NA 2001-66-000Recalls & reported problems
Depuy (Ireland) · FEMORAL HD IMPACTOR REPL TIP · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy femoral head impactor tip is a surgical tool used to seat the ball of a hip implant. Many of its 88 reports describe the tip wearing, scratching, denting, or breaking, sometimes into two pieces. The reports include no deaths and no injuries. Reports rose from 3 in the prior 12 months to 11 in the last 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 88
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 88 adverse event reports that list this exact device (DI 10603295080138), received between Sep 20, 2017 and Jun 5, 2026 — classified as 88 malfunctions.
The most frequently reported specific device problem is “Naturally Worn” (46 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Naturally Worn46
- Break22
- Material Twisted/Bent21
- Scratched Material21
- Material Deformation16
- Crack4
- Device-Device Incompatibility4
- Material Discolored2
Patient problems
- No Clinical Signs, Symptoms or Conditions66
- No Consequences Or Impact To Patient21
- No Code Available5
- No Information1
Recent adverse event reports
The 25 most recent of 88 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 5, 2026Break; Crack
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT THE DEVICE CRACKED AND BROKE INTO 2 PIECES. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. IT WAS REPORTED THAT THERE WAS NO FURTHER INFORMATION REGARDING THE ISSUE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25409835 ↗MalfunctionMay 21, 2026Material Deformation; Scratched Material
IT WAS REPORTED THAT THE IMPACTOR TIPS DENTED AND SCRATCHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25238231 ↗MalfunctionMay 21, 2026Material Deformation; Scratched Material
IT WAS REPORTED THAT THE IMPACTOR TIPS DENTED AND SCRATCHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25238230 ↗MalfunctionMay 21, 2026Material Deformation; Scratched Material
IT WAS REPORTED THAT THE IMPACTOR TIPS DENTED AND SCRATCHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25238229 ↗MalfunctionMay 21, 2026Material Deformation; Scratched Material
IT WAS REPORTED THAT THE IMPACTOR TIPS DENTED AND SCRATCHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25238228 ↗MalfunctionApr 21, 2026Material Deformation; Scratched Material
IT WAS REPORTED THAT THE FEMORAL HEAD IMPACTOR TIP WAS DENTED, SCRATCHED, AND DEFORMED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24942483 ↗MalfunctionApr 16, 2026Break
IT WAS REPORTED THAT THE HEAD OF THE IMPACTOR TIP, BROKE INTO 2 PIECES, ALL PIECES RECOVERED. THE ISSUE WAS OBSERVED DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. IT WAS REPORTED THAT THERE WAS NO FURTHER INFORMATION REGARDING THE ISSUE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24897115 ↗MalfunctionFeb 20, 2026Break
IT WAS REPORTED THAT THE REPL TIP FEMORAL HD IMPACTOR AND MODULAR CALCAR PLANER LRG WERE DAMAGED/BROKEN FOLLOWING SURGERY AND/OR DURING THE WASH. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24401784 ↗MalfunctionDec 31, 2025Break; Break
IT WAS REPORTED THAT THE HEAD PUSHER TIPS WERE BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23936763 ↗MalfunctionDec 31, 2025Break; Break
IT WAS REPORTED THAT THE HEAD PUSHER TIPS WERE BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23936762 ↗MalfunctionDec 31, 2025Break; Break
IT WAS REPORTED THAT THE HEAD PUSHER TIPS WERE BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23936761 ↗MalfunctionJul 3, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT THE BOSS ADAPTOR CROSS THREADED STEMS NO LONGER THREAD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22398593 ↗MalfunctionJul 3, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT THE BOSS ADAPTOR CROSS THREADED STEMS NO LONGER THREAD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22398584 ↗MalfunctionJul 3, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT THE BOSS ADAPTOR CROSS THREADED STEMS NO LONGER THREAD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22398579 ↗MalfunctionOct 15, 2024Material Deformation; Naturally Worn; Material Deformation; Naturally Worn
IT WAS REPORTED THAT THE INTERNAL THREADS OF THE DEVICE WERE DAMAGED. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20451447 ↗MalfunctionMar 13, 2024Break; Break
