This DePuy MBT Revision Tibial Broach is a surgical broach. The 40 adverse event reports describe malfunctions such as missing parts, device-device incompatibility, breakage, and bent or deformed parts; accounts also mention dull or worn instruments, poor fit with handles, and trouble securing stem trials. There were no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
40
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 10603295096078), received between Jul 18, 2017 and Oct 10, 2024 — classified as 40 malfunctions.

The most frequently reported specific device problem is “Component Missing” (21 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Component Missing21
  • Device-Device Incompatibility7
  • Appropriate Device Problem Term/Code Not Available2
  • Break2
  • Material Deformation2
  • Material Twisted/Bent2
  • Scratched Material2
  • Connection Problem1

Patient problems

  • No Consequences Or Impact To Patient33
  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions2
  • No Code Available2
  • No Information2
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 10, 2024Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT ON (B)(6) 2024, THE BROACH IS WARPED AND DOES NOT FIT ON HANDLE. THE METAL TIP BROKE OFF ON WRENCH. UNKNOWN SURGICAL DELAY.

    Patient problems: Insufficient Information

    MAUDE report 20417725 ↗
  • MalfunctionJan 4, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE TRIAL STEM IS UNABLE TO LOCK ON TO BROACH HANDLE. ITEM DEFECTIVE. NO MISSING PIECES. NO FURTHER INFORMATION IS AVAILABLE. NOT USED IN SURGERY. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13151451 ↗
  • MalfunctionOct 13, 2021Device-Device Incompatibility; Scratched Material; Device-Device Incompatibility; Scratched Material

    "CONNECTION IN THE SLEEVE BROACH WAS NICKED BY THE SAW BLADE. IT WASN'T ABLE TO HITCH ON TO THE HANDLE AFTER THAT."

    Patient problems: Insufficient Information

    MAUDE report 12624450 ↗
  • MalfunctionApr 26, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE BROACHES ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11726957 ↗
  • MalfunctionOct 2, 2020Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT THE PRODUCT D961673 IS WORN AND NEEDS TO BE REPLACED, THE BLACK END THAT GOES OVER THE IMPLANT TO TIGHTEN WAS SLIPPING WHEN THE SURGEON WAS PUTTING IMPLANTS TOGETHER. PRODUCT D217863109 AND D217863111 ARE NOT HOLDING STEM TRIALS WELL DUE TO THE WEAR AT THE CONNECTION SITE. NEEDS TO BE REPLACED. NO SURGICAL DELAY.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 10623941 ↗
  • MalfunctionMay 11, 2020Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE TWO TIBIAL BROACHES WOULDN¿T SECURELY HOLD THE STEM TRIALS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10043730 ↗
  • MalfunctionNov 20, 2019Dull, Blunt; Dull, Blunt

    IT WAS REPORTED THAT THE BROACH WAS VERY DULL AND NEEDS TO BE REPLACED. SHIP REPLACEMENT TO EAST COOPER FOR P1 DELIVERY. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9348210 ↗
  • MalfunctionOct 24, 2019Device-Device Incompatibility; Device-Device Incompatibility

    REFURBISH WON'T ATTACH TO BROACH HANDLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9234227 ↗
  • MalfunctionMay 28, 2019Component Missing; Material Twisted/Bent; Component Missing; Material Twisted/Bent

    AN MBT SLEEVE WAS MISSING THE ORANGE RING IN FEMALE PIECE THAT ATTACHES TO STEM TRIALS. NO TIME WAS ADDED TO THE CASE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8646030 ↗
  • MalfunctionMay 10, 2019Scratched Material; Scratched Material

    IT WAS REPORTED THAT THE RUBBER GASKET IS TORN IN 29MM SIGMA TIBIAL BROACH.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8601002 ↗
  • MalfunctionApr 2, 2019Failure to Clean Adequately; Failure to Clean Adequately

