SIGMA 2024-40-100Recalls & reported problems

Depuy (Ireland) · SIGMA HP PARTIAL KNEE TIBIAL CUTTING BLOCK LM/RL · Class 1

All models of SIGMA (380)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy SIGMA HP partial knee tibial cutting block is a surgical guide used to make precise bone cuts during knee replacement. The 43 reports mostly describe wear and tear on the instrument: bent or broken blocks and chisels, damaged cutting slots that a saw blade cannot pass through, and parts that no longer fit together. None of the reports mention deaths or injuries, and most note no effect on the patient. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
43
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 10603295082972), received between Aug 18, 2017 and Apr 1, 2026 — classified as 43 malfunctions.

The most frequently reported specific device problem is “Naturally Worn” (15 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Naturally Worn15
  • Device-Device Incompatibility11
  • Material Deformation10
  • Break5
  • Material Twisted/Bent4
  • Scratched Material4
  • Mechanical Jam3
  • Appropriate Device Problem Term/Code Not Available2

Patient problems

  • No Clinical Signs, Symptoms or Conditions24
  • No Consequences Or Impact To Patient14
  • Insufficient Information3
  • No Information3
  • No Code Available1

Recent adverse event reports

The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 1, 2026Break; Material Deformation

    ADDITIONAL INFORMATION RECEIVED: 1. DID IT BROKE INTO TWO OR MORE PIECES? IF NO, PLEASE SPECIFY WHAT IS THE ALLEGED DEFICIENCY OF THE INSTRUMENT. IS IT BENT, CRACKED, STRIPPED, WORN OR CROSS THREADED? - NO. THE PIN WAS CROSS THREADED IN THE PIN HOLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24756217 ↗
  • MalfunctionMay 14, 2025Device-Device Incompatibility; Material Deformation; Material Integrity Problem; Device-Device Incompatibility; Material Deformation; Material Integrity Problem

    ADDITIONAL INFORMATION RECEIVED: A. WAS THERE ANY SURGICAL DELAY RELATED TO THE EVENT? THIS DID NOT HAPPEN DURING THE PROCEDURE SO NOT APPLICABLE. B. PLEASE VERIFY IF THERE WAS ANY ADVERSE EVENT/PATIENT HARM INVOLVED? THIS DID NOT HAPPEN DURING THE PROCEDURE SO NOT APPLICABLE. C. WHAT DO YOU MEAN BY DAMAGED FOR THE PRODUCT CODES: 202440100 AND 202440200? 202440100 - (B)(4) / (10) (B)(6) - SIGMA HP UNI TIB CUT BLK LM/R / (10) (B)(6) - SIGMA HP UNI TIB CUT BLK LM/RL ¿ CUTTING SLOT DAMAGED, BLADE HAS HIT IT AND NOW UNABLE TO PUT SAW BLADE THROUGH HOLE EASILY ¿ IN ONE PIECE STILL 202440100 - (B)(4) / (10) (B)(6) - SIGMA HP UNI TIB CUT BLK LM/RL ¿ CUTTING SLOT DAMAGED, BLADE HAS HIT IT AND NOW UNABLE TO PUT SAW BLADE THROUGH HOLE EASILY ¿ IN ONE PIECE STILL 202440200 - (B)(4) / (10) (B)(6) - SIGMA HP UNI TIB CUT BLK RM/LL - CUTTING SLOT DAMAGED, BLADE HAS HIT IT AND NOW UNABLE TO PUT SAW BLADE THROUGH HOLE EASILY ¿ IN ONE PIECE STILL ALL ITEMS ABOVE HAVE A CUTTING SLOT ¿ WHERE THE SAW BLADE HAS REPEATEDLY HIT THIS DURING SURGERY IT HAD WARPED THE METAL CAUSING DAMAGE AND THE SAW BLADE NO LONGER CAN FIT THROUGH THE CUTTING SLOT. ALL ITEMS ARE STILL IN ONE PIECE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22018346 ↗
  • MalfunctionMay 14, 2025Device-Device Incompatibility; Material Deformation; Material Integrity Problem; Device-Device Incompatibility; Material Deformation; Material Integrity Problem

    IT WAS REPORTED THAT THE ITEMS WERE DAMAGED. ISSUES DID NOT OCCUR DURING SURGERY. SIGMA HP UNI TIB CUT BLK LM/RL (X2) ¿ CUTTING SLOT DAMAGED, BLADE HAS HIT IT AND NOW UNABLE TO PUT SAW BLADE THROUGH HOLE EASILY. SIGMA HP UNI TIB CUT BLK RM/LL - CUTTING SLOT DAMAGED, BLADE HAS HIT IT AND NOW UNABLE TO PUT SAW BLADE THROUGH HOLE EASILY. SIGMA HP UNI ANTERIOR CHISEL ¿ CHISEL END SNAPPED INTO TWO PIECES, PIECES BOTH DISPOSED OF. SIGMA HP UNI AP KEEL OSTEOTOME ¿ METAL KEEL END BENT, NO LONGER FIT FOR USE. SIGMA HP UNI FB KEEL OSTEOTOME - METAL KEEL END BENT, NO LONGER FIT FOR USE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22018341 ↗
  • MalfunctionDec 2, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE METAL BACK AND ALL POLY GOUGE BROKE. LM AND RM TIBIA CUT GUIDES ARE BENT.

