Skip to content

ATTUNE 1504-10-107Recalls & reported problems

Depuy (Ireland) · ATTUNE KNEE SYSTEM FEMORAL POSTERIOR STABILIZED SIZE 7 LEFT CEMENTED · Class 3 · Implant

All models of ATTUNE (1475)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Attune knee implant is part of a total knee replacement. Many reports describe infections that required the implant to be removed, and others describe loosening, pain and loss of motion. The 159 reports include no deaths and 148 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
158
Deaths reported
0
Injuries reported
147
Recalls naming it
0

Summary

The FDA's MAUDE database contains 158 adverse event reports that list this exact device (DI 10603295041641), received between Oct 18, 2017 and Jul 29, 2026 — classified as 147 injuries, 11 malfunctions.

The most frequently reported specific device problem is “Loss of or Failure to Bond” (31 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 2 reportsApr 2024: 2 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem97
  • Loss of or Failure to Bond31
  • Insufficient Device Problem Information11
  • Difficult to Open or Remove Packaging Material8
  • Malposition of Device4
  • Use of Device Problem4
  • Device Contaminated During Manufacture or Shipping3
  • Loosening of Implant Not Related to Bone-Ingrowth2

Patient problems

  • Pain66
  • No Code Available64
  • Unspecified Infection32
  • Adhesion(s)21
  • Joint Laxity16
  • Insufficient Information11
  • Loss of Range of Motion11
  • Not Applicable8

Recent adverse event reports

The 25 most recent of 158 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 29, 2026Migration

    It was reported that the femur is in extreme valgus.DOI: Unknown.DOR: (b)(6)2026.Affected side: Left Hip.

    Patient problems: Insufficient Information

    MAUDE report 50077230 ↗
  • InjuryMar 10, 2026Adverse Event Without Identified Device or Use Problem

    SUBJECT ID: (B)(6). STUDY NO: (B)(4). CLINICAL ADVERSE EVENT RECEIVED FOR REVISION- INFECTION DEVICE RELATED: NO INFORMATION PROVIDED. PROCEDURE RELATED: NO INFORMATION PROVIDED. DATE OF EVENT: NO INFORMATION PROVIDED. DATE OF IMPLANT: (B)(6) 2025. DATE OF REVISION: (B)(6) 2026. DEVICE LOCATION: LEFT. TREATMENT/IMPACT: WERE DEPUY SYNTHES COMPONENTS REMOVED: YES (FEMORAL, TIBIAL).

    Patient problems: Unspecified Infection

    MAUDE report 24560750 ↗
  • InjuryJan 30, 2026Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO INFECTION. LEFT TOTAL KNEE ARTHROPLASTY WAS DONE. DOI: (B)(6) 2021. DOR: (B)(6) 2026. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Unspecified Infection

    MAUDE report 24221453 ↗
  • InjuryOct 15, 2025Loss of or Failure to Bond; Loss of or Failure to Bond

    SUBJECT ID: (B)(6) STUDY NO: (B)(6) CLINICAL ADVERSE EVENT RECEIVED FOR FAILURE OF LEFT TOTAL KNEE ARTHROPLASTY. LOOSENING OF TIBIAL AND FEMORAL COMPONENTS. MEDDRA CODE - 10021539 DEVICE RELATED: DEFINITELY RELATED PROCEDURE RELATED: POSSIBLY RELATED DATE OF EVENT: (B)(6) 2024 DATE OF IMPLANT: (B)(6) 2013 DATE OF REVISION: (B)(6) 2024 DEVICE LOCATION: LEFT TREATMENT/IMPACT: REQUIRES IN PATIENT HOSPITALIZATION, RESULTS IN A MEDICAL OR SURGICAL INTERVENTION TO PREVENT LIFE-THREATENING ILLNESS OR INJURY OR PERMANENT IMPAIRMENT TO A BODY STRUCTURE OR FUNCTION, MEDICAL INTERVENTION/MODIFICATION, REVISION OF LEFT TOTAL KNEE ARTHROPLASTY, COMPONENTS REMOVED: FEMORAL COMPONENT, TIBIAL BASE, TIBIAL INSERT, AND PATELLAR COMPONENT (FOR PREVIOUSLY RESURFACED PATELLA).

