ATTUNE 2545-00-155Recalls & reported problems

Depuy (Ireland) · ATTUNE A-P CHAMFER BLOCK SIZE 5 · Class 1

All models of ATTUNE (1475)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Depuy ATTUNE A-P chamfer block is an orthopedic cutting instrument. Reports mostly describe parts deforming or failing to cut, with some breaking or corroding; patient impact was rarely noted. There are 40 reports, no deaths and no injuries, and no recalls name this device. Reports rose to 13 in the last 12 months from 6 the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
39
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 39 adverse event reports that list this exact device (DI 10603295395287), received between Sep 25, 2019 and Feb 12, 2026 — classified as 39 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (22 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 6 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 12 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation22
  • Failure to Cut12
  • Naturally Worn4
  • Break3
  • Corroded3
  • Device-Device Incompatibility2
  • Incorrect Measurement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions26
  • No Consequences Or Impact To Patient11
  • Insufficient Information2
  • No Code Available2

Recent adverse event reports

The 25 most recent of 39 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 12, 2026Failure to Cut

    IT WAS REPORTED THAT THEIR CUT BLOCKS WERE 10+ YEARS OLD. NEED TO REPLACE THEM AS THEY HAVE HAD 2+ CASES THAT CUTS HAD SEEMED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24320518 ↗
  • MalfunctionFeb 12, 2026Failure to Cut

    IT WAS REPORTED THAT THEIR CUT BLOCKS WERE 10+ YEARS OLD. NEED TO REPLACE THEM AS THEY HAVE HAD 2+ CASES THAT CUTS HAD SEEMED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24320514 ↗
  • MalfunctionFeb 12, 2026Failure to Cut

    IT WAS REPORTED THAT THEIR CUT BLOCKS WERE 10+ YEARS OLD. NEED TO REPLACE THEM AS THEY HAVE HAD 2+ CASES THAT CUTS HAD SEEMED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24320443 ↗
  • MalfunctionFeb 12, 2026Failure to Cut

    IT WAS REPORTED THAT THEIR CUT BLOCKS WERE 10+ YEARS OLD. NEED TO REPLACE THEM AS THEY HAVE HAD 2+ CASES THAT CUTS HAD SEEMED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24320441 ↗
  • MalfunctionFeb 12, 2026Failure to Cut

    IT WAS REPORTED THAT THEIR CUT BLOCKS WERE 10+ YEARS OLD. NEED TO REPLACE THEM AS THEY HAVE HAD 2+ CASES THAT CUTS HAD SEEMED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24320436 ↗
  • MalfunctionFeb 12, 2026Failure to Cut

    IT WAS REPORTED THAT THEIR CUT BLOCKS WERE 10+ YEARS OLD. NEED TO REPLACE THEM AS THEY HAVE HAD 2+ CASES THAT CUTS HAD SEEMED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24320434 ↗
  • MalfunctionFeb 12, 2026Failure to Cut

    IT WAS REPORTED THAT THEIR CUT BLOCKS WERE 10+ YEARS OLD. NEED TO REPLACE THEM AS THEY HAVE HAD 2+ CASES THAT CUTS HAVE SEEMED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24320388 ↗
  • MalfunctionFeb 12, 2026Failure to Cut

    IT WAS REPORTED THAT THEIR CUT BLOCKS WERE 10+ YEARS OLD. NEED TO REPLACE THEM AS THEY HAVE HAD 2+ CASES THAT CUTS HAVE SEEMED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24320387 ↗
  • MalfunctionFeb 12, 2026Failure to Cut

    IT WAS REPORTED THAT THEIR CUT BLOCKS WERE 10+ YEARS OLD. NEED TO REPLACE THEM AS THEY HAVE HAD 2+ CASES THAT CUTS HAVE SEEMED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24320385 ↗
  • MalfunctionFeb 12, 2026Failure to Cut

    IT WAS REPORTED THAT THEIR CUT BLOCKS WERE 10+ YEARS OLD. NEED TO REPLACE THEM AS THEY HAVE HAD 2+ CASES THAT CUTS HAVE SEEMED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24320383 ↗
  • MalfunctionFeb 12, 2026Failure to Cut

    IT WAS REPORTED THAT THEIR CUT BLOCKS WERE 10+ YEARS OLD. NEED TO REPLACE THEM AS THEY HAVE HAD 2+ CASES THAT CUTS HAVE SEEMED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24320381 ↗
  • MalfunctionFeb 12, 2026Failure to Cut

    IT WAS REPORTED THAT THEIR CUT BLOCKS WERE 10+ YEARS OLD. NEED TO REPLACE THEM AS THEY HAVE HAD 2+ CASES THAT CUTS HAVE SEEMED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24320377 ↗
  • MalfunctionSep 8, 2025Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT THE PIN HOLES WERE STRIPPED. METAL SHAVINGS BEING PRODUCED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22992410 ↗
  • MalfunctionJun 27, 2025Break; Break

    IT WAS REPORTED THAT INSTRUMENTS WERE BROKEN. ATTUNE EM TIBIAL ANKLE CLAMP BLACK PART BROKEN. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22350292 ↗
  • MalfunctionMay 19, 2025Break; Break