    Patient problems: Insufficient Information

    MAUDE report 20824665 ↗
  • MalfunctionJun 26, 2024Material Deformation; Naturally Worn; Material Deformation; Naturally Worn

    ADDITIONAL INFORMATION RECEIVED STATES THAT THE ISSUE WAS THAT ITEMS WERE BENT AND WORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19620099 ↗
  • MalfunctionApr 3, 2024Naturally Worn; Naturally Worn

    ADDITIONAL INFORMATION WAS RECEIVED AND IT WAS STATED THAT THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19030824 ↗
  • MalfunctionNov 6, 2023Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT INSTRUMENTS NEED REPLACING DUE TO WEAR AND TEAR AND ARE NO LONGER USABLE FOR SURGERY. THERE WAS NO ADVERSE EFFECT TO THE PATIENT OR SURGERY

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18081019 ↗
  • MalfunctionJul 18, 2023Break; Material Deformation; Break; Material Deformation

    ADDITIONAL INFORMATION RECEIVED. A. WHAT PART OF THE INSTRUMENT BROKE? THE BIT AT THE TOP. B. DID IT BREAK INTO TWO OR MORE PIECES? IF NO, PLEASE SPECIFY, WHAT IS THE ALLEGED DEFICIENCY, IS IT BENT, CRACKED, STRIPPED, SCRATCHED, WORN, OR CROSS-THREADED? NO, IT'S BENT. C. PLEASE VERIFY, IF THE INSTRUMENT WAS USED, DURING SURGERY OR NOT? YES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17343633 ↗
  • MalfunctionMay 24, 2023Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT BROKE DURING KNEE SURGERY TODAY AT EPSOM HOSPITAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16991538 ↗
  • MalfunctionApr 28, 2023Scratched Material; Scratched Material

    IT WAS REPORTED THE SURFACES OF THE INSTRUMENTS ARE DEEPLY GOUGED AND SCRAPED FROM SAW USAGE. NO ADVERSE AFFECTS ON THE PATIENT. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16831613 ↗
  • MalfunctionFeb 27, 2023Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE WAS REPORTED FOR UNKNOWN REASON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16443291 ↗
  • MalfunctionJan 9, 2023Contamination /Decontamination Problem; Scratched Material; Contamination /Decontamination Problem; Scratched Material

    IT WAS REPORTED THAT THE INSTRUMENT WAS DAMAGED AND COULD NOT BE PROPERLY CLEANED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16121601 ↗
  • MalfunctionOct 28, 2022Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn

    IT WAS REPORTED THE LM CUTTING BLOCK ON SIGMA PARTIAL SET IS DAMAGED. WONT ACCEPT PINS DUE TO WEAR AND TEAR. NO SURGICAL DELAY. NO ADVERSE AFFECTS ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15693804 ↗
  • MalfunctionJul 26, 2022Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT BOTH OF THE TIBIAL CUTTING BLOCKS ARE SEVERELY DAMAGED THROUGH WEAR AND TEAR. NO PATIENT HARM OR SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15103196 ↗
  • MalfunctionFeb 4, 2022Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THESE BLOCKS ARE VERY WORN DUE TO HIGH USAGE. THE SAW SPACE IS DEFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13451376 ↗
  • MalfunctionFeb 4, 2022Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THESE BLOCKS ARE VERY WORN DUE TO HIGH USAGE. THE SAW SPACE IS DEFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13451374 ↗
  • MalfunctionFeb 4, 2022Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THESE BLOCKS ARE VERY WORN DUE TO HIGH USAGE. THE SAW SPACE IS DEFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13451373 ↗
  • MalfunctionSep 22, 2021Naturally Worn; Naturally Worn

    WE HAVE A HANDLE THAT¿S REQUIRED ON A SET AS IT ONLY HAS ONE AND SHOULD HAVE TWO. THE OTHERS ARE DUE TO WEAR ON TEAR ON THE CURRENT ONES SO THEY NEED REPLACING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12517199 ↗
  • MalfunctionSep 22, 2021Naturally Worn; Naturally Worn

    WE HAVE A HANDLE THAT¿S REQUIRED ON A SET AS IT ONLY HAS ONE AND SHOULD HAVE TWO.THE OTHERS ARE DUE TO WEAR ON TEAR ON THE CURRENT ONES SO THEY NEED REPLACING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12517196 ↗
  • MalfunctionSep 14, 2021Naturally Worn; Naturally Worn

    INSTRUMENTS ARE BEING REPORTED DUE TO WEAR AND TEAR ON THE CURRENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12468458 ↗
  • MalfunctionSep 14, 2021Naturally Worn; Naturally Worn

    INSTRUMENTS ARE BEING REPORTED DUE TO WEAR AND TEAR ON THE CURRENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12468450 ↗
  • MalfunctionJul 1, 2021Material Twisted/Bent; Naturally Worn; Material Twisted/Bent; Naturally Worn

    IT WAS REPORTED THAT SSD NOTICED THE CUTTING BLOCKS WERE VERY WORN AND NO LONGER FUNCTIONAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12105243 ↗
  • MalfunctionJul 1, 2021Material Twisted/Bent; Naturally Worn; Material Twisted/Bent; Naturally Worn

    IT WAS REPORTED THAT SSD NOTICED THE CUTTING BLOCKS WERE VERY WORN AND NO LONGER FUNCTIONAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12105242 ↗
  • MalfunctionMar 31, 2021Device-Device Incompatibility; Device-Device Incompatibility

    SURGEON JAMMED A DRILL PIN IN A SIGMA PARTIAL CUTTING BLOCK WHEN REMOVING THE BLOCK AFTER CUTTING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11598710 ↗
  • MalfunctionNov 24, 2020Scratched Material; Scratched Material

    IT WAS REPORTED THAT DURING POST CLEANING INSPECTION THESE CUT BLOCKS WERE FOUND TO HAD DEEP GROVES CUT INTO THE TOPS OF THE BLOCKS. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10894848 ↗

Other INSTRUMENT, CUTTING, ORTHOPEDIC devices

All INSTRUMENT, CUTTING, ORTHOPEDIC devices →