TRI-LOCK 2012-03-070Recalls & reported problems
Depuy (Ireland) · TRI-LOCK BPS W/GRIPTION BROACH SZ 7 · Class 1
All models of TRI-LOCK (21)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10603295082125), received between Mar 16, 2018 and Jan 5, 2023 — classified as 1 injury, 16 malfunctions.
The most frequently reported specific device problem is “Naturally Worn” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Naturally Worn7
- Device-Device Incompatibility3
- Scratched Material3
- Break2
- Material Deformation2
- Physical Resistance/Sticking1
Patient problems
- No Consequences Or Impact To Patient13
- No Clinical Signs, Symptoms or Conditions3
- Not Applicable1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
InjuryJan 5, 2023Break; Break
IT WAS REPORTED THAT THE SIZE 7 TRILOCK BROACH BROKE WHILE INSIDE PATIENTS FEMORAL CANAL REQUIRING ADDITIONAL OPERATIVE TIME, EXTENDED OSTEOTOMY AND CABLES TO RETRIEVE AND REPAIR FEMUR. THERE WAS NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16101462 ↗MalfunctionDec 17, 2021Device-Device Incompatibility; Device-Device Incompatibility
THE #7 TRI LOCK BROACH WOULD NOT ATTACH TO THE BROACH HANDLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13029348 ↗MalfunctionFeb 9, 2021Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE SIZE 7 TRI LOCK BROACH DID NOT ENGAGE TO THE BROACH HANDLE. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11301516 ↗MalfunctionJul 6, 2020Physical Resistance/Sticking; Physical Resistance/Sticking
SURGEON COMPLAINED THAT FEMORAL BROACHES WERE GETTING STUCK IN THE CANAL AND WAS CONCERNED THEY WERE OUT OF SPEC.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10235563 ↗MalfunctionSep 16, 2019Break; Break
TIP OF TRILOCK SIZE 7 BROACH BROKE LEAVING BROACH IN PATIENT WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? YES. IF YES, NUMBER OF MINUTES: 20 MINUTES. ACTION TAKEN WHEN EVENT OCCURRED? BURR TO ALLOW VICEGRIP ACCESS TO PROXIMAL BROACH. WAS PROCEDURE SUCCESSFULLY COMPLETED? YES.
Patient problems: Not Applicable
MAUDE report 9070939 ↗MalfunctionJul 18, 2018Device-Device Incompatibility; Scratched Material; Device-Device Incompatibility; Scratched Material
IT WAS REPORTED THAT THE TRI LOCK BROACH SZ.7 WOULD NOT SEAT ONTO THE BROACH HANDLE. UPON INSPECTION, NECK THAT ATTACHES TO BROACH IS SCARRED AND WOULD NOT ALLOW SEATING. NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7698252 ↗MalfunctionApr 30, 2018Naturally Worn; Naturally Worn
IT WAS REPORTED THAT DUE TO SEVERAL COMPLAINTS BY THE SURGEON, THE BROACHER FROM (B)(6) HOSPITAL (7 SETS) ARE BEING RETURNED FOR FURTHER TESTING. SIZES 0-7 FORM EACH SET WILL BE RETURNED. THE INITIAL COMPLAINTS WERE MADE REGARDING BROACH TO IMPLANT MISMATCH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7473591 ↗MalfunctionApr 30, 2018Naturally Worn; Naturally Worn
IT WAS REPORTED THAT DUE TO SEVERAL COMPLAINTS BY THE SURGEON, THE BROACHER FROM (B)(6) HOSPITAL ( 7 SETS) ARE BEING RETURNED FOR FURTHER TESTING. SIZES (B)(6) FORM ECH SET WILL BE RETURNED. THE INITIAL COMPLAINTS WERE MADE REGARDING BROACH TO IMPLANT MISMATCH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7473476 ↗MalfunctionApr 30, 2018Naturally Worn; Naturally Worn