IT WAS REPORTED THAT THE SALES REP WAS INFORMED THAT THE HEAD IMPACTOR BROKE DURING CLEANING. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18894366 ↗MalfunctionNov 16, 2023Break; Break
IT WAS REPORTED THAT BOTH INSERTER AND HEAD IMPACTORS ARE CHIPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18156106 ↗MalfunctionNov 16, 2023Break; Break
IT WAS REPORTED THAT BOTH INSERTER AND HEAD IMPACTORS ARE CHIPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18156100 ↗MalfunctionJun 20, 2023Naturally Worn; Naturally Worn
ADDITIONAL INFORMATION WAS RECEIVED AND STATED THAT THE INCIDENT DID NOT OCCUR DURING SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17165008 ↗MalfunctionJun 1, 2023Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE ITEM WAS NOT FIT FOR PATIENT USE DUE TO WEAR AND TEAR. THE WEAR AND TEAR IS GENERAL WEAR AND TEAR WE SEE ON THIS SET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17041591 ↗MalfunctionMay 31, 2023Naturally Worn; Naturally Worn
DEVICE WAS REPORTED FOR GENERAL WEAR AND TEAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17032417 ↗MalfunctionApr 4, 2023Naturally Worn; Scratched Material; Naturally Worn; Scratched Material
IT WAS REPORTED THAT DEVICE IS DAMAGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16676265 ↗MalfunctionApr 4, 2023Naturally Worn; Scratched Material; Naturally Worn; Scratched Material
IT WAS REPORTED THAT TRIAL HEAD IS DAMAGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16674933 ↗MalfunctionDec 15, 2022Naturally Worn; Naturally Worn
STAFF HAVE INDICATED SEVERAL ACETABULAR REAMER HANDLES ARE OLD, CRACKED AND NEED TO BE REPLACED DUE TO WEAR AND TEAR OVER MANY CASES. THEY HAVE ALSO INDICATED 6X FEMORAL HEAD IMPACTOR TIPS ARE WEARING AWAY ALSO DUE TO WEAR AND TEAR OVER TIME AND REQUIRE REPLACEMENT. TWO CORAIL BROACH HANDLES ARE EXTREMELY LOOSE AND NO LONGER FUNCTION PROPERLY WITH THE TRIAL BROACH. LASTLY, A PINNACLE VERSION GUIDE REQUIRES REPLACEMENT AS YOU CAN NO LONGER READ THE WRITING ON THE TOP OF THE INSTRUMENT AS IT¿S FADED OVER TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15992833 ↗MalfunctionDec 15, 2022Naturally Worn; Naturally Worn
STAFF HAVE INDICATED SEVERAL ACETABULAR REAMER HANDLES ARE OLD, CRACKED AND NEED TO BE REPLACED DUE TO WEAR AND TEAR OVER MANY CASES. THEY HAVE ALSO INDICATED 6X FEMORAL HEAD IMPACTOR TIPS ARE WEARING AWAY ALSO DUE TO WEAR AND TEAR OVER TIME AND REQUIRE REPLACEMENT. TWO CORAIL BROACH HANDLES ARE EXTREMELY LOOSE AND NO LONGER FUNCTION PROPERLY WITH THE TRIAL BROACH. LASTLY, A PINNACLE VERSION GUIDE REQUIRES REPLACEMENT AS YOU CAN NO LONGER READ THE WRITING ON THE TOP OF THE INSTRUMENT AS IT¿S FADED OVER TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15992826 ↗
FDA clearance and classification
The FDA regulates product code HWA (impactor) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 2001-66-000
Has the NA 2001-66-000 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10603295080138), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 2001-66-000?
The FDA's MAUDE database lists 88 adverse event reports for this device, received between Sep 20, 2017 and Jun 5, 2026. In those reports, the most common specific device problems are “Naturally Worn” (46), “Break” (22) and “Material Twisted/Bent” (21). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 2001-66-000?
None of the reports are classified as deaths, 0 as injuries and 88 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 2001-66-000 FDA approved?
Devices of this type (product code HWA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).
Who makes the NA 2001-66-000, and is it still sold?
The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand NA, model 2001-66-000). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other IMPACTOR devices
- PINNACLE 2217-50-041Depuy (Ireland) · IMPACTOR4,276 reports
- ATTUNE 2544-01-006Depuy (Ireland) · IMPACTOR1 recall2,528 reports
- ATTUNE 2544-01-003Depuy (Ireland) · IMPACTOR1 recall1,491 reports
- ATTUNE 2544-01-017Depuy (Ireland) · IMPACTOR1 recall1,454 reports
- ATTUNE 2544-01-004Depuy (Ireland) · IMPACTOR1 recall873 reports
- DURALOC 2360-71-000Depuy Orthopaedics, Inc. · IMPACTOR715 reports
- SPECIALIST 2581-11-000Depuy (Ireland) · IMPACTOR703 reports
- INHANCE 6205-10-103Depuy (Ireland) · IMPACTOR334 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page