pain18
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available10
- No Clinical Signs, Symptoms or Conditions7
- Bone Fracture(s)6
- Joint Laxity6
Recent adverse event reports
The 25 most recent of 122 reports, in the reporter's own words as published by the FDA.
InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Unspecified Infection
MAUDE report 50923234 ↗InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Unspecified Infection
MAUDE report 50923219 ↗InjurySep 17, 2026Osseointegration Problem
Patient problems: Inadequate Osseointegration; Unspecified Infection
MAUDE report 50810916 ↗InjurySep 17, 2026Osseointegration Problem
Patient problems: Inadequate Osseointegration; Unspecified Infection
MAUDE report 50810878 ↗MalfunctionMay 21, 2026Fracture
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGERY THE SCREW BROKE AND WAS DISCARDED. THERE WERE NO PATIENT CONSEQUENCES. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25244342 ↗InjuryMay 1, 2026Osseointegration Problem
IT WAS REPORTED THAT PATIENT REVISED TO A HEMI CTA AFTER A FAILED REVERSE SHOULDER REVISION. PREVIOUSLY HAD AN ANATOMIC SHOULDER IN 2020. IT WAS STATED THAT METAGLENE BASEPLATE WAS LOOSE. LOOSENING INTERFACE WAS MARKED AS NON-CEMENTED. DOI-(B)(6) 2024 DOR- (B)(6) 2026 AFFECTED SIDE- RIGHT SHOULDER
Patient problems: Inadequate Osseointegration
MAUDE report 25056808 ↗InjuryApr 29, 2026Adverse Event Without Identified Device or Use Problem
SUBJECT ID: (B)(6). STUDY NO: (B)(4). CLINICAL ADVERSE EVENT RECEIVED FOR DISPLACED TYPE 2B SCAPULAR SPINE FRACTURE, DEVICE (RELATEDNESS): PROBABLE, PROCEDURE (RELATEDNESS): CAUSAL RELATIONSHIP, DATE OF EVENT: 23 FEB 2023, DATE OF IMPLANT: NO INFORMATION PROVIDED, DATE OF REVISION: ON (B)(6) 2025, DEVICE LOCATION: NO INFORMATION PROVIDED. TREATMENT/IMPACT: SURGICAL INTERVENTION FOR THE REGISTRY SHOULDER: YES, DATE OF SURGICAL INTERVENTION: ON (B)(6) 2026, OPEN REDUCTION INTERNAL FIXATION (ORIF): YES, COMPONENTS REMOVED (MARK NONE OR ALL THAT APPLY.) :(PLEASE COMPLETE A PATIENT COMPLETION/DISCONTINUATION FORMNBSP; NONE: YES. DEPUY SYNTHES PRODUCTS USED: TYPE: SCREW, CATALOG NUMBER: 130960315, LOT NUMBER: RL295828, DESCRIPTION: CENTRAL METAGLENE SCREW 6MM X 15MM, TYPE: SCREW, CATALOG NUMBER: 130790024, LOT NUMBER: 5547930, DESCRIPTION: DELTA XTEND LOCKING SCREW 4.5 X 24, TYPE: SCREW, CATALOG NUMBER: 130770018, LOT NUMBER: 5819790, DESCRIPTION: DELTA XTEND NON-LOCKING SCREW 4.5 X 18, TYPE: SCREW, CATALOG NUMBER: 130790024, LOT NUMBER: 5552950, DESCRIPTION: DELTA XTEND LOCKING SCREW 4.5 X 24, TYPE: SCREW, CATALOG NUMBER: 130770018, LOT NUMBER: 5829670, DESCRIPTION: DELTA XTEND NON-LOCK
Patient problems: Bone Fracture(s)
MAUDE report 25034067 ↗InjuryApr 29, 2026Adverse Event Without Identified Device or Use Problem
SUBJECT ID: (B)(6) STUDY NO: (B)(4). CLINICAL ADVERSE EVENT RECEIVED FOR DISPLACED TYPE 2B SCAPULAR SPINE FRACTURE DEVICE (RELATEDNESS): PROBABLE PROCEDURE (RELATEDNESS): CAUSAL RELATIONSHIP DATE OF EVENT: 23 FEB 2023 DATE OF IMPLANT: NO INFORMATION PROVIDED DATE OF REVISION: (B)(6) 2025 DEVICE LOCATION: NO INFORMATION PROVIDED. TREATMENT/IMPACT: SURGICAL INTERVENTION FOR THE REGISTRY SHOULDER: YES, DATE OF SURGICAL INTERVENTION: (B)(6) 2026, OPEN REDUCTION INTERNAL FIXATION (ORIF): YES, COMPONENTS REMOVED (MARK NONE OR ALL THAT APPLY.) :( PLEASE COMPLETE A PATIENT COMPLETION/DISCONTINUATION FORMNBSP; NONE: YES . DEPUY SYNTHES PRODUCTS USED: TYPE: SCREW CATALOG NUMBER: 130960315 LOT NUMBER: RL295828 DESCRIPTION: CENTRAL METAGLENE SCREW 6MM X 15MM. TYPE: SCREW CATALOG NUMBER: 130790024 LOT NUMBER: 5547930 DESCRIPTION: DELTA XTEND LOCKING SCREW 4.5 X 24. TYPE: SCREW CATALOG NUMBER: 130770018 LOT NUMBER: 5819790 DESCRIPTION: DELTA XTEND NON-LOCKING SCREW 4.5 X 18. TYPE: SCREW CATALOG NUMBER: 130790024 LOT NUMBER: 5552950 DESCRIPTION: DELTA XTEND LOCKING SCREW 4.5 X 24. TYPE: SCREW CATALOG NUMBER: 130770018 LOT NUMBER: 5829670 DESCRIPTION: DELTA XTEND NON-LOCKING SCREW 4.5 X 18. TYPE:
Patient problems: Bone Fracture(s)
MAUDE report 25033928 ↗InjuryFeb 25, 2026Osseointegration Problem
IT WAS REPORTED THAT FAILED GLENIOD DISTAL STEM LEFT IN PLACE. IT WAS STATED THAT BASE PLATE AND SCREWS WERE LOOSE. LOOSENING INTERFACE WAS MARKED AS NON-CEMENTED. DOI: (B)(6) 2024. DOR: (B)(6) 2026. AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Inadequate Osseointegration
MAUDE report 24443165 ↗InjuryDec 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THE PATIENT WAS REVISED DUE TO PAINFUL RIGHT SHOULDER. IT WAS NOTED THE GLENOSPHERE APPEARED TO BE TILTED FROM METAGLENE. THE IMPLANTS WERE REMOVED AND REPLACED. IT WAS NOTED THERE WAS A POSSIBLE INFECTION IN THE SHOULDER AND A POSSIBLE BREAK ON THE POST AT THE BOTTOM OF GLENOSPHERE AND PIECE IN THE METAGLENE POST. DOI: (B)(6) 2019. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Pain; Unspecified Infection
MAUDE report 23935683 ↗InjuryDec 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ALL IMPLANTS WERE REMOVED DUE TO INFECTION. AFFECTED SIDE- RIGHT SHOULDER.
Patient problems: Unspecified Infection
MAUDE report 23818288 ↗InjuryNov 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISION OF REVERSE LEFT SHOULDER. REASON FOR REVISION ¿ INFECTION.
Patient problems: Unspecified Infection
MAUDE report 23558496 ↗InjuryOct 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL CASE IN REGARD TO (B)(4) TO CAPTURE THE PREVIOUS PARTIAL REVISION. PARTIAL REVISION; PROCEDURE: REMOVAL OF ALL COMPONENTS EXCEPT THE STEM; RETAINED COMPONENT: STEM; NEW COMPONENTS INSERTED: EPI BODY, CTA HEAD; INDICATION: SUSPECTED INFECTION; STEM WAS WELL-FIXED AND COULD NOT BE REMOVED; CTA HEAD PLACED ONTO RETAINED STEM; PURULENT FLUID OBSERVED IN BONE AND ON IMPLANTS; SWABS TAKEN FOR MICROBIOLOGICAL TESTING; PATIENT COMMENCED ON HIGH-DOSE ANTIBIOTICS.
Patient problems: Unspecified Infection
MAUDE report 23372456 ↗InjurySep 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED TO HEMIARTHROPLASTY DUE TO INSTABILITY. DOI: (B)(6) 2022. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Joint Laxity
MAUDE report 23031116 ↗InjuryJul 1, 2025Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED THAT PATIENTS METAGLENE LOOSENED POST OP. METAGLENE, 2- 18 NON LOCKING SCREWS AND 2- 24MM LOCKING SCREWS BECAME LOOSE AT NON-CEMENTED INTERFACE. CEMENT MANUFACTURER WAS UNKNOWN. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Inadequate Osseointegration
MAUDE report 22378619 ↗InjuryJul 1, 2025Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED THAT PATIENTS METAGLENE LOOSENED POST OP. METAGLENE, 2- 18 NON-LOCKING SCREWS AND 2- 24MM LOCKING SCREWS BECAME LOOSE AT NON-CEMENTED INTERFACE. CEMENT MANUFACTURER WAS UNKNOWN. DOI:(B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Inadequate Osseointegration
MAUDE report 22378562 ↗InjuryJun 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REVERSE SHOULDER PROSTHESIS WAS IMPLANTED IN OUR HOSPITAL FOUR YEARS AGO. ITS FUNCTION IS EXCELLENT. A FIRM, ELASTIC BULGE HAS FORMED ANTERIORLY IN THE AREA OF THE DELTOIDEOPECTORAL APPROACH. RADIOLOGICAL EXAMINATION ON (B)(6) 2024, REVEALED A DISLOCATION OF THE GLENOSPHERE RELATIVE TO THE METAGLENE. THERE IS AN INDICATION FOR REPLACEMENT OF THE GLENOSPHERE AND, IF NECESSARY, ALSO THE METAGLENE IN CONJUNCTION WITH AN INLAY REPLACEMENT. THE REVISION SURGERY TOOK PLACE ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED WITH NO ISSUES NOTED.
