SELF CENTERING 2055-13-000Recalls & reported problems
Depuy (Ireland) · SELF CENTERING MODULAR CATHCART HANDLE ANGLED · Class 1
All models of SELF CENTERING (29)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy self-centering Cathcart handle, a surgical instrument, has 323 reports, including 2 injuries and no deaths. Many describe contamination, with hospitals finding germs on the handles, along with trial parts that stick, jam or will not fit. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 322
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 322 adverse event reports that list this exact device (DI 10603295084488), received between Sep 15, 2017 and Sep 21, 2026 — classified as 2 injuries, 320 malfunctions.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (132 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Contamination /Decontamination Problem132
- Device-Device Incompatibility120
- Mechanical Jam42
- Break21
- Device Contamination with Chemical or Other Material13
- Physical Resistance/Sticking12
- Naturally Worn6
- Corroded5
Patient problems
- No Clinical Signs, Symptoms or Conditions211
- No Consequences Or Impact To Patient86
- Insufficient Information21
- Not Applicable6
- No Code Available2
- No Information2
Recent adverse event reports
The 25 most recent of 322 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 21, 2026Device-Device Incompatibility
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50845881 ↗MalfunctionMay 20, 2026Device-Device Incompatibility
IT WAS REPORTED THAT THE BIPOLAR HEAD TRIAL WON'T COME OFF ANGLED INSERTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25230819 ↗MalfunctionMay 18, 2026Corroded; Contamination /Decontamination Problem
IT WAS REPORTED THAT DURING WASHING OF THE SELF-CENTERING MODULAR CATHCART HANDLE ANGLED DEVICE, BROWN FLUID LEAKED FROM THE END. WATER HAD LIKELY ENTERED UNDER THE BALL BEARINGS AND CAUSED RUST. NO DELAYS OCCURRED DURING THE HIP SURGERY. NO PATIENT INJURIES OCCURRED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25204762 ↗MalfunctionMay 14, 2026Device-Device Incompatibility
IT WAS REPORTED THAT DURING A HIP PROCEDURE, ONE OF THE BALL BEARINGS WAS STUCK AND THE TRIAL DOES NOT GO ON THE HANDLE. NO DELAYS DURING DURING THE CASE AND NO PATIENT INJURIES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25176604 ↗MalfunctionMay 6, 2026Contamination /Decontamination Problem
IT WAS REPORTED THAT THE HOSPITAL FOUND BIOBURDEN IN ALL OF THEIR BIPOLAR ENDO HEAD TRIAL HANDLES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25092874 ↗MalfunctionMay 6, 2026Contamination /Decontamination Problem
IT WAS REPORTED THAT THE HOSPITAL FOUND BIOBURDEN IN ALL OF THEIR BIPOLAR ENDO HEAD TRIAL HANDLES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25092813 ↗MalfunctionMay 6, 2026Contamination /Decontamination Problem
IT WAS REPORTED THAT THE HOSPITAL FOUND BIOBURDEN IN ALL OF THEIR BIPOLAR ENDO HEAD TRIAL HANDLES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25092797 ↗MalfunctionMay 6, 2026Contamination /Decontamination Problem
IT WAS REPORTED THAT THE HOSPITAL FOUND BIOBURDEN IN ALL OF THEIR BIPOLAR ENDO HEAD TRIAL HANDLES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25092767 ↗MalfunctionApr 2, 2026Contamination /Decontamination Problem
IT WAS REPORTED THAT THE SPRING MECHANISM IN THE DEVICES HOLDS BIO BURDEN, RESULTING IN THE FACILITY'S DECISION TO DISCONTINUE STERILIZATION BY SPD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24769127 ↗MalfunctionMar 19, 2026Contamination /Decontamination Problem
