Skip to content

SIGMA 2024-46-010Recalls & reported problems

Depuy (Ireland) · SIGMA HP PARTIAL KNEE ALL POLY TIB KEEL OSTEOTOME · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
17
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10603295083160), received between Aug 17, 2017 and Aug 7, 2025 — classified as 17 malfunctions.

The most frequently reported specific device problem is “Break” (15 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break15
  • Appropriate Device Problem Term/Code Not Available1
  • Material Deformation1

Patient problems

  • No Consequences Or Impact To Patient8
  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information2
  • No Information2

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2025Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE RED HANDLED KEEL OSTEOTOME WAS BROKEN. VERSION STICK BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22737925 ↗
  • MalfunctionMay 14, 2025Material Deformation; Material Deformation

    ADDITIONAL INFORMATION RECEIVED: A. WAS THERE ANY SURGICAL DELAY RELATED TO THE EVENT? THIS DID NOT HAPPEN DURING THE PROCEDURE SO NOT APPLICABLE. B. PLEASE VERIFY IF THERE WAS ANY ADVERSE EVENT/PATIENT HARM INVOLVED? THIS DID NOT HAPPEN DURING THE PROCEDURE SO NOT APPLICABLE. C. WHAT DO YOU MEAN BY DAMAGED FOR THE PRODUCT CODES: 202440100 AND 202440200? 202440100 - (B)(4) / (B)(6) - SIGMA HP UNI TIB CUT BLK LM/RL ¿ CUTTING SLOT DAMAGED, BLADE HAS HIT IT AND NOW UNABLE TO PUT SAW BLADE THROUGH HOLE EASILY ¿ IN ONE PIECE STILL 202440100 - (B)(4) / (B)(6) - SIGMA HP UNI TIB CUT BLK LM/RL ¿ CUTTING SLOT DAMAGED, BLADE HAS HIT IT AND NOW UNABLE TO PUT SAW BLADE THROUGH HOLE EASILY ¿ IN ONE PIECE STILL 202440200 - (B)(4) / (B)(6) - SIGMA HP UNI TIB CUT BLK RM/LL - CUTTING SLOT DAMAGED, BLADE HAS HIT IT AND NOW UNABLE TO PUT SAW BLADE THROUGH HOLE EASILY ¿ IN ONE PIECE STILL ALL ITEMS ABOVE HAVE A CUTTING SLOT ¿ WHERE THE SAW BLADE HAS REPEATEDLY HIT THIS DURING SURGERY IT HAD WARPED THE METAL CAUSING DAMAGE AND THE SAW BLADE NO LONGER CAN FIT THROUGH THE CUTTING SLOT. ALL ITEMS ARE STILL IN ONE PIECE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22018349 ↗
  • MalfunctionDec 2, 2024Break; Break

    IT WAS REPORTED THAT THE METAL BACK AND ALL POLY GOUGE BROKE. LM AND RM TIBIA CUT GUIDES ARE BENT.

    Patient problems: Insufficient Information

    MAUDE report 20824741 ↗
  • MalfunctionFeb 26, 2024Break; Break

    SURGEON WAS PREPPING THE TIBIAL WITH THE FB KEEL AND AS HE SAID THE BONE WAS VERY HARD THAT THE INSTRUMENT SNAPPED AT THE TIP (WHERE THE KEEL IS). NO PATIENT HARM, NO DELAYS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18778540 ↗
  • MalfunctionOct 3, 2023Break; Break

    IT WAS REPORTED THAT THE T5 RM/LL AND LM/RL FEMUR IMPLANTS WERE MISSING. THE TIBIAL OSTEOTOME IS USED DURING THE PREPARATION OF THE TIBIAL SIGMA BASE, I.E. AT THE END OF THE OPERATION. TIBIAL OSTEOPOROSIS ALL POLY BROKE DURING THE INTERVENTION. THE INTERVENTION WAS ABLE TO TAKE PLACE NORMALLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17860142 ↗
  • MalfunctionJun 22, 2023Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT LOCKING CAP WAS BROKEN OFF ONE END.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17183888 ↗
  • MalfunctionOct 4, 2022Break; Break

    IT WAS REPORTED THAT THE END CAP HAS BROKEN OFF KEEL OSTEOTOMES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15536634 ↗
  • MalfunctionFeb 27, 2020Break; Break

    IMPACTING HEAD OF THE 202446010 OSTEOTOME HAS BROKEN OFF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9763765 ↗
  • MalfunctionJan 19, 2020Break; Break

    IT WAS REPORTED THAT BOTH GOUGES BROKE AT THE BACK OF THE HANDLE. BOTH WERE BROKEN DURING STERILIZATION, NOT DURING SURGERY. NEEDS REPLACEMENT BOTH P1 TO DUKE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9604991 ↗
  • MalfunctionJan 9, 2019Break; Break

    IT WAS REPORTED THAT ALL POLYOSTEOTOME WAS FOUND BROKEN UPON CLEANING UP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8230959 ↗
  • MalfunctionJan 7, 2019Break; Break

    IT WAS REPORTED THAT AS THE SURGEON WAS USING THE ALL POLY GAUGE FOR THE TIBIA, THE SILVER GOUGE PIECE BROKE OFF INTO ONE PIECE. THE PIECE WAS RETRIEVED FROM THE PATIENT, BUT WAS THROWN OUT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8221931 ↗
  • MalfunctionOct 24, 2018Break; Break

    END CAP HAS BROKEN OFF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8000213 ↗
  • MalfunctionAug 13, 2018Break; Break

    IMPACTING HEAD OF OSTEOTOME HAS BROKEN OFF. IMPACTING HEAD WAS NOT RETURNED. THIS CASE HAS BEEN REPORTED BY THE LOAN KIT TECHNICIAN DURING LOAN KIT INSPECTION. IT HAS BEEN FORWARDED TO DEPUY ENGINEERING FOR ASSESSMENT. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7776819 ↗
  • MalfunctionJul 2, 2018Break; Break

    IT WAS REPORTED THAT THE TIP BROKE OFF OF THE ALL POLY KEEL OSTEOTOME.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7654267 ↗
  • MalfunctionJun 6, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE KEEL OSTEOTOME HAD BEEN LEFT/CLOGGED IN THE LOOPED SECTION OF THE DEVICE AFTER CLEANING/STERILIZATION WHICH RESULTED IN A NEW KIT HAVING TO BE OPENED AND A REPLACEMENT DEVICE HAVING TO BE SUBSTITUTED IN. WAS UNAWARE OF WHAT THE INSTRUMENT WAS OR LOOKED LIKE THE REP SHOWED ME AN IMAGE FROM A SURGICAL TECHNIQUE AND POINT TO THE AREA IN WHICH SHE HAS SEEN BONE COLLECTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7573544 ↗
  • MalfunctionNov 17, 2017Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT HAS BROKEN INTO THREE PIECES.

    Patient problems: No Information

    MAUDE report 7041607 ↗
  • MalfunctionAug 17, 2017Break; Break

    IT WAS REPORTED THAT THE GOUGE TIP BROKE OFF.

    Patient problems: No Information

    MAUDE report 6800376 ↗