Patient problems: Foreign Body Reaction; Unspecified Tissue Injury
MAUDE report 22136080 ↗MalfunctionMay 29, 2025Fracture; Fracture
IT WAS REPORTED THAT THE LOCKING SCREW HEADS BROKE WHEN TRYING TO ENGAGE INTO THE METAGLENE. THERE WAS NO ADDITIONAL INFORMATION AVAILABLE. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22103635 ↗InjuryJan 9, 2025Osseointegration Problem; Osseointegration Problem
PATIENT WAS REVISED AS THE GLENOID IMPLANT WAS LOOSE. IT WAS AN UNKNOWN EVENT AND ALL IMPLANTS WERE REMOVED. SURGICAL DELAY WAS UNKNOWN. AFFECTED SIDE: UNKNOWN SHOULDER.
Patient problems: Inadequate Osseointegration
MAUDE report 21114289 ↗InjuryJan 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT HAD INFECTED SHOULDER. ALL IMPLANTS REMOVED AND ANTIBIOTIC SPACER PLACED. NO DELAYS IN SURGERY. DOI: (B)(6) 2023. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Unspecified Infection
MAUDE report 21081186 ↗InjuryJan 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A REMOVAL OF IMPLANTS DUE TO INFECTION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. NO FURTHER INFORMATION WAS PROVIDED. ACTIVITY LEVEL - YES. DOI: (B)(6) 2024. DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Unspecified Infection
MAUDE report 21072944 ↗InjuryNov 21, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENTS LEFT SHOULDER IS BEING REVISED DUE TO INSTABILITY. DOI: (B)(6) 2021. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Joint Laxity
MAUDE report 20752640 ↗InjuryNov 21, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENTS LEFT SHOULDER IS BEING REVISED DUE TO INSTABILITY. DOI: (B)(6) 2021, DOR: (B)(6) 2024, AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Joint Laxity
MAUDE report 20752329 ↗InjuryAug 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT WAS REVISED DUE INFECTION. IMPLANT WAS REMOVED. DOI: (B)(6) 2023. DOR: (B)(6) 2024. AFFECTED SIDE: UNKNOWN.
Patient problems: Unspecified Infection
MAUDE report 20054488 ↗InjuryAug 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT WAS REVISED DUE INFECTION. IMPLANT WAS REMOVED. DOI: (B)(6) 2023. DOR: (B)(6) 2024. AFFECTED SIDE: UNKNOWN.
Patient problems: Unspecified Infection
MAUDE report 20054408 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K062250, K081620, K120174, K192855. One submission often covers many models, so check the product and lot details in each record.
- Z-0834-2018 · Aug 31, 2017 · TerminatedDELTA XTEND" Reverse Shoulder System Modular Centered and Eccentric Epiphysis Implants intended for use in total shoulder or hemi-shoulder replacement procedures in patients with n
- Z-0833-2018 · Aug 31, 2017 · TerminatedDELTA XTEND" Reverse Shoulder System Modular Centered and Eccentric Epiphysis Implants are intended for use in total shoulder or hemi-shoulder replacement procedures in patients wi
- Z-0532-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard humeral PE Cup Dia 38/+3 STD, sterile, Depuy France; REF 130738203. The device is used as a shoulder implant.
- Z-0539-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42+6R RET, sterile, Depuy France; REF 130742106. The device is used as a shoulder implant.
- Z-0533-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humerus PE Cup, Dia 38/+6 STD, sterile, Depuy France; REF 130738206. The device is used as a shoulder implant.
- Z-0534-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 38/+9 STD, sterile, Depuy France; REF 130738209. The device is used as a shoulder implant.
- Z-0537-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+6 STD, sterile, Depuy France; REF 130742206. The device is used as a shoulder implant.
- Z-0535-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 38/+6R RET, sterile, Depuy France; REF 130738106. The device is used as a shoulder implant.
- Z-0538-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+9 STD, sterile, Depuy France; REF 130742209. The device is used as a shoulder implant.
- Z-0536-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+3 STD, sterile, Depuy France; REF 130742203. The device is used as a shoulder implant.
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page