IT WAS REPORTED THAT THE HANDLE CAN NO LONGER BE CLEANED. BLOOD GETTING CAUGHT BEHIND THE BEARINGS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24640746 ↗MalfunctionMar 3, 2026Device-Device Incompatibility
IT WAS REPORTED THAT THEY WERE DOING A HEMI. THE CURVED WHITE HANDLE THAT ATTACHES TO THE HEMI TRIALS WAS TOUGH TO GET ON AND OFF. THIS DID NOT SLOW THE SURGERY DOWN AND THE SURGERY PROCEEDED AS NORMAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24494668 ↗MalfunctionFeb 23, 2026Contamination /Decontamination Problem
IT WAS REPORTED BEHIND THE INSTRUMENTS` SPRING BALLS BLOOD, FLUID, TISSUE, ETC. IS STILL APPARENT AFTER REPROCESSING. INSTRUMENT IS NOT POSSIBLE TO BE REPROCESSED PROPERLY. NO ADVERSE PATIENT IMPACT, NO SURGICAL DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24421026 ↗MalfunctionFeb 20, 2026Device-Device Incompatibility
IT WAS REPORTED THAT S/C TRIAL HANDLE AND S/C TRIAL HANDLE ANGLED WERE DUE TO AN INABILITY TO DISASSEMBLE THE INSTRUMENTS, WHICH PREVENTED CLEANING BEHIND THE SPRING BALLS. THE EVENT WAS IDENTIFIED IN THE CENTRAL STERILE SERVICES DEPARTMENT, AND THE CUSTOMER DECLINED TO REPROCESS THE INSTRUMENTS. THERE WAS NO ADVERSE PATIENT IMPACT OR SURGICAL DELAY ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24400196 ↗MalfunctionFeb 19, 2026Contamination /Decontamination Problem
IT WAS REPORTED THAT AFTER THE SURGICAL PROCEDURE, BEHIND THE INSTRUMENTS SPRING BALLS BLOOD, FLUID, TISSUE, WERE STILL APPARENT AFTER REPROCESSING. THE INSTRUMENT WAS NOT POSSIBLE TO BE REPROCESSED PROPERLY. NO ADVERSE PATIENT IMPACT, NO SURGICAL DELAY REPORTED. S/C TRIAL HANDLE ANGLED: 205513000: NOT APPLICABLE. LIST ANY J&J PRODUCTS THAT WERE UTILIZED DURING THE CASE BUT DID NOT CONTRIBUTE TO THE REPORTED EVENT. WAS THERE ANY REPORTED PATIENT OR USER HARM? NO. WAS THERE A SIGNIFICANT DELAY? NO.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24393169 ↗MalfunctionFeb 18, 2026Corroded; Contamination /Decontamination Problem
IT WAS REPORTED THAT DURING STERILE PROCESSING IT WAS OBSERVED THAT THE HANDLE WAS HOLDING BIO BURDEN AND RUST IN THE ATTACHMENT PIECE FOR THE HEAD TRIALS AND COULD NOT BE CLEANED PROPERLY. THE EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24384657 ↗MalfunctionFeb 13, 2026Contamination /Decontamination Problem
IT WAS REPORTED THAT IT DID NOT HAPPEN IN SURGERY, SPD FOUND DEBRIS COMING OUT OF THE LOCKING MECHANISM AND CANNOT BE CLEANED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24333262 ↗MalfunctionFeb 12, 2026Device-Device Incompatibility; No Apparent Adverse Event
IT WAS REPORTED THAT THE INSTRUMENT WAS NOT HOLDING PROPERLY DUE TO WEAR. IT WAS REPORTED THAT THERE WAS NO DELAY IN THE PROCEDURE DUE TO THE EVENT. IT WAS REPORTED THAT THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24321691 ↗MalfunctionFeb 5, 2026Device-Device Incompatibility
IT WAS REPORTED THAT IT IS VERY DIFFICULT TO GET THE MOD ENDO HEAD TRIALS ON AND OFF THE WHITE HANDLE, AS THEY TEND TO GET STUCK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24270603 ↗MalfunctionJan 13, 2026Device-Device Incompatibility
IT WAS REPORTED THAT THE HANDLES DID NOT CONNECT PROPERLY TO THE TRIALS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24053799 ↗MalfunctionJan 2, 2026Device-Device Incompatibility; No Apparent Adverse Event
