Quick summary

This DePuy Pinnacle 32mm Poly Impactor Tip is a surgical instrument used to seat hip implant components. The 50 reports describe instruments wearing out with normal use — cracking, scratching, deforming and breaking, and tips and trial shells that no longer connect or thread smoothly. No deaths and no injuries were reported; most mentions note no effect on the patient. There are no recalls naming this device. Reports rose from 2 in the prior 12 months to 11 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
49
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 49 adverse event reports that list this exact device (DI 10603295098966), received between Jan 26, 2018 and May 6, 2026 — classified as 49 malfunctions.

The most frequently reported specific device problem is “Naturally Worn” (22 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 4 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 5 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Naturally Worn22
  • Scratched Material17
  • Material Deformation16
  • Crack9
  • Device-Device Incompatibility8
  • Break7
  • Material Twisted/Bent4
  • Unintended Movement2

Patient problems

  • No Clinical Signs, Symptoms or Conditions31
  • No Consequences Or Impact To Patient12
  • Insufficient Information6

Recent adverse event reports

The 25 most recent of 49 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 6, 2026Device-Device Incompatibility

    ALL EQUIPMENT PANS HAVE BECOME CRACKED THROUGH REGULAR USE. IMPACTOR TIPS NO LONGER CONNECT SMOOTHLY. TRIAL SHELLS NO LONGER THREAD TOGETHER SMOOTHLY. THERE WAS NO REPORTED PATIENT OR USER HARM. THERE WAS NO SIGNIFICANT DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25093522 ↗
  • MalfunctionMay 6, 2026Device-Device Incompatibility

    ALL EQUIPMENT PANS HAVE BECOME CRACKED THROUGH REGULAR USE. IMPACTOR TIPS NO LONGER CONNECT SMOOTHLY. TRIAL SHELLS NO LONGER THREAD TOGETHER SMOOTHLY. THERE WAS NO REPORTED. PATIENT OR USER HARM. THERE WAS NO SIGNIFICANT DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25093514 ↗
  • MalfunctionMay 6, 2026Device-Device Incompatibility

    ALL EQUIPMENT PANS HAVE BECOME CRACKED THROUGH REGULAR USE. IMPACTOR TIPS NO LONGER CONNECT SMOOTHLY. TRIAL SHELLS NO LONGER THREAD TOGETHER SMOOTHLY. THERE WAS NO REPORTED PATIENT OR USER HARM. THERE WAS NO SIGNIFICANT DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25093513 ↗
  • MalfunctionMay 6, 2026Device-Device Incompatibility

    ALL EQUIPMENT PANS HAVE BECOME CRACKED THROUGH REGULAR USE. IMPACTOR TIPS NO LONGER CONNECT SMOOTHLY. TRIAL SHELLS NO LONGER THREAD TOGETHER SMOOTHLY. THERE WAS NO REPORTED PATIENT OR USER HARM. THERE WAS NO SIGNIFICANT DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25093510 ↗
  • MalfunctionMay 5, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT ALL EQUIPMENT PANS HAVE BECOME CRACKED THROUGH REGULAR USE. THE IMPACTOR TIPS NO LONGER CONNECT SMOOTHLY, AND THE TRIAL SHELLS NO LONGER THREAD TOGETHER SMOOTHLY. IT WAS REPORTED THAT THERE WAS NO SIGNIFICANT DELAY, AND NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25084911 ↗
  • MalfunctionFeb 3, 2026Crack; Material Deformation

    IT WAS REPORTED THAT SEVERAL INSTRUMENTS HAVE BEEN DEEMED NO LONGER USABLE DUE TO BEING CRACKED, CROSS THREADED WITH WEAR AND TEAR OVER TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24246491 ↗
  • MalfunctionFeb 3, 2026Crack; Material Deformation

    IT WAS REPORTED THAT SEVERAL INSTRUMENTS HAVE BEEN DEEMED NO LONGER USABLE DUE TO BEING CRACKED, CROSS THREADED WITH WEAR AND TEAR OVER TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24246490 ↗
  • MalfunctionFeb 3, 2026Crack; Material Deformation

