NA 1363-47-000Recalls & reported problems

Depuy Orthopaedics, Inc. · MODULAR CATHCART FRACTURE HEAD HIP BALL 47mm OD +5mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy modular metal hip ball (Cathcart fracture head, 47mm) is the femoral head component used in partial hip replacement. The 33 reports mostly describe infection around the hip implant, along with pain, dislocation of the joint, and one case where the stem separated from the cement during revision surgery; 32 injuries and no deaths were reported. Most reports came in the last 12 months (7, versus 3 the year before). No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
0
Injuries reported
33
Recalls naming it
0

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 10603295032793), received between Oct 20, 2017 and Sep 23, 2026 — classified as 33 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).

1 other recall cover devices cleared under the same FDA submission (K903084); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 3 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem23
  • Device Dislodged or Dislocated4
  • Insufficient Device Problem Information3
  • Naturally Worn2
  • Degraded1
  • Device Contaminated During Manufacture or Shipping1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Unspecified Infection21
  • No Code Available6
  • Pain6
  • Joint Dislocation4
  • Not Applicable2
  • Fluid Discharge1
  • Joint Laxity1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • InjurySep 23, 2026Device Dislodged or Dislocated

    It was reported that the hip was dislocating over and over.DOI: (b)(6) 2026.DOR: (b)(6) 2026.Affected side: Unknown.

    Patient problems: Joint Dislocation

    MAUDE report 50872368 ↗
  • InjuryAug 27, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that Patient presented with infection near closure. Dr opted for head exchange and irrigation and debridement. Replaced head and insert. Surgery was successful.DOI: (b)(6) 2026.DOR: (b)(6) 2026.Affected Side: Left Hip.

    Patient problems: Unspecified Infection

    MAUDE report 50477459 ↗
  • InjuryMar 12, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT SHOWED SIGNS OF INFECTION WITH A DRAINING WOUND. AFFECTED SIDE: LEFT HIP.

    Patient problems: Fluid Discharge; Unspecified Infection

    MAUDE report 24582147 ↗
  • InjuryJan 7, 2026Adverse Event Without Identified Device or Use Problem

    PATIENT WAS INFECTED, ALL IMPLANTS WERE EXTRACTED AND A PROSTALAC SPACER WAS IMPLANTED. DOI: (B)(6) 2025, DOR: (B)(6) 2025, AFFECTED SIDE: UNKNOWN HIP.

    Patient problems: Unspecified Infection

    MAUDE report 23997416 ↗
  • InjuryDec 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE DUE TO INFECTION. THE STEM SEPARATED FROM THE CEMENT MANTLE WHEN DISSOCIATING THE TAPER. NO HARM TO THE PATIENT WAS REPORTED. THE PROCEDURE WAS DELAYED FOR ABOUT THIRTY (30) MINUTES WHILE THE NEW CEMENT HARDENED. DOI: (B)(6) 2025, DOR: (B)(6) 2025, AFFECTED SIDE: UNKNOWN.

    Patient problems: Unspecified Infection

    MAUDE report 23716763 ↗
  • InjuryNov 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE HIP WAS INFECTED. DOCTOR PERFORMED AN I&D AND SWAPPED THE HEAD AND TAPER ADAPTER. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 23544398 ↗
  • InjuryNov 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO INFECTION. DOI: UNKNOWN. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 23544306 ↗
  • InjuryNov 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO IRRIGATION AND DEBRIDEMENT WITH HEAD EXCHANGE DUE TO INFECTION. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 23535704 ↗
  • InjuryFeb 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO INFECTION AND WASHOUT. DOI: UNKNOWN, DOR: UNKNOWN, AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 21374650 ↗
  • InjuryJan 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO INFECTION AND NEEDED A NEW HEADBALL AND SPACER. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. AFFECTED SIDE: UNKNOWN HIP.

    Patient problems: Unspecified Infection

    MAUDE report 21126793 ↗
  • InjuryNov 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT SURGEON HAD TO REPLACE HARDWARE FROM A FRACTURED FEMUR AND NOTED AN INFECTION. DOI: UNKNOWN. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 20673910 ↗
  • InjuryJul 26, 2024Degraded; Naturally Worn; Degraded; Naturally Worn

    THAYER SCHOOL OF ENGINEERING AT DARTMOUTH FIRST REPORT OF RETRIEVALS FOR STATEMENT OF WORK 10 FOR Q2 OF 2024. A DEPUY SYNTHES IMPLANT WAS REVISED AND REVIEWED FOR ANALYSIS REASON FOR REVISION: REVISION DUE TO PAIN AND OTHER. THE SLEEVE SHOWS FRETTING AND CORROSION.

    Patient problems: Pain

    MAUDE report 19840139 ↗
  • InjuryMar 21, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2024 PATIENT WAS REVISED DUE TO INFECTION. ITEMS WERE REMOVED AND REPLACED. NO DELAY.DOI: (B)(6) 2024. DOR: (B)(6) 2024. AFFECTED SIDE: UNKNOWN.