    Patient problems: Insufficient Information

    MAUDE report 23294716 ↗
  • InjuryOct 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PATIENT WAS INFECTED WITH MERSA SO ALL OF HER PREVIOUS IMPLANTS WERE TAKEN OUT TO BE REPLACED WITH ANTIBIOTIC SPACERS. NO PATIENT CONSEQUENCES

    Patient problems: Unspecified Infection

    MAUDE report 23289986 ↗
  • InjurySep 11, 2025Loss of or Failure to Bond; Loss of or Failure to Bond

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO ASEPTIC LOOSENING OF FEMUR AT CEMENT TO IMPLANT INTERFACE. DEPUY CEMENT WAS USED. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Insufficient Information

    MAUDE report 23024849 ↗
  • InjuryAug 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS EXPLANTED DUE TO INFECTION. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Unspecified Infection

    MAUDE report 22689303 ↗
  • InjuryJul 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED AS A STAGE 1 KNEE REVISION. SURGEON TOOK OUT ALL COMPONENTS AND PUT IN AN ANTIBIOTIC SPACER TO RULE OUT INFECTION. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Unspecified Infection

    MAUDE report 22522482 ↗
  • MalfunctionJul 10, 2025Difficult to Open or Remove Packaging Material; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT THE FEMORAL COMPONENT WASN'T ABLE TO BE DELIVERED TO STERILE FIELD. IT WAS STUCK IN THE OUTER CONTAINER. AFFECTED SIDE: LEFT KNEE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22468965 ↗
  • MalfunctionApr 29, 2025Difficult to Open or Remove Packaging Material; Difficult to Open or Remove Packaging Material

    ADDITIONAL INFORMATION WAS RECEIVED: A. WAS THERE ANY PATIENT HARM/CONSEQUENCE RELATED TO THE EVENT? NO PATIENT WAS HARMED. B. WAS THERE ANY SURGICAL DELAY RELATED TO THE EVENT? CASE WAS DELAYED. C. COULD YOU PLEASE PROVIDE VALID THE LOT NUMBER OF THE REPORTED PRODUCT? THE STICKER IS WITH THE FEMUR THAT WAS RETURNED BACK TO THE OFFICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21930036 ↗
  • MalfunctionNov 18, 2024Difficult to Open or Remove Packaging Material; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT THE STERILE IMPLANT WAS NOT ABLE TO BE DELIVERED TO STERILE FIELD. IT WAS STUCK IN OUTER BOX. DOE: (B)(6) 2024. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Insufficient Information

    MAUDE report 20711221 ↗
  • MalfunctionSep 26, 2024Difficult to Open or Remove Packaging Material; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT THE STERILE IMPLANT WAS NOT ABLE TO BE TAKEN OUT OF THE BOX BY OR PERSONNEL. IT WAS LIKE IT WAS STUCK INSIDE OUTER BOX. DOE: (B)(6) 2024. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Insufficient Information

    MAUDE report 20310504 ↗
  • InjuryAug 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED WITH CHRONIC DRAINAGE FROM KNEE AND SCHEDULED FOR IRRIGATION AND DEBRIDEMENT (I&D) WITH HARDWARE REMOVAL ON (B)(6) 2024. DOI: (B)(6) 2024. DOE: (B)(6) 2024. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Fluid Discharge; Unspecified Infection

    MAUDE report 19992138 ↗
  • InjuryJul 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PLANNED INLAY REVISION OF PS-ATTUNE-KNEE-TEP LEFT SIDE AFTER ABOUT ONE YEAR OF IMPLANTATION BECAUSE PATIENT HAD ISSUES STRETCHING THE KNEE. INTRAOPERATIVELY THE PE-INLAY SHOWED STRANGE SCRATCHES. SURGEON DECIDED TO CHANGE COMPLETE KNEE-TEP TO A COMPETITOR PRODUCT.

    Patient problems: Loss of Range of Motion

    MAUDE report 19836932 ↗
  • InjuryMay 1, 2024Loss of or Failure to Bond; Loss of or Failure to Bond

    PATIENT WAS REVISE DUE TO AESEPTIC LOOSENING OF THE PATELLA, TIBIA AND FEMORAL COMPONENT AT THE CEMENT TO IMPLANT INTERFACE. DEPUY CEMENT WAS USED. FULL ATTUNE CRS WAS RE-IMPLANTED, THERE WAS NO X-RAYS AVAILABLE AT THIS TIME, IMPLANTS NOT BEING RETURNED. DOI: (B)(6) 2019, DOR: (B)(6) 2024, AFFECTED SIDE: LEFT KNEE.