    IT WAS REPORTED THAT SEVERAL INSTRUMENTS HAVE BEEN DEEMED NO LONGER USABLE DUE TO BEING CRACKED, CROSS THREADED WITH WEAR AND TEAR OVER TIME. THESE INSTRUMENTS DO GET REPLACED EVERY SO OFTEN DUE TO USAGE OVER TIME AS MENTIONED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24246367 ↗
  • MalfunctionFeb 3, 2026Crack; Material Deformation

    IT WAS REPORTED THAT SEVERAL INSTRUMENTS HAVE BEEN DEEMED NO LONGER USABLE DUE TO BEING CRACKED, CROSS THREADED WITH WEAR AND TEAR OVER TIME. THESE INSTRUMENTS DO GET REPLACED EVERY SO OFTEN DUE TO USAGE OVER TIME AS MENTIONED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24246364 ↗
  • MalfunctionJan 30, 2026Crack; Material Deformation

    SEVERAL INSTRUMENTS HAVE BEEN DEEMED NO LONGER USABLE DUE TO BEING CRACKED, CROSS THREADED ETC WITH WEAR AND TEAR OVER TIME. THESE INSTRUMENTS DO GET REPLACED EVERY SO OFTEN DUE TO USAGE OVER TIME AS MENTIONED. NO FURTHER INFORMATION IS AVAILABLE. PLEASE SEND REPLACEMENTS DIRECT TO THE ACCOUNT: [REDACTED] [REDACTED] [REDACTED] [REDACTED] N/C PO: [REDACTED] BILL TO: [REDACTED] SHIP TO: [REDACTED] PINN STRAIGHT CUP IMPACTOR - 221750041 (221750041): NOT APPLICABLE DURALOC STR CUP IMPACTOR - 236071000 (236071000): NOT APPLICABLE PINN CUP IMPACTOR TIP 32MM - 221750007 (221750007): NOT APPLICABLE PINN POLY IMPACTOR TIP 36MM - 221750008 (221750008): NOT APPLICABLE QUICKSET EZ CLEAN GRATER HAND. - 979810000 (979810000): NOT APPLICABLE PINNACLE VERSION GUIDE - 221750044 (221750044): NOT APPLICABLE APEX HOLE ELIM TPD HEX DRIVER - 234601000 (234601000): NOT APPLICABLE SUMMIT UNIVERSAL BROACH HANDLE - 257000000 (257000000): NOT APPLICABLE UNIVERSAL STEM INSERT HANDLE - 259807460 (259807460): NOT APPLICABLE CALCAR MILL MEDIUM - 200148000 (200148000): NOT APPLICABLE MODULAR BOX OSTEOTOME - 259807530 (259807530): NOT APPLICABLE ARTICUL TR BALL GRVD 36-2 - 253150000 (253150000): NOT APPLICABLE ART

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24221179 ↗
  • MalfunctionOct 15, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE MEDICAL DEVICE REPROCESSING STAFF NOTICED THAT THE PINNACLE CUP INSERTER THREADS WERE CROSS THREADED, AND ALL THREE SIZES OF THE IMPACTOR TIPS WERE ALSO CROSS THREADED AND SO WAS THE DURALOC IMPACTOR HANDLE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23304123 ↗
  • MalfunctionJul 16, 2025Break; Break

    IT WAS REPORTED THAT THE ITEM WAS BROKEN AMD NEEDS REPLACING. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22521049 ↗
  • MalfunctionOct 30, 2024Contamination /Decontamination Problem; Material Deformation; Naturally Worn; Contamination /Decontamination Problem; Material Deformation; Naturally Worn

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL HIP REPLACEMENT SURGERY ON (B)(6) 2024. THE THREADS ON THE IMPACTOR TIP HAVE WORN DOWN AND CAN NO LONGER BE PROPERLY CLEANED. NO FURTHER DETAILS WERE PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 20566371 ↗
  • MalfunctionSep 10, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE 28 AND 32 IMPACTOR TIP THREADS ARE BAD.

