Skip to content

TRI-LOCK 2012-06-300Recalls & reported problems

Depuy (Ireland) · TRI-LOCK BPS W/GRIPTION CASE COMPLETE · Class 1

Quick summary

This DePuy Tri-Lock surgical instrument tray and broach pan set is used to hold and organize orthopedic surgical tools. The 27 reports mostly describe trays and cases peeling apart, cracking, chipping, fading and delaminating after years of repeated cleaning and sterilization. No deaths or injuries were reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
27
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 10603295082323), received between Oct 9, 2017 and Apr 1, 2024 — classified as 27 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Peeled/Delaminated11
  • Break9
  • Material Discolored9
  • Delamination5
  • Appropriate Device Problem Term/Code Not Available1
  • Contamination /Decontamination Problem1

Patient problems

  • No Consequences Or Impact To Patient22
  • No Clinical Signs, Symptoms or Conditions5
  • No Information5

Recent adverse event reports

The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 1, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE METAL AND PLASTIC IS BREAKING AND CHIPPING OFF OLD TRAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19011464 ↗
  • MalfunctionApr 1, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE METAL AND PLASTIC IS BREAKING AND CHIPPING OFF OLD TRAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19011410 ↗
  • MalfunctionAug 10, 2021Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE TRAYS WERE PEELING AND FALLING APART.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12302369 ↗
  • MalfunctionAug 10, 2021Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE TRAYS WERE PEELING AND FALLING APART.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12302368 ↗
  • MalfunctionJun 1, 2021Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE T-HANDLE HAS A CRACK IN IT AND NEEDS REPLACED. THIS DID NOT OCCUR DURING THE CASE, BUT DISCOVERED BEFOREHAND. ALSO, THE TRILOCK GRAPHIC CASE HAS PLASTIC PEELING OFF AND NEEDS REPLACED PER SPD GUIDELINES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11916441 ↗
  • MalfunctionJul 13, 2020Break; Material Discolored; Break; Material Discolored

    TRILOCK BROACH PAN HAVE FADED AND HAVE STARTED TO CHIP AFTER BEING IN ROTATION FOR APPROX. 10 YEARS AND THOUSANDS OF CASE REPROCESSING'S.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10267854 ↗
  • MalfunctionJul 13, 2020Break; Material Discolored; Break; Material Discolored

    TRILOCK BROACH PAN HAVE FADED AND HAVE STARTED TO CHIP AFTER BEING IN ROTATION FOR APPROX. 10 YEARS AND THOUSANDS OF CASE REPROCESSINGS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10267847 ↗
  • MalfunctionJul 13, 2020Break; Material Discolored; Break; Material Discolored

    TRILOCK BROACH PAN HAVE FADED AND HAVE STARTED TO CHIP AFTER BEING IN ROTATION FOR APPROX. 10 YEARS AND THOUSANDS OF CASE REPROCESSINGS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10267843 ↗
  • MalfunctionJul 13, 2020Break; Material Discolored; Break; Material Discolored

    TRILOCK BROACH PAN HAVE FADED AND HAVE STARTED TO CHIP AFTER BEING IN ROTATION FOR APPROX. 10 YEARS AND THOUSANDS OF CASE REPROCESSING'S.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10267840 ↗
  • MalfunctionJul 13, 2020Break; Material Discolored; Break; Material Discolored

    TRILOCK BROACH PAN HAVE FADED AND HAVE STARTED TO CHIP AFTER BEING IN ROTATION FOR APPROX. 10 YEARS AND THOUSANDS OF CASE REPROCESSINGS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10267824 ↗
  • MalfunctionJul 13, 2020Break; Material Discolored; Break; Material Discolored

    TRILOCK BROACH PAN HAVE FADED AND HAVE STARTED TO CHIP AFTER BEING IN ROTATION FOR APPROX. 10 YEARS AND THOUSANDS OF CASE REPROCESSINGS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10267821 ↗
  • MalfunctionJul 13, 2020Break; Material Discolored; Break; Material Discolored

    TRILOCK BROACH PAN HAVE FADED AND HAVE STARTED TO CHIP AFTER BEING IN ROTATION FOR APPROX. 10 YEARS AND THOUSANDS OF CASE REPROCESSING'S.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10267820 ↗
  • MalfunctionJul 13, 2020Break; Material Discolored; Break; Material Discolored

