Skip to content

NA 2178-63-170Recalls & reported problems

Depuy (Ireland) · M.B.T. REVISION STRAIGHT REAMER 10mm · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
17
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10603295096276), received between Jul 27, 2017 and Sep 30, 2021 — classified as 17 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device-Device Incompatibility6
  • Dull, Blunt4
  • Material Twisted/Bent4
  • Material Deformation3
  • Naturally Worn3
  • Loose or Intermittent Connection2
  • Appropriate Device Problem Term/Code Not Available1
  • Failure to Cut1

Patient problems

  • No Consequences Or Impact To Patient8
  • No Clinical Signs, Symptoms or Conditions4
  • No Information3
  • Insufficient Information2
  • Not Applicable2

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT WHILE REAMING FOR A CEMENT SPACER THE 10MM REAMER GOT STUCK SO THE SURGEON, RESIDENT WERE TRYING TO PULL OUT WITH PLIERS AND ALSO TAPPING OUT WITH VICE GRIPS AND MALLET. IT CAME OUT PUT IT MESSED UP THE END OF THE REAMER AND IT WILL NOT ATTACH TO THE T-HANDLE CORRECTLY NOW. NEEDS REPLACED. NO SURGICAL DELAY

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12554094 ↗
  • MalfunctionSep 29, 2021Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE ATTACHMENT END OF 10MM REAMER IS BURRED SO IT GETS STUCK ON MODIFIED HUDSON ADAPTER. MODIFIED HUDSON ADAPTER IS STRIPPED THEREFORE THE REAMERS WILL NOT WORK PROPERLY WHEN ATTACHED TO THE POWER. THE CONNECTING END OF THE FEM ADAPTER TORQUE WRENCH IS WORN AND WILL NOT HOLD FEM ADAPTER FIRMLY IN PLACE WHEN APPLYING TORQUE. PLEASE SEND REPLACEMENTS P1 TO TRIDENT ATTN: OR/JASON OZMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12548841 ↗
  • MalfunctionApr 30, 2021Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT THE MBT REAMERS WERE SLIPPING IN THE HUDSON ADAPTER AND T-HANDLE DURING THE CASE. AFTER THE CASE, THE INSTRUMENTS SHOWED WEAR DUE TO USAGE. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11758420 ↗
  • MalfunctionApr 20, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE SIZE 10 AND 12 REVISION REAMER WOULD STICK WHILE ATTACHED IN T-HANDLE AND WAS DIFFICULT REMOVE. ALL OTHER SIZES USED WOULD NOT STICK. THERE WAS A 30 SECOND DELAY IN THE CASE. LOT NUMBERS ON BOTH REAMERS UNAVAILABLE NOT ABLE TO READ.

    Patient problems: Insufficient Information

    MAUDE report 11693997 ↗
  • MalfunctionDec 29, 2020Failure to Cut; Device-Device Incompatibility; Failure to Cut; Device-Device Incompatibility

    IT WAS REPORTED THAT T HANDLE, CHUCKS, AND REAMERS ARE ALL DULL AND DO NOT STAY LOCKED TOGETHER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11090074 ↗
  • MalfunctionDec 22, 2020Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT ALL REAMERS WERE LOOSE ON THE ADAPTERS AND SPUN FREELY. ADAPTER WOULD SPIN FREELY WHEN ATTACHED TO REAMERS OR ANY OTHER ATTACHMENT. T-HANDLE WAS ALSO LOOSE. ATTACHMENT POINT SEEMS TO BE WEAK AND CAME FREE AT FIRST SIGN OF RESISTANCE. NO PIECES BROKE OFF IN PATIENT. ALL PIECES WERE REMOVED FROM PATIENT. SURGERY TIME WAS EXTENDED BY APPROXIMATELY 30 AS A RESULT OF DEPUY INSTRUMENTS NOT FUNCTIONING PROPERLY.

    Patient problems: Insufficient Information

    MAUDE report 11057538 ↗
  • MalfunctionOct 3, 2019Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE SURGEON WAS REAMING WITH THE HUDSON ADAPTOR AND REAMERS THE HUDSON ATTACHMENT AND THE REAMERS STRIPPED AND HAD DIFFICULTY REAMING THE FEMUR AND THE TIBIA. THEY TRIED THIS WITH BOTH SETS OF OUR HUDSON¿S AND T-HANDLES AND REAMERS. ALSO THE SIZE 4 TC3 BOX TRIAL WOULD NOT STAY ON THE FEMUR TRIAL SO THINK THE TORQUE SCREW MUST BE STRIPPED. A SURGICAL DELAY OF 15-20 MINUTES.

    Patient problems: Not Applicable

    MAUDE report 9152460 ↗
  • MalfunctionOct 3, 2019Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE SURGEON WAS REAMING WITH THE HUDSON ADAPTOR AND REAMERS THE HUDSON ATTACHMENT AND THE REAMERS STRIPPED AND HAD DIFFICULTY REAMING THE FEMUR AND THE TIBIA. THEY TRIED THIS WITH BOTH SETS OF OUR HUDSON¿S AND T-HANDLES AND REAMERS. ALSO THE SIZE 4 TC3 BOX TRIAL WOULD NOT STAY ON THE FEMUR TRIAL SO THINK THE TORQUE SCREW MUST BE STRIPPED. A SURGICAL DELAY OF 15-20 MINUTES.