    IT WAS REPORTED THAT THE DEVICE IS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22045872 ↗
  • MalfunctionApr 21, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT A SIZE 5 AP CHAMFER CUT BLOCK WAS DAMAGED AND NEEDED REPLACING. THE DAMAGE IS AT THE CORNERS OF THE CUT SLOTS THE BLOCK HAS DEVELOPED SHARP EDGES WHICH ARE A RISK TO SURGEON STERILITY DURING THE CASE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21872576 ↗
  • MalfunctionMar 11, 2025Incorrect Measurement; Device-Device Incompatibility; Incorrect Measurement; Device-Device Incompatibility

    IT WAS REPORTED THAT THERE WAS A VERY APPARENT DISCREPANCY BETWEEN THE SIZE 5 CUTTING BLOCK AND THE SIZE 5 TRIAL. STAFF CUT WITH THE SIZE 5 FOUR-IN-ONE BLOCK AND THE SIZE 5 TRIAL WOULD NOT FIT. AFTER CLEANING IT UP WITH FREEHAND CUTS, STAFF WERE ABLE TO GET THE TRIAL TO FIT. BECAUSE THEY WERE PERFORMING A BILATERAL TOTAL KNEE REPLACEMENT, THEY USED THE SAME BLOCK ON THE CONTRA LATERAL SIDE WITH THE SAME RESULT, ALLOWING THEM TO RULE OUT THE POSSIBILITY OF A WARPED/FAULTY FEMORAL TRIAL.

    Patient problems: Insufficient Information

    MAUDE report 21574974 ↗
  • MalfunctionFeb 7, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE CUTTING BLOCK HAS JAGGED EDGES ON THE BACK OF IT. THE PATELLA CLAMP WILL NOT SECURELY ACCEPT ATTACHMENTS. SURGICAL DELAY WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED FOR REPORTING. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 21337771 ↗
  • MalfunctionApr 12, 2024Material Deformation; Material Deformation

    AT A GENERAL CHECK OF THE ATTUNE AP CUTTING BLOCKS THE CENTRAL STERILIZATION DEPARTMENT DETECTED SUBSTANTIAL SHARP EDGES AT THE CORNERS OF THE CUTTING SLOTS. THERE WERE NO IMPLICATIONS DURING ANY SURGERY! NO FURTHER INFORMATION AVAILABLE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19096100 ↗
  • MalfunctionApr 12, 2024Material Deformation; Material Deformation

    AT A GENERAL CHECK OF THE ATTUNE AP CUTTING BLOCKS THE CENTRAL STERILIZATION DEPARTMENT DETECTED SUBSTANTIAL SHARP EDGES AT THE CORNERS OF THE CUTTING SLOTS. THERE WERE NO IMPLICATIONS DURING ANY SURGERY! NO FURTHER INFORRMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19096099 ↗
  • MalfunctionApr 12, 2024Material Deformation; Material Deformation

    AT A GENERAL CHECK OF THE ATTUNE AP CUTTING BLOCKS THE CENTRAL STERILIZATION DEPARTMENT DETECTED SUBSTANTIAL SHARP EDGES AT THE CORNERS OF THE CUTTING SLOTS. THERE WERE NO IMPLICATIONS DURING ANY SURGERY! NO FURTHER INFORRMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19096095 ↗
  • MalfunctionApr 12, 2024Material Deformation; Material Deformation

    AT A GENERAL CHECK OF THE ATTUNE AP CUTTING BLOCKS THE CENTRAL STERILIZATION DEPARTMENT DETECTED SUBSTANTIAL SHARP EDGES AT THE CORNERS OF THE CUTTING SLOTS. THERE WERE NO IMPLICATIONS DURING ANY SURGERY! NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19096094 ↗
  • MalfunctionApr 12, 2024Material Deformation; Material Deformation

    AT A GENERAL CHECK OF THE ATTUNE AP CUTTING BLOCKS THE CENTRAL STERILIZATION DEPARTMENT DETECTED SUBSTANTIAL SHARP EDGES AT THE CORNERS OF THE CUTTING SLOTS. THERE WERE NO IMPLICATIONS DURING ANY SURGERY! NO FURTHER INFORRMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19096088 ↗
  • MalfunctionApr 12, 2024Material Deformation; Material Deformation

    AT A GENERAL CHECK OF THE ATTUNE AP CUTTING BLOCKS THE CENTRAL STERILIZATION DEPARTMENT DETECTED SUBSTANTIAL SHARP EDGES AT THE CORNERS OF THE CUTTING SLOTS. THERE WERE NO IMPLICATIONS DURING ANY SURGERY. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19096067 ↗
  • MalfunctionMar 25, 2024Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE CHAMFER BLOCK IS DEFORMED. CANNOT GET SAW THROUGH. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? NO. WAS PROCEDURE SUCCESSFULLY COMPLETED? YES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18967591 ↗

Other INSTRUMENT, CUTTING, ORTHOPEDIC devices

All INSTRUMENT, CUTTING, ORTHOPEDIC devices →