    TRILOCK BROACH PAN HAVE FADED AND HAVE STARTED TO CHIP AFTER BEING IN ROTATION FOR APPROX. 10 YEARS AND THOUSANDS OF CASE REPROCESSINGS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10267811 ↗
  • MalfunctionJul 13, 2020Break; Material Discolored; Break; Material Discolored

    TRILOCK BROACH PAN HAVE FADED AND HAVE STARTED TO CHIP AFTER BEING IN ROTATION FOR APPROX. 10 YEARS AND THOUSANDS OF CASE REPROCESSINGS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10267810 ↗
  • MalfunctionMay 8, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem

    CUSTOMER HAS BEEN USING THE TRILOCK TRAYS FOR ABOUT 3 MONTHS. NOW THIS ISSUE IS OCCURING: THE TRAYS HAVE BEEN WASHED ABOUT 7 TIMES UP TO NOW AND AFTER THE STERILIZATION THERE IS A KIND OF "CHALK DUST" ON THE TRAYS. THIS HAPPENS ONLY WHEN WASHING THE TRILOCK TRAYS, ALL OTHER TRAYS BEHAVE NORMALLY. THE CAUSE IS UNKNOWN BUT THE "CHALK-LIKE LAYER" IS LOCATED ON THE INSIDE AND OUTSIDE OF THE ENTIRE TRAY. THE FEAR IS THAT PARTICLES OF THIS LAYER ARE ON THE INSTRUMENTS AND CAN COME IN CONTACT WITH THE PATIENT. NO SURGERY DELAY, NO ADVERSE CONSEQUENCES FOR PATIENT REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8592234 ↗
  • MalfunctionDec 26, 2018Peeled/Delaminated; Peeled/Delaminated

    INSTRUMENT PEELING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8195831 ↗
  • MalfunctionDec 26, 2018Peeled/Delaminated; Peeled/Delaminated

    INSTRUMENT PEELING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8195812 ↗
  • MalfunctionDec 26, 2018Peeled/Delaminated; Peeled/Delaminated

    INSTRUMENT PEELING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8195806 ↗
  • MalfunctionDec 26, 2018Peeled/Delaminated; Peeled/Delaminated

    INSTRUMENT PEELING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8195788 ↗
  • MalfunctionDec 26, 2018Peeled/Delaminated; Peeled/Delaminated

    INSTRUMENT PEELING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8195783 ↗
  • MalfunctionDec 26, 2018Peeled/Delaminated; Peeled/Delaminated

    INSTRUMENT PEELING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8195765 ↗
  • MalfunctionAug 24, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE BLACK COATING PART OF THE STERILIZED CASE PEDESTAL PART WAS PEELED OFF AND THE METAL PART WAS EXPOSED DURING THE PREPARATION OF THE SURGERY ON (B)(6) 2018. THE HOSPITAL STAFF TRANSFERRED THE CONTENTS OF THE REPORTED PRODUCT TO THE ALTERNATIVE CASE TO STERILIZE. THERE WAS NO SURGICAL DELAY AND THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED BY THE HOSPITAL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7814426 ↗
  • MalfunctionOct 9, 2017Delamination; Delamination

    IT WAS REPORTED THAT THE RUBBER BROACH HOLDERS ARE STARTING TO LOSE PARTS. SEPTEMBER 15, 2017 - EVALUATION OF THE RETURNED CASES FOUND SOME OF THE BRACKETS ARE DELAMINATING. (B)(6).

    Patient problems: No Consequences Or Impact To Patient; No Information

    MAUDE report 6928172 ↗
  • MalfunctionOct 9, 2017Delamination; Delamination

    IT WAS REPORTED THAT THE RUBBER BROACH HOLDERS ARE STARTING TO LOSE PARTS. SEPTEMBER 15, 2017 - EVALUATION OF THE RETURNED CASES FOUND SOME OF THE BRACKETS ARE DELAMINATING. (B)(6).

    Patient problems: No Consequences Or Impact To Patient; No Information

    MAUDE report 6928163 ↗
  • MalfunctionOct 9, 2017Delamination; Delamination

    IT WAS REPORTED THAT THE RUBBER BROACH HOLDERS ARE STARTING TO LOSE PARTS. SEPTEMBER 15, 2017 - EVALUATION OF THE RETURNED CASES FOUND SOME OF THE BRACKETS ARE DELAMINATING. (B)(6).

    Patient problems: No Consequences Or Impact To Patient; No Information

    MAUDE report 6928154 ↗

Other TRAY, SURGICAL, INSTRUMENT devices

All TRAY, SURGICAL, INSTRUMENT devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.