ATTUNE 1506-60-002Recalls & reported problems
Depuy (Ireland) · ATTUNE KNEE SYSTEM REVISION ROTATING PLATFORM TIBIAL BASE CEMENTED SIZE 2 · Class 3 · Implant
All models of ATTUNE (1475)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10603295042860), received between Sep 23, 2021 and Sep 19, 2025 — classified as 8 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Contaminated During Manufacture or Shipping” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem8
- Device Contaminated During Manufacture or Shipping1
- Difficult to Open or Remove Packaging Material1
- Off-Label Use1
Patient problems
- Unspecified Infection3
- Fall2
- Insufficient Information2
- Bone Fracture(s)1
- Fluid Discharge1
- Loss of Range of Motion1
- Muscular Rigidity1
- Pain1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjurySep 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED WITH DRAINAGE SUGGESTING HE WAS STILL INFECTED. ALL WELL FIXED, WELL ALIGNED COMPONENTS WERE REMOVED. A THOROUGH DEBRIDEMENT WAS PERFORMED. BONE WAS THEN PREPARED FOR NEW IMPLANTS WHICH ARE PLACED WITH HIGH DOSE ANTIBIOTIC CEMENT ALONG WITH RESORBABLE BEADS THAT CONTAINED BOTH ANTIFUNGAL AND ANTIBIOTICS. NO DELAY WAS NOTED. PREVIOUS CEMENT WAS DEPUY 2 WITH GENT. AFFECTED SIDE: LEFT KNEE.
Patient problems: Fluid Discharge; Unspecified Infection
MAUDE report 23099807 ↗InjuryJun 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS ORIGINALLY OPERATED ON FOR AN INFECTED NON-UNION OF FUSION. SINCE CULTURES WERE NEGATIVE, SURGEON CHOSE TO REIMPLANT THE PATIENT. CONTINUED DRAINAGE SUGGESTS INFECTION. SINCE THE PATIENT CONTINUES TO DRAIN DESPITE NEGATIVE CULTURES, THE SURGEON CHOSE TO REMOVE ALL COMPONENTS AND CONVERT TO A STATIC SPACER UTILIZING LPS COMPONENTS, PLACEMENT OF RESORBABLE ANTIBIOTIC BEADS AND HIGH DOSE ANTIBIOTIC CEMENT TO COAT THE COMPONENTS. ALL COMPONENTS WERE WELL FIXED BUT LACK OF BONY INGROWTH ALSO SUGGESTS PERSISTENT INFECTION. AND I&D WAS PERFORMED ON (B)(6) 2025, WHERE THE POLY AND PIN WERE REPLACED. DEPUY 2 WITH GENTAMICIN WAS LIKELY UTILIZED, AS THAT IS THE SURGEON'S NORM. NO DELAY OCCURRED AND NO PATIENT HARM WAS NOTED. DOI: (B)(6) 2025, DOR: (B)(6) 2025, AFFECTED SIDE: LEFT KNEE.
Patient problems: Unspecified Infection
MAUDE report 22214727 ↗InjuryDec 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT FELL AND FRACTURED TIBIA. THE TIBIA IMPLANTS WERE EXPLANTED, AND NEW ONES WERE INSERTED WITH A LONGER STEM TO BYPASS THE FRACTURE. THERE WAS NO SURGICAL DELAY OR CONSEQUENCES TO THE PATIENT. DOI: (B)(6) 2023. DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT KNEE.
Patient problems: Bone Fracture(s); Fall
MAUDE report 21018809 ↗InjuryNov 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SUBJECT ID: (B)(6), STUDY NO: (B)(6). CLINICAL ADVERSE EVENT RECEIVED, FOR RIGHT KNEE MOBILIZATION: DEVICE (RELATEDNESS): POSSIBLY. PROCEDURE (RELATEDNESS): PROBABLY. DATE OF EVENT: (B)(6) 2024. DATE OF REVISION: NO REVISION. DATE OF IMPLANT: (B)(6) 2024. DEVICE LOCATION: RIGHT. TREATMENT/IMPACT: HOSPITALIZATION/CLOSED REDUCTION/MANIPULATION.
Patient problems: Insufficient Information
MAUDE report 20693254 ↗InjurySep 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
CLINICAL ADVERSE EVENT RECEIVED FOR RIGHT KNEE MOBILIZATION IS THIS EVENT RELATED TO THE STUDY DEVICE? POSSIBLY IS THIS EVENT RELATED TO THE STUDY PROCEDURE? PROBABLY. DEVICE LOCATION: RIGHT. DATE OF EVENT: 05 SEP 2024. DATE OF IMPLANT: (B)(6) 2024. TREATMENT/IMPACT: CLOSED REDUCTION/MANIPULATION.