    Patient problems: Unspecified Infection

    MAUDE report 18956552 ↗
  • InjuryOct 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT HAD INFECTION. SURGEON PERFORMED A ONE STAGE REVISION. IT WAS UNKNOWN SURGICAL DELAY. DOI: (B)(6) 2023, DOR: (B)(6) 2023, AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 18010408 ↗
  • InjuryOct 17, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT, ATTEMPTED REDUCTION OF DISLOCATED RIGHT HIP IN ER RESULTED IN ROTATION OF ENTIRE PROSTHESIS (STEM AND UNIPOLAR HEAD) IN THE FEMORAL CANAL AND HIP JOINT. NO RESULTING FEMORAL FRACTURE. IMPLANTS WERE PULLED AND REPLACED WITH REVISION TOTAL HIP COMPONENTS. THERE WAS LOOSENING OF STEM. LOOSENING INTERFACE IS NON-CEMENTED. NO ADDITIONAL INFORMATION. DOI: (B)(6) 2023. DOR: (B)(6) 2023. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 17948346 ↗
  • InjuryAug 4, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PATIENT WAS REVISED DUE TO DISLOCATION. DOI: UNKNOWN. DOR: (B)(6) 2023. AFFECTED SIDE: UNKNOWN.

    Patient problems: Joint Dislocation

    MAUDE report 17458162 ↗
  • InjuryJul 25, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT HAD IRRIGATION AND DEBRIDEMENT (I&D) OF INFECTED HEMI HIP ARTHROPLASTY. THE MODULAR HEMI HEAD AND TAPER SPACER SWAPPED OUT. UNKNOWN SURGICAL DELAY. DOI: (B)(6) 2023. DOR: (B)(6) 2023. AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 17390040 ↗
  • InjuryJul 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT HEMIARTHROPLASTY HEAD AND SPACER WAS EXCHANGED DUE TO INFECTION. DOI: (B)(6) 2023. DOR: (B)(6) 2023. AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 17333840 ↗
  • InjuryJan 4, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD I&D WITH HEAD EXCHANGE FOR INFECTION. THERE WAS NO SURGICAL DELAY. DOI: (B)(6) 2022. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 16093768 ↗
  • InjuryMay 23, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION COMPLETED DUE TO RIGHT HIP PAIN FROM THE UNIPOLAR HEAD IN THE SUPERIOR ACETABULUM. THE UNIPOLAR HEAD WAS REMOVED AND THE STEM REMAINED IN SITU. DOI: (B)(6) 2017, DOR: (B)(6) 2022, AFFECTED SIDE: RIGHT HIP.

    Patient problems: Pain

    MAUDE report 14472031 ↗
  • InjuryMar 16, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO PAIN. DOI: (B)(6), 2018. DOR: (B)(6), 2022. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Pain

    MAUDE report 13790428 ↗
  • InjuryJan 20, 2022Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THERE WAS AN INFECTED HIP HEMI ARTHROPLASTY. PERFORMED INCISION AND DEBRIS WENT AND CHANGED THE UNIPOLAR HEAD AND SPACER. NO ADDITIONAL INFORMATION IS AVAILABLE. DOI: (B)(6) 2021; DOR: (B)(6) 2022; AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 13306714 ↗
  • MalfunctionJun 18, 2021Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    DURING OPENING, THE MATERIAL STICKER WAS ALREADY ON THE IMPLANT - THE CUSTOMER DID NOT WANT DO IMPLANT THAT. A NEW IMPLANT WAS OPENED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12025496 ↗
  • InjuryJan 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE SURGEON CONVERTED A HEMI TO S THA DUE TO INSTABILITY FROM THE GREATER TROCHANTER PULLING OFF THE FEMUR. DOI: (B)(6) 2021 - DOR: (B)(6) 2021 (RIGHT HIP).

    Patient problems: Joint Laxity

    MAUDE report 11193553 ↗
  • InjuryNov 16, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    PATIENT HAD HEMIARTHROPLASTY PERFORMED (B)(6) 2020. WAS BROUGHT BACK TO OR DUE TO FOREIGN BODY (CEMENT) IN ACETABULUM AND SUBLUXATION. THE FEMORAL STEM WAS NOT EXPLANTED. THE STEM WAS NOT LOOSE AND DID NOT MOVE. AFTER FOREIGN BODY REMOVED IT WAS DETERMINED THE 47MM HEAD WAS TOO LARGE AND A 46MM WAS THEN USED. TAPERED NECK (-3) WAS ALSO REPLACED. INTRA OP X-RAYS TAKEN TO CONFIRM PLACEMENT.

    Patient problems: No Code Available

    MAUDE report 10843170 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K903084. One submission often covers many models, so check the product and lot details in each record.

  • Z-1375-2022 · May 18, 2022 · Open, Classified
    MODULAR CATHCART FRACTURE HEAD HIP BALL, sizes 41mm to 60mm inclusive; metallic femoral head prosthesis, Product No. 136341000, 136342000, 136343000, 136344000, 136345000, 13634600

Other Prosthesis, hip, hemi-, femoral, metal ball devices

All Prosthesis, hip, hemi-, femoral, metal ball devices →