PINNACLE 2217-50-048Recalls & reported problems
Depuy (Ireland) · PINNACLE BANTAM ACETABULAR CUP IMPACTOR ADAPTOR · Class 1
All models of PINNACLE (198)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10603295099000), received between Dec 14, 2017 and Oct 17, 2025 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Entrapment of Device2
- Appropriate Device Problem Term/Code Not Available1
- Detachment of Device or Device Component1
- Loose or Intermittent Connection1
- Material Deformation1
- Material Twisted/Bent1
- Naturally Worn1
Patient problems
- No Consequences Or Impact To Patient5
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 17, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT IT WAS DISCOVERED DURING THE REVIEW AFTER WASHING BUT BEFORE THE OPERATION THAT THE GASKET ON ONE OF THE INSTRUMENTS WAS IN POOR CONDITION. IT APPEARS THAT THE RUBBER ON THE INSTRUMENT HAS WORN OUT DUE TO NORMAL USE. THE ITEM WAS NOT USED IN THE SURGERY SO NO AFFECT ON A SURGERY. THE WORN-OUT ITEM HAS BEEN REMOVED FROM THE MODULE AND REPLACED WITH A NEW ONE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23328140 ↗MalfunctionSep 13, 2024Naturally Worn; Naturally Worn
ADDITIONAL INFORMATION RECEIVED STATES THAT THE WELD AT THE HANDLE WAS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20213180 ↗MalfunctionJun 14, 2024Break; Break
ADDITIONAL INFORMATION WAS RECEIVED: WAS/WERE THERE ANY ADVERSE CONSEQUENCE/S THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT? NO. WAS THERE A SURGICAL DELAY? IF YES, WHAT IS THE DURATION OF THE DELAY? NO.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19536472 ↗MalfunctionMay 13, 2020Material Twisted/Bent; Scratched Material; Material Twisted/Bent; Scratched Material
DURING THE COVID PERIOD, HOSPITAL STAFF AT (B)(6) IN (B)(6) EXAMINED A VARIETY OF SETS AND CAME ACROSS SEVERAL DAMAGED INSTRUMENTS (7 TOTAL) IN THEIR PINNACLE BANTAM SET THAT REQUIRE REPLACING: THE PINNACLE IMPACTOR ADAPTER AND SHELL TRIAL WERE CROSS THREADED AND NO LONGER USABLE (221750048 & 221701038). THE 28MM LINER IMPACTOR WAS GAUGED (221750006). THE ARTICULEZE TRIAL BALLS AND NEUTRAL TRIAL LINERS WERE ALSO GAUGED AND NOT USABLE DUE TO WEAR AND TEAR (253070000, 253069000, 221828044 & 221828046). ALL ITEMS WERE DISCARDED. NO PATIENT OR INTRA-OP ISSUES AS THIS DID NOT OCCUR IN THE OPERATING ROOM. PLEASE SEND REPLACEMENTS TO THE FOLLOWING ADDRESS: (B)(6). PATIENT STATUS/ OUTCOME / CONSEQUENCES: NO, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN, IS THE PATIENT PART OF A CLINICAL STUDY: UNKNOWN, (B)(4), DEVICE PROPERTY OF: NONE, DEVICE IN POSSESSION OF: NONE, (B)(4), DEVICE PROPERTY OF: NONE, DEVICE IN POSSESSION OF: NONE, (B)(4), DEVICE PROPERTY OF: NONE, DEVICE IN POSSESSION OF: NONE, (B)(4), DEVICE PROPERTY OF: NONE, DEVICE IN POSSESSION OF: NONE, (B)(4), DEVICE PROPERTY OF: NONE, DEVICE IN POSSESSION OF: NONE
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10053532 ↗MalfunctionJan 17, 2019Entrapment of Device; Entrapment of Device
EVENT: TOTAL HIP REPLACEMENT SURGERY WAS PERFORMED, AT THE TIME OF PLACING THE FINAL ACETABULAR CUP PINNACLE # 40, AND UNSCREWING THE ENTIRE POSITIONER OF THE CUP AFTER BEING IMPACTED ON ITS ACCEPTABILITY. THE PIECE IN QUESTION IS "STUCK IN THE CUP", THIS INCIDENT CAUSES THE SURGEON TO HAVE TO USE OTHER CLAMPS TO UNSCREW THE PIECE FROM THE CUP, RISKING THAT THE SAME CUP MOVED FROM WHERE IT SHOULD GO. FINALLY, THE PIECE IS UNSCREWED IN A SOMEWHAT ABRUPT MANNER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8256978 ↗MalfunctionJan 17, 2019Entrapment of Device; Entrapment of Device
