NA 1363-46-000Recalls & reported problems

Depuy Orthopaedics, Inc. · MODULAR CATHCART FRACTURE HEAD HIP BALL 46mm OD +5mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy modular metal hip ball (a large femoral head used in partial hip replacement) appears in 38 adverse event reports, none involving death and 37 involving injury. Most reports describe the ball dislodging or the hip dislocating after surgery, along with infection, pain, and reactions to the metal implant; a few mention the head wearing down or degrading. Reports stayed level at 3 in each of the last two 12-month periods, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
39
Deaths reported
0
Injuries reported
38
Recalls naming it
0

Summary

The FDA's MAUDE database contains 39 adverse event reports that list this exact device (DI 10603295032786), received between Oct 2, 2017 and Aug 6, 2026 — classified as 38 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (7 reports).

1 other recall cover devices cleared under the same FDA submission (K903084); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem24
  • Device Dislodged or Dislocated7
  • Insufficient Device Problem Information5
  • Naturally Worn3
  • Degraded1

Patient problems

  • Unspecified Infection17
  • No Code Available16
  • Pain12
  • Joint Dislocation7
  • Foreign Body Reaction2
  • Not Applicable2
  • Arthritis1
  • Bone Notching1

Recent adverse event reports

The 25 most recent of 39 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 6, 2026Degraded; Naturally Worn

    Patient problems: Pain

    MAUDE report 50156424 ↗
  • InjuryJul 27, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient was revised due to infection. DOI: (b)(6) 2026.DOR: (b)(6)2026.Affected side: Right hip.

    Patient problems: Unspecified Infection

    MAUDE report 50035760 ↗
  • InjuryMay 29, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO DISLOCATION AND FRACTURE OF HEMI. DOI: (B)(6) 2026. DOR: (B)(6) 2026. AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 25334259 ↗
  • InjuryFeb 13, 2026Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO SIGNS OF INFECTION; I/D WAS PERFORMED AND HEAD/SLEEVE WERE REPLACED. DOI: (B)(6) 2025. DOR: (B)(6) 2026. AFFECTED SIDE: UNKNOWN.

    Patient problems: Unspecified Infection

    MAUDE report 24333795 ↗
  • InjuryOct 30, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ADDITIONAL INFORMATION RECEIVED. IT WAS STATED THAT VISUAL INSPECTION OF THE RETURNED DEVICES REVEALED THE METAL HEAD (MODULAR CATHCART BALL 46MM OD, (B)(4). ASSEMBLED WITH THE HEAD ADAPTOR (B)(4).

    Patient problems: Bone Notching; Foreign Body Reaction; Joint Dislocation

    MAUDE report 23427309 ↗
  • InjuryAug 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A RIGHT HIP HEMIARTHROPLASTY ON (B)(6) 2025. THE PATIENT WAS VERY UNSTABLE AND CAME IN FOR A SECOND OPINION. THE SURGEON DECIDED THAT IT WOULD BE BENEFICIAL FOR THE PATIENT TO DO A CONVERSION TO A TOTAL HIP REPLACEMENT ON THE ACETABULAR SIDE. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Laxity

    MAUDE report 22688393 ↗
  • MalfunctionApr 1, 2025Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE PATIENT REVISED FROM HEMI TO TOTAL. NO IMPLANT FAILURE. NO BROKEN PIECES. NO CASE DELAY. NO LOOSENING. THERE WAS NO PATIENT HARM. DOI: UNKNOWN. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT.

    Patient problems: Insufficient Information

    MAUDE report 21740277 ↗
  • InjurySep 19, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED ARTHRITIS IN HER ACETABULUM. UNKNOWN SURGICAL DELAY. DOI: (B)(6) 2021, DOR: (B)(6) 2024, AFFECTED SIDE: RIGHT HIP.

    Patient problems: Arthritis

    MAUDE report 20257881 ↗
  • InjurySep 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED ON (B)(6) 2024 INCLUDING IRRIGATION AND DEBRIDEMENT DUE TO INFECTION WITH HEAD EXCHANGE. PATIENT WAS ORIGINALLY IMPLANTED ON THE LEFT HIP, ON (B)(6) 2024. THERE WAS NO SURGICAL DELAY.

    Patient problems: Unspecified Infection

    MAUDE report 20250044 ↗
  • InjuryApr 15, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO ACETABULAR WEAR AND PAIN. DOI: (B)(6) 2017 DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Pain

    MAUDE report 19105612 ↗
  • InjuryFeb 16, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTS WERE REMOVED DUE TO INFECTION AND REPLACED WITH A BALL OF CEMENT, THEN REIMPLEMENTED ON (B)(6) 2024. UNKNOWN SURGICAL DELAY. DOI: (B)(6) 2024. DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 18720899 ↗
  • InjuryFeb 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT HAD A HEMI AND WAS PAINFUL POST OP SO WE CONVERTED TO THA WITH DUAL MOBILITY. IT WAS UNKNOWN, IF THERE WAS A SURGICAL DELAY. DOI- UNKNOWN. DOR- (B)(6), 2024. AFFECTED SIDE- RIGHT HIP.

