SELF CENTERING 2055-13-000Recalls & reported problems

Depuy (Ireland) · SELF CENTERING MODULAR CATHCART HANDLE ANGLED · Class 1

Quick summary

This DePuy self-centering Cathcart handle, a surgical instrument, has 323 reports, including 2 injuries and no deaths. Many describe contamination, with hospitals finding germs on the handles, along with trial parts that stick, jam or will not fit. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
324
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 324 adverse event reports that list this exact device (DI 10603295084488), received between Sep 15, 2017 and Sep 21, 2026 — classified as 2 injuries, 322 malfunctions.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (132 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 2 reportsFeb 2024: 3 reportsMar 2024: 2 reportsApr 2024: 3 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 2 reportsSep 2024: 3 reportsOct 2024: 5 reportsNov 2024: 4 reportsDec 2024: 3 reportsJan 2025: 4 reportsFeb 2025: 7 reportsMar 2025: 1 reportsApr 2025: 4 reportsMay 2025: 2 reportsJun 2025: 4 reportsJul 2025: 3 reportsAug 2025: 5 reportsSep 2025: 0 reportsOct 2025: 9 reportsNov 2025: 3 reportsDec 2025: 3 reportsJan 2026: 3 reportsFeb 2026: 7 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 7 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination /Decontamination Problem132
  • Device-Device Incompatibility122
  • Mechanical Jam44
  • Break21
  • Device Contamination with Chemical or Other Material13
  • Physical Resistance/Sticking12
  • Naturally Worn6
  • Corroded5

Patient problems

  • No Clinical Signs, Symptoms or Conditions213
  • No Consequences Or Impact To Patient86
  • Insufficient Information21
  • Not Applicable6
  • No Code Available2
  • No Information2

Recent adverse event reports

The 25 most recent of 324 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 21, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50845881 ↗
  • MalfunctionSep 18, 2026Device-Device Incompatibility; Mechanical Jam

    It was reported that during trialing for the unipolar ball the scrub tech had difficulty applying the mod endo head trial to the handle. It eventually went on but was difficult to remove after the surgery when instruments were returned to their trays. After decontamination the reporter found one or two perimeter ball bearings did not fully depress. The reporter soaked the handle in a milk/lubricant recommended by the hospital's SPD director, which did not resolve the problem. The issue did not cause delay to the surgery. The head trial was not affected. The instrument was removed from use. All available information has been disclosed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820270 ↗
  • MalfunctionSep 3, 2026Device-Device Incompatibility; Mechanical Jam

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50596582 ↗
  • MalfunctionMay 20, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT THE BIPOLAR HEAD TRIAL WON'T COME OFF ANGLED INSERTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25230819 ↗
  • MalfunctionMay 18, 2026Corroded; Contamination /Decontamination Problem

    IT WAS REPORTED THAT DURING WASHING OF THE SELF-CENTERING MODULAR CATHCART HANDLE ANGLED DEVICE, BROWN FLUID LEAKED FROM THE END. WATER HAD LIKELY ENTERED UNDER THE BALL BEARINGS AND CAUSED RUST. NO DELAYS OCCURRED DURING THE HIP SURGERY. NO PATIENT INJURIES OCCURRED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25204762 ↗
  • MalfunctionMay 14, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT DURING A HIP PROCEDURE, ONE OF THE BALL BEARINGS WAS STUCK AND THE TRIAL DOES NOT GO ON THE HANDLE. NO DELAYS DURING DURING THE CASE AND NO PATIENT INJURIES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25176604 ↗
  • MalfunctionMay 6, 2026Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE HOSPITAL FOUND BIOBURDEN IN ALL OF THEIR BIPOLAR ENDO HEAD TRIAL HANDLES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25092874 ↗
  • MalfunctionMay 6, 2026Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE HOSPITAL FOUND BIOBURDEN IN ALL OF THEIR BIPOLAR ENDO HEAD TRIAL HANDLES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25092813 ↗
  • MalfunctionMay 6, 2026Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE HOSPITAL FOUND BIOBURDEN IN ALL OF THEIR BIPOLAR ENDO HEAD TRIAL HANDLES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25092797 ↗
  • MalfunctionMay 6, 2026Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE HOSPITAL FOUND BIOBURDEN IN ALL OF THEIR BIPOLAR ENDO HEAD TRIAL HANDLES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25092767 ↗
  • MalfunctionApr 2, 2026Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE SPRING MECHANISM IN THE DEVICES HOLDS BIO BURDEN, RESULTING IN THE FACILITY'S DECISION TO DISCONTINUE STERILIZATION BY SPD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24769127 ↗
  • MalfunctionMar 19, 2026Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE HANDLE CAN NO LONGER BE CLEANED. BLOOD GETTING CAUGHT BEHIND THE BEARINGS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24640746 ↗
  • MalfunctionMar 3, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT THEY WERE DOING A HEMI. THE CURVED WHITE HANDLE THAT ATTACHES TO THE HEMI TRIALS WAS TOUGH TO GET ON AND OFF. THIS DID NOT SLOW THE SURGERY DOWN AND THE SURGERY PROCEEDED AS NORMAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24494668 ↗
  • MalfunctionFeb 23, 2026Contamination /Decontamination Problem