EVENT: TOTAL HIP REPLACEMENT SURGERY WAS PERFORMED, AT THE TIME OF PLACING THE FINAL ACETABULAR CUP PINNACLE # 40, AND UNSCREWING THE ENTIRE POSITIONER OF THE CUP AFTER BEING IMPACTED ON ITS ACCEPTABILITY. THE PIECE IN QUESTION IS "STUCK IN THE CUP", THIS INCIDENT CAUSES THE SURGEON TO HAVE TO USE OTHER CLAMPS TO UNSCREW THE PIECE FROM THE CUP, RISKING THAT THE SAME CUP MOVED FROM WHERE IT SHOULD GO. FINALLY, THE PIECE IS UNSCREWED IN A SOMEWHAT ABRUPT MANNER. THE PROCEDURE IS FINISHED WITHOUT ANY OTHER INCONVENIENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8256955 ↗MalfunctionAug 6, 2018Break; Detachment of Device or Device Component; Appropriate Device Problem Term/Code Not Available; Break; Detachment of Device or Device Component; Appropriate Device Problem Term/Code Not Available
RUBBER BUNG ON PINNACLE BANTAM CUP IMPACTOR CAME LOOSE AND FELL INTO PELVIC CREVICE INTRAOPERATIVELY. NO PATIENT COMPROMISE AS SURGEON WAS ABLE TO FISH RUBBER BUNG OUT. PATIENT CONSEQUENCE? :NO. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT? : YES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7753475 ↗MalfunctionDec 14, 2017Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT DURING SURGERY WHILE SURGEON IMPACTING THE CUP USING THE ADAPTOR (PART#: 221750048, LOT#: UNKNOWN) CONNECTED TO AN IMPACTOR (PART#: UNKNOWN), THE ADAPTOR GOT DISCONNECTED AND THEN LEFT ON THE CUP. HE REMOVED THE ADAPTOR FROM THE CUP USING A LONG NOSE PLIERS AND SO ON. DOE: (B)(6) 2017; SURGICAL DELAY: 10 MINUTES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7117497 ↗
FDA clearance and classification
The FDA regulates product code HWA (impactor) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the PINNACLE 2217-50-048
Has the PINNACLE 2217-50-048 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10603295099000), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the PINNACLE 2217-50-048?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Dec 14, 2017 and Oct 17, 2025. In those reports, the most common specific device problems are “Break” (2), “Entrapment of Device” (2) and “Detachment of Device or Device Component” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the PINNACLE 2217-50-048?
None of the reports are classified as deaths, 0 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the PINNACLE 2217-50-048 FDA approved?
Devices of this type (product code HWA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).
Who makes the PINNACLE 2217-50-048, and is it still sold?
The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand PINNACLE, model 2217-50-048). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other IMPACTOR devices
- PINNACLE 2217-50-041Depuy (Ireland) · IMPACTOR4,276 reports
- ATTUNE 2544-01-006Depuy (Ireland) · IMPACTOR1 recall2,528 reports
- ATTUNE 2544-01-003Depuy (Ireland) · IMPACTOR1 recall1,491 reports
- ATTUNE 2544-01-017Depuy (Ireland) · IMPACTOR1 recall1,454 reports
- ATTUNE 2544-01-004Depuy (Ireland) · IMPACTOR1 recall873 reports
- DURALOC 2360-71-000Depuy Orthopaedics, Inc. · IMPACTOR715 reports
- SPECIALIST 2581-11-000Depuy (Ireland) · IMPACTOR703 reports
- INHANCE 6205-10-103Depuy (Ireland) · IMPACTOR334 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page