IT WAS REPORTED THAT THE BALL-BEARINGS ON THE UNIPOLAR/BIPOLAR ANGLED HANDLE WAS WORN. TROUBLE ATTACHING HEAD TRIALS ON TO HANDLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23949572 ↗MalfunctionJan 2, 2026Device-Device Incompatibility; No Apparent Adverse Event
IT WAS REPORTED THAT THE BALL-BEARINGS ON THE UNIPOLAR/BIPOLAR ANGLED HANDLE WAS WORN. TROUBLE ATTACHING HEAD TRIALS ON TO HANDLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23949398 ↗MalfunctionDec 22, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT STERILE PROCESSING DEPARTMENT COULD NOT CLEAN THESE INSTRUMENTS DUE TO DESIGN OF BALL BEARING. THE FAILURE WAS NOT DISCOVERED DURING SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23872608 ↗MalfunctionDec 22, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT STERILE PROCESSING DEPARTMENT COULD NOT CLEAN THESE INSTRUMENTS DUE TO DESIGN OF BALL BEARING. THE FAILURE WAS NOT DISCOVERED DURING SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23872592 ↗MalfunctionDec 22, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING THE CASE THE TECH WENT TO PUT THE BIPOLAR HEAD TRIAL ONTO THE INSERTER HANDLE AND IT WOULD NOT SEAT PROPERLY. THE NURSE GRABBED THE SECOND SET THAT WAS AVAILABLE AND THAT INSERTER HANDLE WAS USED. SINCE THERE WAS ANOTHER ONE AVAILABLE THERE WAS NO DELAY TO THE CASE, NOR WAS THERE ANY NEGATIVE EFFECT TO THE PATIENT ASSOCIATED WITH THIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23864715 ↗MalfunctionNov 17, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT HANDLE DOES NOT CONNECT PROPERLY TO TRIAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23570266 ↗
FDA clearance and classification
The FDA regulates product code HWA (impactor) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the SELF CENTERING 2055-13-000
Has the SELF CENTERING 2055-13-000 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10603295084488), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the SELF CENTERING 2055-13-000?
The FDA's MAUDE database lists 322 adverse event reports for this device, received between Sep 15, 2017 and Sep 21, 2026. In those reports, the most common specific device problems are “Contamination /Decontamination Problem” (132), “Device-Device Incompatibility” (120) and “Mechanical Jam” (42); the most common patient problem is “Not Applicable” (6). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SELF CENTERING 2055-13-000?
None of the reports are classified as deaths, 2 as injuries and 320 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the SELF CENTERING 2055-13-000 FDA approved?
Devices of this type (product code HWA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).
Who makes the SELF CENTERING 2055-13-000, and is it still sold?
The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand SELF CENTERING, model 2055-13-000). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other IMPACTOR devices
- PINNACLE 2217-50-041Depuy (Ireland) · IMPACTOR4,276 reports
- ATTUNE 2544-01-006Depuy (Ireland) · IMPACTOR1 recall2,528 reports
- ATTUNE 2544-01-003Depuy (Ireland) · IMPACTOR1 recall1,491 reports
- ATTUNE 2544-01-017Depuy (Ireland) · IMPACTOR1 recall1,454 reports
- ATTUNE 2544-01-004Depuy (Ireland) · IMPACTOR1 recall873 reports
- DURALOC 2360-71-000Depuy Orthopaedics, Inc. · IMPACTOR715 reports
- SPECIALIST 2581-11-000Depuy (Ireland) · IMPACTOR703 reports
- INHANCE 6205-10-103Depuy (Ireland) · IMPACTOR334 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page