    Patient problems: Insufficient Information

    MAUDE report 20180288 ↗
  • MalfunctionAug 9, 2024Device-Device Incompatibility; Unintended Movement; Device-Device Incompatibility; Unintended Movement

    IT WAS REPORTED THAT THE IMPACTORS NOT USED FOR THE CASE BUT THE SLEEVES OF THE IMPACTOR TIPS ARE LOOSE AND EASILY DETACH FROM THE TIP.

    Patient problems: Insufficient Information

    MAUDE report 19947861 ↗
  • MalfunctionJul 19, 2023Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT WORN DAMAGE TO TWO 32 BLUE IMPACTOR HEADS FOR LINER IMPACTION. NO SURGICAL DELAY.

    Patient problems: Insufficient Information

    MAUDE report 17353691 ↗
  • MalfunctionJul 19, 2023Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT WORN DAMAGE TO TWO 32 BLUE IMPACTOR HEADS FOR LINER IMPACTION. NO SURGICAL DELAY.

    Patient problems: Insufficient Information

    MAUDE report 17353654 ↗
  • MalfunctionApr 4, 2023Naturally Worn; Scratched Material; Naturally Worn; Scratched Material

    IT WAS REPORTED THAT TRIAL HEAD IS DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16674913 ↗
  • MalfunctionJan 5, 2023Material Deformation; Naturally Worn; Scratched Material; Material Deformation; Naturally Worn; Scratched Material

    IT WAS REPORTED THAT THE TIP WAS DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16108110 ↗
  • MalfunctionJan 5, 2023Material Deformation; Naturally Worn; Scratched Material; Material Deformation; Naturally Worn; Scratched Material

    IT WAS REPORTED THAT THE INSTRUMENT WAS DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16107025 ↗
  • MalfunctionJan 5, 2023Material Deformation; Naturally Worn; Scratched Material; Material Deformation; Naturally Worn; Scratched Material

    IT WAS REPORTED THAT THE INSTRUMENT WAS DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16106949 ↗
  • MalfunctionJan 3, 2023Material Deformation; Naturally Worn; Scratched Material; Material Deformation; Naturally Worn; Scratched Material

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED, AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16089050 ↗
  • MalfunctionJan 3, 2023Material Deformation; Naturally Worn; Scratched Material; Material Deformation; Naturally Worn; Scratched Material

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16088994 ↗
  • MalfunctionJan 3, 2023Material Deformation; Naturally Worn; Scratched Material; Material Deformation; Naturally Worn; Scratched Material

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED, AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16088965 ↗
  • MalfunctionJan 3, 2023Material Deformation; Naturally Worn; Scratched Material; Material Deformation; Naturally Worn; Scratched Material

    THE 28 AND 32MM LINER IMPACTOR TIPS ARE BURRED AND SCRATCHED UP DUE TO WEAR AND TEAR. THE HEX LINER SCREWDRIVERS ARE STRIPPED ON THE BOTTOM AND HAVE WORN DOWN OVER TIME. THEY NO LONGER CONNECT INTO THE TRIAL LINER. THE PINNACLE STRAIGHT CUP INSERTERS HAVE BROKEN OFF STRIKE PLATES ON THE PROXIMAL END OF THE INSTRUMENT. THEY CAN NO LONGER IMPACT PINNACLE CUPS. THE STANDARD AND HIGH OFFSET NECK TRIALS ARE BURRED ON THE INSIDE AND NO LONGER CONNECT PROPERLY TO A TRIAL BROACH. THERE WERE NO PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16085616 ↗

FDA clearance and classification

The FDA regulates product code HWA (impactor) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the PINNACLE 2217-50-007

Has the PINNACLE 2217-50-007 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10603295098966), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the PINNACLE 2217-50-007?

The FDA's MAUDE database lists 49 adverse event reports for this device, received between Jan 26, 2018 and May 6, 2026. In those reports, the most common specific device problems are “Naturally Worn” (22), “Scratched Material” (17) and “Material Deformation” (16). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PINNACLE 2217-50-007?

None of the reports are classified as deaths, 0 as injuries and 49 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the PINNACLE 2217-50-007 FDA approved?

Devices of this type (product code HWA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).

Who makes the PINNACLE 2217-50-007, and is it still sold?

The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand PINNACLE, model 2217-50-007). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other IMPACTOR devices

All IMPACTOR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page