ATTUNE 1516-50-310Recalls & reported problems
Depuy (Ireland) · ATTUNE KNEE SYSTEM TIBIAL INSERT ROTATING PLATFORM POSTERIOR STABILIZED SIZE 3 10mm AOX · Class 3 · Implant
All models of ATTUNE (1475)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10603295052180), received between Jun 8, 2020 and Jun 30, 2025 — classified as 5 injuries.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem5
Patient problems
- Pain3
- Joint Laxity2
- Adhesion(s)1
- No Code Available1
- Unspecified Tissue Injury1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJun 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED TO UPSIZE POLY FOR KNEE STABILITY DUE TO LIGAMENT INSUFFICIENCY. DOI: UNKNOWN DOR: JUNE 11,2025 AFFECTED SIDE: UNKNOWN.
Patient problems: Joint Laxity; Unspecified Tissue Injury
MAUDE report 22358872 ↗InjuryAug 11, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT WAS REVISED DUE TO INSTABILITY. DOI: (B)(6) 2023 DOR: (B)(6) 2023 AFFECTED SIDE: RIGHT KNEE.
Patient problems: Joint Laxity
MAUDE report 17511022 ↗InjuryJun 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED TO ADDRESS PAIN AND LOOSENING OF THE TIBIAL TRAY AT THE CEMENT TO IMPLANT INTERFACE. COMPETITOR CEMENT WAS USED. THE PATIENT RECENTLY COMPLAINED OF KNEE PAIN AFTER HAVING GONE SEVERAL YEARS PAIN FREE. BONE SCAN INDICATED A POTENTIALLY LOOSE TIBIA. THE SURGEON DECIDED TO ASSESS THE FIXATION OF THE TIBIAL TRAY, HE DESCRIBED IT AS NEITHER LOOSE NOR WELL FIXED. WITH A MILD BLOW WITH A MALLET AND BONE TAMP, THE TRAY WAS REMOVED. NO CEMENT REMAINED ATTACHED TO THE MEAL TRAY. SURGEON QUESTIONS THE FIXATION OF THE FIRST GENERATION ATTUNE TRAYS. DOI: (B)(6) 2015, LEFT KNEE.
Patient problems: Adhesion(s); Pain
MAUDE report 14874147 ↗InjuryMar 5, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED DUE TO PAIN TKA. DOI: UNKNOWN; DOR: (B)(6) 2021 LEFT KNEE.
Patient problems: Pain
MAUDE report 11424232 ↗InjuryJun 8, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT RECEIVED LEFT PRIMARY ATTUNE TKA TO TREAT END-STAGE OSTEOARTHRITIS. THE PATELLA WAS RESURFACED, AND DEPUY CEMENT X 2 WAS UTILIZED. THE SURGEON NOTED THAT THE PATIENT HAD POSTEROLATERAL INSUFFICIENCY DUE TO AN OLD ACL INJURY AND THAT THE POPLITEUS TENDON WAS ABSENT. THE PROCEDURE WAS COMPLETED WITHOUT COMPLICATIONS. PATIENT RECEIVE A LEFT TOTAL REVISION TO TREAT PAIN AND INSTABILITY. UPON ENTERING THE JOINT, THE SURGEON CONFIRMED GROSS INSTABILITY OF THE LEFT KNEE. THE TIBIAL TRAY AND FEMORAL COMPONENT WERE WELL-FIXED BUT REVISED. THERE WAS NO REPORTED PRODUCT PROBLEM WITH THE EXPLANTED TIBIAL INSERT. THE PATELLA WAS RETAINED. THE SURGEONS NOTE THE PATELLAR TENDON WAS THINNING AT THE TIBIAL TUBERCLE AND WAS REATTACHED WITH SUTURE. THE PATIENT RECEIVED COMPETITOR REVISION PRODUCTS UTILIZING COMPETITOR CEMENT. THE PROCEDURE WAS COMPLETED WITHOUT COMPLICATIONS. DOI: (B)(6) 2015. DOR: (B)(6) 2016. LEFT KNEE.
Patient problems: No Code Available; Pain
MAUDE report 10130014 ↗
Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices
- ATTUNE 1506-10-005Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing442 reports
- ATTUNE 1506-10-004Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing442 reports
- ATTUNE 1518-20-038Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing3 deaths393 reports
- LCS SIGMA 1294-33-130Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing1 recall310 reports
- ATTUNE 1518-20-035Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing342 reports
- P.F.C. SIGMA 96-0102Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing1 death319 reports
- ATTUNE 1518-20-041Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing308 reports
- ATTUNE 1506-10-007Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing284 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.