    REV. G) FOR INSTRUMENTS OF THIS SYSTEM AS THERE IS A RUBBER-O-RING IN THE INSTRUMENT SO THAT THE TRIAL STEM WILL HAVE A TIGHT FIT. BECAUSE OF THIS THE INSTRUMENT IS DIFFICULT TO CLEAN. THE GIVEN REF IS AN EXAMPLE. NO SURGERY DELAY, NO ADVERSE CONSEQUENCES FOR PATIENT REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8475931 ↗
  • MalfunctionMar 29, 2019Component Missing; Material Deformation; Naturally Worn; Component Missing; Material Deformation; Naturally Worn

    THE 2 INSTRUMENTS ARE WORN AND THE COUPLING MECHANISM HAS BECOME FAULTY DUE TO THE STEM RETAINING RING IN THE COUPLING MECHANISM, HAVING BEEN LOST. THIS DID NOT OCCUR DURING A PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8465414 ↗
  • MalfunctionFeb 4, 2019Fitting Problem; Fitting Problem

    THE STEM TRIAL DOES NOT CORRECTLY ENGAGE WITH BROACH.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8301177 ↗
  • MalfunctionJan 30, 2019Loose or Intermittent Connection; Loose or Intermittent Connection

    THE TIBIAL BROACH COULD NOT HOLD STEM TRIAL IN PLACE. THE SPRING INSIDE WAS LOOSE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8288864 ↗
  • MalfunctionJan 17, 2019Component Missing; Component Missing

    WORN INSTRUMENTS, NEED TO BE REPLACED. SOME OF THE "O RINGS" ARE MISSING FROM THE REAMERS

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8258608 ↗
  • MalfunctionOct 30, 2018Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE DOCTOR AND SCRUB NURSE STRUGGLED CONNECTING BROACHES AND REAMERS TO TRIAL STEMS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8019233 ↗
  • MalfunctionAug 13, 2018Component Missing; Component Missing

    IT WAS REPORTED THAT THE 29 SLEEVE BROACH IS MISSING WASHER SO STEM TRIAL WILL NOT STAY ATTACHED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7777230 ↗
  • MalfunctionAug 9, 2018Component Missing; Component Missing

    IT WAS REPORTED THAT THE 29 SLEEVE BROACH IS MISSING WASHER SO STEM TRIAL WILL NOT STAY ATTACHED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7767743 ↗
  • MalfunctionAug 9, 2018Component Missing; Component Missing

    IT WAS REPORTED THAT THE 29 SLEEVE BROACH IS MISSING WASHER SO STEM TRIAL WILL NOT STAY ATTACHED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7767715 ↗
  • MalfunctionAug 9, 2018Component Missing; Component Missing

    IT WAS REPORTED THAT THE 29 SLEEVE BROACH IS MISSING WASHER SO STEM TRIAL WILL NOT STAY ATTACHED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7767626 ↗
  • MalfunctionAug 9, 2018Component Missing; Component Missing

    IT WAS REPORTED THAT THE 29 SLEEVE BROACH IS MISSING WASHER SO STEM TRIAL WILL NOT STAY ATTACHED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7767620 ↗
  • MalfunctionAug 9, 2018Component Missing; Component Missing

    IT WAS REPORTED THAT THE 29 SLEEVE BROACH IS MISSING WASHER SO STEM TRIAL WILL NOT STAY ATTACHED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7767095 ↗
  • MalfunctionAug 9, 2018Component Missing; Component Missing

    IT WAS REPORTED THAT THE 29 SLEEVE BROACH IS MISSING WASHER SO STEM TRIAL WILL NOT STAY ATTACHED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7767085 ↗
  • MalfunctionAug 9, 2018Component Missing; Component Missing

    IT WAS REPORTED THAT THE 29 SLEEVE BROACH IS MISSING WASHER SO STEM TRIAL WILL NOT STAY ATTACHED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7767063 ↗
  • MalfunctionAug 9, 2018Component Missing; Component Missing

    IT WAS REPORTED THAT THE 29 SLEEVE BROACH IS MISSING WASHER SO STEM TRIAL WILL NOT STAY ATTACHED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7767051 ↗