Patient problems: Loss of Range of Motion; Muscular Rigidity
MAUDE report 20278260 ↗InjuryAug 18, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM DEPUY SYNTHES JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PRIMARY SURGERY WAS PERFORMED WITH THE IMPLANTS IN QUESTION. AFTER THE SURGERY, THE PATIENT FELL AND HAD A PERIPROSTHETIC FEMORAL IMPLANT FRACTURE. THE REVISION SURGERY IS SCHEDULED ON AN UNKNOWN DATE TO FIX THE AFFECTED AREA WITH PLATE. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Fall
MAUDE report 17574686 ↗InjuryOct 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT IMPLANTS WERE INSERTED ON 2 DIFFERENT DATES AS A PARTIAL REVISION WAS PERFORMED ON THE SECOND DATE, (B)(6) 21. A DER WAS SUBMITTED AT THAT TIME. FIRST 3 ITEMS LISTED WERE LEFT FROM (B)(6) 2020 PROCEDURE, THE REST WERE PLACED IN (B)(6) 2021. PATIENT COMPLAINED OF TIBIAL PAIN. XRAYS APPEARED TO HAVE LUCENCIES ON THE TIBIAL SIDE SO A TIBIAL SIDE REVISION WAS PLANNED. INSERT AND FEMORAL COMPONENT WERE REMOVED FOR EXPOSURE, THEN TIBIAL TRAY AND STEM WERE REMOVED, BUT THE TIBIAL SLEEVE WAS FOUND TO BE WELL FIXED. THE SURGEON WANTED TO SHORTEN THE ADAPTER ON THE FEMORAL SLEEVE AND WHEN TRYING TO REMOVE THE ADAPTER THE FEMORAL SLEEVE AND STEM WERE EXTRACTED. AT THIS POINT, A TIBIAL AND FEMORAL REVISION WERE PERFUMED UNEVENTFULLY. THERE WAS NO PATIENT HARM. DOI: (B)(6) 2020, DOR: (B)(6) 2022, AFFECTED SIDE: RIGHT HIP.
Patient problems: Pain
MAUDE report 15526019 ↗InjuryApr 26, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT REMAINS INFECTED. ALL COMPONENTS WERE REMOVED AND A SPACER WAS FASHIONED UTILIZING LPS COMPONENTS. CEMENT WAS DEPUY 2 WITH GENT BUT NO QUANTITY OR LOT INFORMATION IS AVAILABLE AS IT IS HOSPITAL OWNED INVENTORY. IT WAS ALSO REPORTED THAT TIBIAL AND FEMORAL COMPONENTS WERE LOOSE AT THE BONE TO IMPLANT INTERFACE. DOI: (B)(6) 2022; DOR: (B)(6) 2022; AFFECTED SIDE: RIGHT KNEE.
Patient problems: Unspecified Infection
MAUDE report 14207955 ↗MalfunctionSep 23, 2021Off-Label Use; Difficult to Open or Remove Packaging Material; Device Contaminated During Manufacture or Shipping; Off-Label Use; Difficult to Open or Remove Packaging Material; Device Contaminated During Manufacture or Shipping
THIS PC IS RELATED TO (B)(4) WHICH REPORTS ABOUT THE INFECTION AFTER THE FIRST SURGERY. THIS PC REPORTS ABOUT THE TRACES OF DIRT ON THE SURFACE OF THE TRAY USED IN THE REVISION SURGERY. ON (B)(6) 2021, THE PATIENT UNDERWENT THE REVISION SURGERY VIA TKA DUE TO INFECTION. THE TRAY IN QUESTION WAS PREPARED FOR THE REVISION SURGERY. IN THE REVISION SURGERY, IT WAS FOUND THAT THERE WERE TRACES OF DIRT ON THE SURFACE OF THE TRAY. THE SURGEON WIPED THE TRAY WITH A GAUGE, AND IT WAS USED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN 30 MINUTES DELAY. THE SALES REP RECOMMENDED TO USE ANOTHER NEW TRAY, BUT THE SURGEON SAID THAT THERE IS NO TRAY WITHOUT TRACES OF DIRT OR A PROTECTIVE COVER ON THE J&J PRODUCT. AND THE SURGEON USED THE DIRTY TRAY AS IT WAS.
Patient problems: Insufficient Information
MAUDE report 12519572 ↗
Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices
- ATTUNE 1506-10-005Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing442 reports
- ATTUNE 1506-10-004Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing442 reports
- ATTUNE 1518-20-038Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing3 deaths393 reports
- LCS SIGMA 1294-33-130Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing1 recall310 reports
- ATTUNE 1518-20-035Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing342 reports
- P.F.C. SIGMA 96-0102Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing1 death319 reports
- ATTUNE 1518-20-041Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing308 reports
- ATTUNE 1506-10-007Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing284 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page