    Patient problems: Not Applicable

    MAUDE report 9152456 ↗
  • MalfunctionSep 27, 2019Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT ALL REAMERS ARE STRIPPED AND DO NOT STAY ATTACHED TO THE HUDSON. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9129106 ↗
  • MalfunctionFeb 14, 2019Dull, Blunt; Naturally Worn; Dull, Blunt; Naturally Worn

    IT WAS REPORTED THAT THE SIZE 10-18 MBT REVISION REAMERS ARE DULL AND SKIPPING ON ADAPTER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8339860 ↗
  • MalfunctionFeb 14, 2019Dull, Blunt; Naturally Worn; Dull, Blunt; Naturally Worn

    IT WAS REPORTED THAT THE SIZE 10-18 MBT REVISION REAMERS ARE DULL AND SKIPPING ON ADAPTER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8339732 ↗
  • MalfunctionDec 3, 2018Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE TOP OF THE REAMERS WERE ROUNDED OUT AND WOULD NOT STAY LOCKED IN THE HUDSON ADAPTER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8124007 ↗
  • MalfunctionJul 10, 2018Material Deformation; Naturally Worn; Material Deformation; Naturally Worn

    IT WAS REPORTED THAT THE INSTRUMENTS NEED A REPAIR.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7673005 ↗
  • MalfunctionJan 29, 2018Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT THE MBT REAMERS SLIPPING IN THE MODIFIED HUDSON ADAPTER WHILE REAMING BONE. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7224927 ↗
  • MalfunctionNov 15, 2017Dull, Blunt; Dull, Blunt

    COMPLAINT DESCRIPTION: IT WAS REPORTED THAT THE REAMERS ARE SCARRED AT THE UNION FROM USAGE OVER THE YEARS. / / INVESTIGATION METHOD: THE COMPLAINT SAMPLE CONSISTED OF (4) 217863136 MODIFIED HUDSON ADAPTER, LOT NUMBERS TBPKE, H0609, AND TBLTW. THE REMAINING REPORTED 217863136 MODIFIED HUDSON ADAPTER WERE NOT RECEIVED FOR EXAMINATION. EXAMINATION OF THE FOUR SUBMITTED ADAPTERS CONFIRMED THE CONNECTION FEATURES THAT ATTACH TO THE REAMER ARE WORN/DAMAGED. THE DAMAGE IS CONSISTENT WITH THE REAMERS MATING FEATURE STRIPPING AFTER BEING SUBJECTED TO HARD CORTICAL BONE AND HEAVY USAGE CAUSING DAMAGE/WEAR TO THE ADAPTER. THE ROOT CAUSE IS ATTRIBUTED TO PRODUCT WEAR OUT. BASED ON THE INVESTIGATION'S DETERMINATION OF DEVICE WEAR FROM NORMAL USE AND SERVICING, NO CORRECTIVE ACTION IS BEING PURSUED. CONTINUE TO MONITOR VIA SEP-419. / / INVESTIGATION SUMMARY: IT WAS REPORTED THAT THE REAMERS ARE SCARRED AT THE UNION FROM USAGE OVER THE YEARS.

    Patient problems: No Consequences Or Impact To Patient; No Information

    MAUDE report 7033291 ↗
  • MalfunctionNov 15, 2017Dull, Blunt; Dull, Blunt

    COMPLAINT DESCRIPTION: IT WAS REPORTED THAT THE REAMERS ARE SCARRED AT THE UNION FROM USAGE OVER THE YEARS. / / INVESTIGATION METHOD: THE COMPLAINT SAMPLE CONSISTED OF (4) 217863136 MODIFIED HUDSON ADAPTER, LOT NUMBERS TBPKE, H0609, AND TBLTW. THE REMAINING REPORTED 217863136 MODIFIED HUDSON ADAPTER WERE NOT RECEIVED FOR EXAMINATION. EXAMINATION OF THE FOUR SUBMITTED ADAPTERS CONFIRMED THE CONNECTION FEATURES THAT ATTACH TO THE REAMER ARE WORN/DAMAGED. THE DAMAGE IS CONSISTENT WITH THE REAMERS MATING FEATURE STRIPPING AFTER BEING SUBJECTED TO HARD CORTICAL BONE AND HEAVY USAGE CAUSING DAMAGE/WEAR TO THE ADAPTER. THE ROOT CAUSE IS ATTRIBUTED TO PRODUCT WEAR OUT. BASED ON THE INVESTIGATION'S DETERMINATION OF DEVICE WEAR FROM NORMAL USE AND SERVICING, NO CORRECTIVE ACTION IS BEING PURSUED. CONTINUE TO MONITOR VIA SEP-419. / / INVESTIGATION SUMMARY: IT WAS REPORTED THAT THE REAMERS ARE SCARRED AT THE UNION FROM USAGE OVER THE YEARS.

    Patient problems: No Consequences Or Impact To Patient; No Information

    MAUDE report 7033228 ↗
  • MalfunctionJul 27, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING ROUTINE INSPECTION REAMER FAILED TO STAY ATTACHED TO REAMER HANDLES OR HUDSON END. REAMER APPEAR TO BE ROUND OFF AND WILL NOT TURN WHEN ENGAGED WITH POWER REAMER OR WITH HAND REAMER. TIPS APPEAR TO BE WORN.

    Patient problems: No Information

    MAUDE report 6746797 ↗

Other REAMER devices

All REAMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page