    IT WAS REPORTED BEHIND THE INSTRUMENTS` SPRING BALLS BLOOD, FLUID, TISSUE, ETC. IS STILL APPARENT AFTER REPROCESSING. INSTRUMENT IS NOT POSSIBLE TO BE REPROCESSED PROPERLY. NO ADVERSE PATIENT IMPACT, NO SURGICAL DELAY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24421026 ↗
  • MalfunctionFeb 20, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT S/C TRIAL HANDLE AND S/C TRIAL HANDLE ANGLED WERE DUE TO AN INABILITY TO DISASSEMBLE THE INSTRUMENTS, WHICH PREVENTED CLEANING BEHIND THE SPRING BALLS. THE EVENT WAS IDENTIFIED IN THE CENTRAL STERILE SERVICES DEPARTMENT, AND THE CUSTOMER DECLINED TO REPROCESS THE INSTRUMENTS. THERE WAS NO ADVERSE PATIENT IMPACT OR SURGICAL DELAY ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24400196 ↗
  • MalfunctionFeb 19, 2026Contamination /Decontamination Problem

    IT WAS REPORTED THAT AFTER THE SURGICAL PROCEDURE, BEHIND THE INSTRUMENTS SPRING BALLS BLOOD, FLUID, TISSUE, WERE STILL APPARENT AFTER REPROCESSING. THE INSTRUMENT WAS NOT POSSIBLE TO BE REPROCESSED PROPERLY. NO ADVERSE PATIENT IMPACT, NO SURGICAL DELAY REPORTED. S/C TRIAL HANDLE ANGLED: 205513000: NOT APPLICABLE. LIST ANY J&J PRODUCTS THAT WERE UTILIZED DURING THE CASE BUT DID NOT CONTRIBUTE TO THE REPORTED EVENT. WAS THERE ANY REPORTED PATIENT OR USER HARM? NO. WAS THERE A SIGNIFICANT DELAY? NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24393169 ↗
  • MalfunctionFeb 18, 2026Corroded; Contamination /Decontamination Problem

    IT WAS REPORTED THAT DURING STERILE PROCESSING IT WAS OBSERVED THAT THE HANDLE WAS HOLDING BIO BURDEN AND RUST IN THE ATTACHMENT PIECE FOR THE HEAD TRIALS AND COULD NOT BE CLEANED PROPERLY. THE EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24384657 ↗
  • MalfunctionFeb 13, 2026Contamination /Decontamination Problem

    IT WAS REPORTED THAT IT DID NOT HAPPEN IN SURGERY, SPD FOUND DEBRIS COMING OUT OF THE LOCKING MECHANISM AND CANNOT BE CLEANED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24333262 ↗
  • MalfunctionFeb 12, 2026Device-Device Incompatibility; No Apparent Adverse Event

    IT WAS REPORTED THAT THE INSTRUMENT WAS NOT HOLDING PROPERLY DUE TO WEAR. IT WAS REPORTED THAT THERE WAS NO DELAY IN THE PROCEDURE DUE TO THE EVENT. IT WAS REPORTED THAT THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24321691 ↗
  • MalfunctionFeb 5, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT IT IS VERY DIFFICULT TO GET THE MOD ENDO HEAD TRIALS ON AND OFF THE WHITE HANDLE, AS THEY TEND TO GET STUCK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24270603 ↗
  • MalfunctionJan 13, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT THE HANDLES DID NOT CONNECT PROPERLY TO THE TRIALS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24053799 ↗
  • MalfunctionJan 2, 2026Device-Device Incompatibility; No Apparent Adverse Event

    IT WAS REPORTED THAT THE BALL-BEARINGS ON THE UNIPOLAR/BIPOLAR ANGLED HANDLE WAS WORN. TROUBLE ATTACHING HEAD TRIALS ON TO HANDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23949572 ↗
  • MalfunctionJan 2, 2026Device-Device Incompatibility; No Apparent Adverse Event

    IT WAS REPORTED THAT THE BALL-BEARINGS ON THE UNIPOLAR/BIPOLAR ANGLED HANDLE WAS WORN. TROUBLE ATTACHING HEAD TRIALS ON TO HANDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23949398 ↗
  • MalfunctionDec 22, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT STERILE PROCESSING DEPARTMENT COULD NOT CLEAN THESE INSTRUMENTS DUE TO DESIGN OF BALL BEARING. THE FAILURE WAS NOT DISCOVERED DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23872608 ↗
  • MalfunctionDec 22, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT STERILE PROCESSING DEPARTMENT COULD NOT CLEAN THESE INSTRUMENTS DUE TO DESIGN OF BALL BEARING. THE FAILURE WAS NOT DISCOVERED DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23872592 ↗

Other IMPACTOR devices

All IMPACTOR devices →