PINNACLE 2217-50-060Recalls & reported problems
Depuy (Ireland) · PINNACLE POLY IMPACTOR TIP 40mm · Class 1
All models of PINNACLE (198)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10603295099093), received between Jul 24, 2023 and May 14, 2026 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Break” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break10
- Device-Device Incompatibility6
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions16
- Insufficient Information1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 14, 2026Break
IT WAS REPORTED THAT DURING A TOTAL HIP REPLACEMENT, UPON EXAMINATION IN RESETTING THE INSTRUMENTS, THE 40MM PINNACLE IMPACTOR TIP WAS SEPARATING AT THE CONNECTION OF METAL AND PINK POLY. THE INSTRUMENT WAS REMOVED FROM THE SET AND DISPOSED. THE TOTAL HIP REPLACEMENT WENT AS EXPECTED WITH NO DELAYS OR ISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25176774 ↗MalfunctionMay 6, 2026Device-Device Incompatibility
IT WAS REPORTED THAT ALL EQUIPMENT PANS HAVE BECOME CRACKED THROUGH REGULAR USE, THE IMPACTOR TIPS NO LONGER CONNECT SMOOTHLY. THE TRIAL SHELLS NO LONGER THREAD TOGETHER SMOOTHLY. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25094309 ↗MalfunctionMay 6, 2026Device-Device Incompatibility
IT WAS REPORTED THAT ALL EQUIPMENT PANS HAVE BECOME CRACKED THROUGH REGULAR USE, THE IMPACTOR TIPS NO LONGER CONNECT SMOOTHLY. THE TRIAL SHELLS NO LONGER THREAD TOGETHER SMOOTHLY. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25094302 ↗MalfunctionMay 6, 2026Device-Device Incompatibility
IT WAS REPORTED THAT ALL EQUIPMENT PANS HAVE BECOME CRACKED THROUGH REGULAR USE, THE IMPACTOR TIPS NO LONGER CONNECT SMOOTHLY. THE TRIAL SHELLS NO LONGER THREAD TOGETHER SMOOTHLY. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25094301 ↗MalfunctionMay 6, 2026Device-Device Incompatibility
IT WAS REPORTED THAT ALL EQUIPMENT PANS HAVE BECOME CRACKED THROUGH REGULAR USE, THE IMPACTOR TIPS NO LONGER CONNECT SMOOTHLY. THE TRIAL SHELLS NO LONGER THREAD TOGETHER SMOOTHLY. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25094299 ↗MalfunctionMay 6, 2026Device-Device Incompatibility
ALL EQUIPMENT PANS HAVE BECOME CRACKED THROUGH REGULAR USE. IMPACTOR TIPS NO LONGER CONNECT SMOOTHLY. TRIAL SHELLS NO LONGER THREAD TOGETHER SMOOTHLY. THERE WAS NO REPORTED PATIENT OR USER HARM. THERE WAS NO SIGNIFICANT DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25093512 ↗MalfunctionMay 5, 2026Device-Device Incompatibility
IT WAS REPORTED THAT ALL EQUIPMENT PANS HAVE BECOME CRACKED THROUGH REGULAR USE. THE IMPACTOR TIPS NO LONGER CONNECT SMOOTHLY, AND THE TRIAL SHELLS NO LONGER THREAD TOGETHER SMOOTHLY. IT WAS REPORTED THAT THERE WAS NO SIGNIFICANT DELAY, AND NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25085056 ↗MalfunctionMay 28, 2025Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THERE WAS A BROKEN INSTRUMENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22097096 ↗MalfunctionMay 23, 2025Break; Break
IT WAS REPORTED THE INSTRUMENTS WERE BROKEN. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22077582 ↗MalfunctionMay 23, 2025Break; Break
IT WAS REPORTED THE INSTRUMENTS WERE BROKEN. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22077482 ↗MalfunctionMay 23, 2025Break; Break
IT WAS REPORTED THE INSTRUMENTS WERE BROKEN. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22077425 ↗MalfunctionMay 23, 2025Break; Break
IT WAS REPORTED THE INSTRUMENTS WERE BROKEN. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22077414 ↗MalfunctionMay 23, 2025Break; Break
IT WAS REPORTED THE INSTRUMENTS WERE BROKEN. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22077406 ↗MalfunctionAug 15, 2024Unintended Movement; Unintended Movement
IT WAS REPORTED THAT THE REAMER HANDLE END BROKE AND THE PINK IMPACTOR TIP COLLAR WAS LOOSE. SURGICAL DELAY WAS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19992842 ↗MalfunctionJul 23, 2024Break; Break
IT WAS REPORTED THAT THE IMPACTOR TIP IS BROKEN.
Patient problems: Insufficient Information
MAUDE report 19804307 ↗MalfunctionJan 16, 2024Break; Break
IT WAS REPORTED THAT THE KINCISE WONT FIRE PROPERLY AFTER LIMITED USE.PLASTIC TRAYS BROKEN IN TRIALS, NOT COMPLAINT.(DEPUY HIP CORE CASE 2+ 4 MIDDLE TRAY, DEPUY HIP TRAY 1 +4 MIS, SUMMIT TRAPPED HIP)
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18516912 ↗MalfunctionJul 24, 2023Break; Break
ADDITIONAL INFORMATION RECEIVED. A. DID IT BREAK INTO TWO OR MORE PIECES? IF NO, PLEASE SPECIFY, THE ALLEGED DEFICIENCY OF THE INSTRUMENTS. WERE THEY BENT, STRIPPED, CROSS THREADED, ETC.? YES. THE METAL PART THAT THREADS ON TO AN IMPACTOR WAS NOT ATTACHED TO THE PLASTIC TIP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17379265 ↗
FDA clearance and classification
The FDA regulates product code HWA (impactor) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the PINNACLE 2217-50-060
Has the PINNACLE 2217-50-060 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10603295099093), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the PINNACLE 2217-50-060?
The FDA's MAUDE database lists 17 adverse event reports for this device, received between Jul 24, 2023 and May 14, 2026. In those reports, the most common specific device problems are “Break” (10), “Device-Device Incompatibility” (6) and “Unintended Movement” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the PINNACLE 2217-50-060?
None of the reports are classified as deaths, 0 as injuries and 17 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the PINNACLE 2217-50-060 FDA approved?
Devices of this type (product code HWA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).
Who makes the PINNACLE 2217-50-060, and is it still sold?
The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand PINNACLE, model 2217-50-060). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other IMPACTOR devices
- PINNACLE 2217-50-041Depuy (Ireland) · IMPACTOR4,276 reports
- ATTUNE 2544-01-006Depuy (Ireland) · IMPACTOR1 recall2,528 reports
- ATTUNE 2544-01-003Depuy (Ireland) · IMPACTOR1 recall1,491 reports
- ATTUNE 2544-01-017Depuy (Ireland) · IMPACTOR1 recall1,454 reports
- ATTUNE 2544-01-004Depuy (Ireland) · IMPACTOR1 recall873 reports
- DURALOC 2360-71-000Depuy Orthopaedics, Inc. · IMPACTOR715 reports
- SPECIALIST 2581-11-000Depuy (Ireland) · IMPACTOR703 reports
- INHANCE 6205-10-103Depuy (Ireland) · IMPACTOR334 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page