    Patient problems: Pain

    MAUDE report 18699160 ↗
  • InjuryDec 6, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS INFECTED AND BOARDED FOR A HIP IRRIGATION AND DEBRIDEMENT (I&D) WITH HEAD SWAP. WHILE DISLOCATING THE FEMUR THE DOCTOR NOTICED THE STEM MOVING. THE STEM WAS LOOSE. CEMENT MANTEL WAS REMOVED AND A NEW HEMI WAS PUT IN AFTER A WASH OUT WAS PERFORMED. DOI: (B)(6) 2023 DOR: (B)(6) 2023 AFFECTED SIDE: LEFT HIP

    Patient problems: Unspecified Infection

    MAUDE report 18277205 ↗
  • InjuryAug 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THERE WAS AN INFECTION IN THE AREA. IT WAS UNKNOWN, IF THERE WAS A SURGICAL DELAY. DOI- (B)(6) 2023. DOR- UNKNOWN. AFFECTED SIDE- RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 17556035 ↗
  • InjuryDec 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION COMPLETED DUE TO FAILED HEMIARTHROPLASTY. PATIENT HAD EVIDENCE OF METALLOSIS IN THE ACETABULUM AND PSEUDOTUMOR IN CAPSULAR STRUCTURES AND GLUTEUS. REVISED TO PINNACLE 52 CUP 36 ID POLYETHYLENE LINER AND 36+6 CERAMIC TS HEAD. THERE WAS NO SURGICAL DELAY. DOI: (B)(6) 2022. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.

    Patient problems: Foreign Body Reaction; Hypersensitivity/Allergic reaction

    MAUDE report 16059877 ↗
  • InjuryJun 10, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT REMOVAL OF IMPLANTS DUE TO EROSION OF ACETABULUM/ DISLOCATION. THERE WAS NO SURGICAL DELAY. DOI: (B)(6) 2022, DOR: MAY 25,2022, AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation; Pocket Erosion

    MAUDE report 14653792 ↗
  • InjuryJun 3, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO INFECTION. NEW HEAD AND SPACER IMPLANTED. THERE WAS NO SURGICAL DELAY. DOI: UNKNOWN; DOR: (B)(6) 2022; AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 14594676 ↗
  • InjuryJan 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED ACETABULAR PAIN. DOI: (B)(6) 2021, DOE: (B)(6) 2021, LEFT HIP.

    Patient problems: Pain

    MAUDE report 13192246 ↗
  • InjuryJul 22, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT HAD HIP REVISION FROM HEMI TO TOTAL WITH A CONSTRAINED LINER DUE TO MULTIPLE DISLOCATIONS AFTER A FALL. DOI: (B)(6) 2014, DOR: (B)(6) 2020, AFFECTED SIDE: LEFT HIP.

    Patient problems: Fall; Joint Dislocation; No Code Available

    MAUDE report 10308560 ↗
  • InjuryJul 22, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT A VERY ELDERLY LADY WITH DEMENTIA WHO BECAME A CHRONIC DISLOCATER. THE SURGEON TOOK ALL IMPLANTS OUT AND LEFT A GIRDLESTONE. DOI: (B)(6) 2019; DOR: (B)(6) 2020; LEFT HIP.

    Patient problems: Joint Dislocation; No Code Available

    MAUDE report 10308006 ↗
  • InjuryJul 9, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT HAS INFECTION. DID A WASHOUT AND HEAD SWAP. DOI: (B)(6) 2020; DOR: (B)(6) 2020; AFFECTED SIDE: UNKNOWN.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 10251327 ↗
  • InjuryJun 8, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO INFECTION. DOI: (B)(6) 2019, DOR: (B)(6). 2020, LEFT HIP.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 10130416 ↗
  • InjuryJun 2, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE REVISION WAS DUE TO PAINFUL RIGHT HIP AND WAS CONVERTED HEMI TO TOTAL HIP. THERE IS NO ADDITIONAL INFORMATION. DOI: (B)(6) 2018. DOR: (B)(6) 2020; AFFECTED SIDE: RIGHT HIP.

    Patient problems: No Code Available; Pain

    MAUDE report 10108633 ↗
  • InjuryMay 5, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT HAD CHRONIC DISLOCATION. DOI: (B)(6) 2020, DOR: (B)(6), 2020; LEFT HIP.

    Patient problems: Joint Dislocation; No Code Available

    MAUDE report 10025605 ↗
  • InjuryMar 27, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE HIP BECAME INFECTED. SURGEON DID A SIMPLE WASH OUT AND REPLACED THE UNIPOLAR HEAD AND CATHCART. THE PRODUCTS WERE NOT DEFECTIVE BUT THE SURGEON REMOVED AS MUCH INFECTED TISSUE AND HARDWARE AS POSSIBLE. DOI: UNKNOWN, DOR: (B)(6) 2020, LEFT HIP.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 9891538 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K903084. One submission often covers many models, so check the product and lot details in each record.

  • Z-1375-2022 · May 18, 2022 · Open, Classified
    MODULAR CATHCART FRACTURE HEAD HIP BALL, sizes 41mm to 60mm inclusive; metallic femoral head prosthesis, Product No. 136341000, 136342000, 136343000, 136344000, 136345000, 13634600

Other Prosthesis, hip, hemi-, femoral, metal ball devices

All Prosthesis, hip, hemi-, femoral, metal ball devices →