    Patient problems: Insufficient Information

    MAUDE report 19219914 ↗
  • InjuryApr 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS BEING REVISED DUE TO INFECTION. ALL HARDWARE FROM A PRIMARY TOTAL KNEE WAS REMOVED AND A CEMENT SPACER WAS PLACED IN THE PATIENT. DOI: UNKNOWN, DOR: (B)(6) 2024, AFFECTED SIDE: LEFT KNEE.

    Patient problems: Unspecified Infection

    MAUDE report 19192040 ↗
  • InjuryApr 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION WAS RECEIVED: WAS THERE ANY SURGICAL DELAY RELATED TO THE EVENT? NO. WHAT IS SPECIFIC ALLEGATION AGAINST THE REPORTED PRODUCT? SUSPECTED INFECTION, NO WAY TO VERIFY.

    Patient problems: Unspecified Infection

    MAUDE report 19042812 ↗
  • InjuryMar 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS PRESENTED WITH INFECTION AND INSTABILITY DUE TO LOOSENING OF THE PATELLA AND FEMORAL COMPONENTS AT THE BONE TO CEMENT INTERFACE. CEMENT MANUFACTURER WAS UNKNOWN. ALL IMPLANTS WERE REVISED. DOI: UNKNOWN. DOR: (B)(6) 2024. AFFECTED SIDE: UNKNOWN.

    Patient problems: Joint Laxity; Unspecified Infection

    MAUDE report 18979964 ↗
  • InjuryMar 1, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO KNEE PAIN AND STIFFNESS. SURGICAL DELAY UNKNOWN. ACTIVITY LEVEL - YES DOI: (B)(6) 2019 DOR: (B)(6) 2024 AFFECTED SIDE: LEFT KNEE

    Patient problems: Muscular Rigidity; Pain

    MAUDE report 18815602 ↗
  • InjuryFeb 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS PRESENTED AT HOSPITAL WITH INFECTED KNEE, 10 MONTHS POST TKA. PLAN TO I&D. PROBABLE TKA REMOVAL ON (B)(6) 2024. PLAN TO REMOVE. IT WAS UNKNOWN BOUT SURGICAL DELAY, DOI: (B)(6) 2023. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Unspecified Infection

    MAUDE report 18766566 ↗
  • InjuryFeb 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT HAD INSTABILITY AND NEEDED A REVISION. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Joint Laxity

    MAUDE report 18711105 ↗
  • InjuryDec 15, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A POOR MOTION AND COMPLAINED OF PAIN AND INSTABILITY. A REVISION TKA WAS PERFORMED UNEVENTFULLY. A SIGNIFICANT POSTERIOR RELEASE WAS PERFORMED AND ADDITIONAL DISTAL FEMORAL RESECTION WAS NECESSARY TO GAIN EXTENSION. NO PATIENT HARM OCCURRED. NO DELAY WAS ENCOUNTERED. DOI: (B)(6) 2023, DOR: (B)(6) 2023, AFFECTED SIDE: LEFT KNEE.

    Patient problems: Joint Laxity; Loss of Range of Motion; Pain

    MAUDE report 18337847 ↗
  • InjuryDec 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO PAINFUL KNEE. NO FURTHER INFORMATION IS AVAILABLE. DOI: (B)(6) 2020. DOR: (B)(6) 2023. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Pain

    MAUDE report 18319467 ↗
  • InjuryOct 6, 2023Loss of or Failure to Bond; Adverse Event Without Identified Device or Use Problem; Loss of or Failure to Bond; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED INDICATES THAT THE PATIENT WAS EXPERIENCING PAIN.

    Patient problems: Insufficient Information; Pain

    MAUDE report 17886025 ↗
  • InjuryAug 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    VOLUNTARY REPORT FROM HEALTH AUTHORITY.

    Patient problems: Pain

    MAUDE report 17583779 ↗

Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices

All Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page