IT WAS REPORTED THAT DUE TO SEVERAL COMPLAINTS BY THE SURGEON, THE BROACHER FROM (B)(6) HOSPITAL ( 7 SETS) ARE BEING RETURNED FOR FURTHER TESTING. SIZES 0-7 FORM EACH SET WILL BE RETURNED. THE INITIAL COMPLAINTS WERE MADE REGARDING BROACH TO IMPLANT MISMATCH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7473443 ↗MalfunctionApr 30, 2018Naturally Worn; Naturally Worn
IT WAS REPORTED THAT DUE TO SEVERAL COMPLAINTS BY THE SURGEON, THE BROACHER FROM THOMAS JEFFERSON UNIVE HOSPITAL ( 7 SETS) ARE BEING RETURNED FOR FURTHER TESTING. SIZES 0-7 FORM ECH SET WILL BE RETURNED. THE INITIAL COMPLAINTS WERE MADE REGARDING BROACH TO IMPLANT MISMATCH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7473162 ↗MalfunctionApr 30, 2018Naturally Worn; Naturally Worn
IT WAS REPORTED THAT DUE TO SEVERAL COMPLAINTS BY THE SURGEON, THE BROACHER FROM (B)(6) HOSPITAL ( 7 SETS) ARE BEING RETURNED FOR FURTHER TESTING. SIZES 0-7 FORM EACH SET WILL BE RETURNED. THE INITIAL COMPLAINTS WERE MADE REGARDING BROACH TO IMPLANT MISMATCH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7472527 ↗MalfunctionApr 30, 2018Naturally Worn; Naturally Worn
IT WAS REPORTED THAT DUE TO SEVERAL COMPLAINTS BY THE SURGEON, THE BROACHER FROM (B)(6) HOSPITAL ( 7 SETS) ARE BEING RETURNED FOR FURTHER TESTING. SIZES 0-7 FORM ECH SET WILL BE RETURNED. THE INITIAL COMPLAINTS WERE MADE REGARDING BROACH TO IMPLANT MISMATCH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7472320 ↗MalfunctionApr 25, 2018Naturally Worn; Naturally Worn
IT WAS REPORTED THAT DUE TO SEVERAL COMPLAINTS BY THE SURGEON, THE BROACHER FROM THOMAS JEFFERSON UNIVE HOSPITAL ( 7 SETS) ARE BEING RETURNED FOR FURTHER TESTING. SIZES 0-7 FORM ECH SET WILL BE RETURNED. THE INITIAL COMPLAINTS WERE MADE REGARDING BROACH TO IMPLANT MISMATCH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7457745 ↗MalfunctionMar 20, 2018Material Deformation; Material Deformation
IT WAS REPORTED THAT DUE TO THE ONGOING COMPLAINTS OF THE SURGEON REGARDING BROACH TO IMPLANT MISMATCH. ALL SIZE 0-7 BROACHES WILL BE REMOVED FROM THE TRILOCK SETS FOR FURTHER TESTING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7353323 ↗MalfunctionMar 20, 2018Material Deformation; Material Deformation
IT WAS REPORTED THAT DUE TO THE ONGOING COMPLAINTS OF THE SURGEON REGARDING BROACH TO IMPLANT MISMATCH. ALL SIZE 0-7 BROACHES WILL BE REMOVED FROM THE TRILOCK SETS FOR FURTHER TESTING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7353313 ↗MalfunctionMar 16, 2018Scratched Material; Scratched Material
IT WAS REPORTED THAT DUE TO THE ONGOING COMPLAINTS OF THE SURGEON REGARDING BROACH TO IMPLANT MISMATCH. ALL SIZE 0-7 BROACHES WILL BE REMOVED FROM THE TRILOCK SETS FOR FURTHER TESTING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7344441 ↗MalfunctionMar 16, 2018Scratched Material; Scratched Material
IT WAS REPORTED THAT DUE TO THE ONGOING COMPLAINTS OF THE SURGEON REGARDING BROACH TO IMPLANT MISMATCH. ALL SIZE 0-7 BROACHES WILL BE REMOVED FROM THE TRILOCK SETS FOR FURTHER TESTING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7344438 ↗
Other BROACH devices
- NA 2598-07-550Depuy (Ireland) · BROACH323 reports
- SUMMIT 2570-00-000Depuy (Ireland) · BROACH269 reports
- NA 952212500FDepuy (Ireland) · BROACH245 reports
- ACTIS 2010-01-050Depuy (Ireland) · BROACH235 reports
- CORAIL 9522-11-500Depuy (Ireland) · BROACH210 reports
- ACTIS 2010-01-060Depuy (Ireland) · BROACH186 reports
- ACTIS 2010-01-040Depuy (Ireland) · BROACH172 reports
- ACTIS 2010-01-070Depuy (Ireland) · BROACH162 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page