Skip to content

NA 2274-36-000Recalls & reported problems

Depuy (Ireland) · DEPTH GAUGE · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy depth gauge is a surgical instrument used to measure screw depth during bone fixation. The 31 reports mostly describe the tool breaking, bending, or coming apart during use or inspection, along with a few cases of it giving wrong measurements or not fitting with other instruments. No deaths and no injuries were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
30
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 10603295109198), received between Aug 13, 2018 and Aug 6, 2026 — classified as 30 malfunctions.

The most frequently reported specific device problem is “Break” (17 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break17
  • Material Deformation4
  • Material Twisted/Bent4
  • Detachment of Device or Device Component3
  • Device-Device Incompatibility3
  • Incorrect Measurement3
  • Defective Device1
  • Naturally Worn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • No Consequences Or Impact To Patient11
  • Insufficient Information4

Recent adverse event reports

The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50158752 ↗
  • MalfunctionAug 28, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT 25MM DRILL WAS BROKEN AND DEPTH GAUGE WAS BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22917664 ↗
  • MalfunctionAug 26, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THERE WAS A MISSING ATTACHMENT. THE ISSUE WAS IDENTIFIED DURING LOAN KIT INSPECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22893797 ↗
  • MalfunctionJun 30, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THERE IS A BENT COIL ON A DEPTH GAUGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22358206 ↗
  • MalfunctionFeb 13, 2025Break; Break

    IT WAS REPORTED THAT THE NUT THAT HOLDS THE GAUGE TOGETHER BROKE. NO SURGICAL DELAY. GAUGE BEING RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21377329 ↗
  • MalfunctionDec 6, 2024Break; Break

    IT WAS REPORTED THAT THE TIP OF DEPTH GAUGE WAS BROKE OFF AFTER MEASURING SCREW LENGTH. THERE WAS NO SURGICAL DELAY AS TIP WAS EASILY REMOVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20866054 ↗
  • MalfunctionAug 28, 2024Break; Break

    IT WAS REPORTED THAT IMPACTOR HANDLE, RESECTION WING AND DEPTH GAUGE WERE BROKEN.

    Patient problems: Insufficient Information

    MAUDE report 20085467 ↗
  • MalfunctionMay 20, 2024Incorrect Measurement; Material Deformation; Incorrect Measurement; Material Deformation

    IT WAS REPORTED THAT UPON INSPECTION, THE SHAFT OF THE DEPTH GAUGE WAS BENT AND UNABLE TO MEASURE. THIS WAS NOTICED PRIOR TO THE CASE STARTING. THE DEPTH GAUGE WAS DISPOSED OF AND NO HARM OCCURRED TO THE PATIENT AND THERE WAS NO DELAY IN SURGICAL TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19350677 ↗
  • MalfunctionFeb 21, 2024Break; Break

    IT WAS REPORTED THAT PINNACLE BONE SCREW DEPTH GAUGE BROKE DURING SURGERY SET UP. SURGICAL DELAY UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 18750753 ↗
  • MalfunctionFeb 18, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT DEPTH GAUGE IS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18728556 ↗
  • MalfunctionJan 4, 2024Break; Break

    IT WAS REPORTED THAT THE DEPTH GAUGE BROKE. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? : NO, WAS PROCEDURE SUCCESSFULLY COMPLETED? : YES, WERE FRAGMENTS GENERATED? -: NO, IF YES, WERE THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION? : UNKNOWN, PATIENT STATUS/ OUTCOME / CONSEQUENCES : NO, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN, IS THE PATIENT PART OF A CLINICAL STUDY: UNKNOWN, (B)(4). DEVICE PROPERTY OF : NONE, DEVICE IN POSSESSION OF : UVM BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST.: TRUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18444661 ↗
  • MalfunctionDec 27, 2023Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE FLEXIBLE DEPTH GAUGE WAS BENT. THERE WAS NO PATIENT CONSEQUENCE OR SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18401054 ↗
  • MalfunctionNov 28, 2023Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT WHILE MEASURING SCREW WITH DEPTH GAUGE, SPRING MALFUNCTIONED AND WOULD NOT ATTACH TO GAUGE. BROKEN PIECE DISPOSED OF. THERE WAS NO HARM TO PATIENT AND NO DELAY IN SURGICAL TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18215932 ↗
  • MalfunctionNov 1, 2022Break; Break

    IT WAS REPORTED THAT THE ACETABULAR SCREW DEPTH GAUGE BROKE. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15708294 ↗
  • MalfunctionSep 27, 2022Break; Break

    IT WAS REPORTED THAT THE DEPTH GAUGE BROKE DURING CLEANING. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15490555 ↗
  • MalfunctionJan 19, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    SPRING ON DEPTH GAUGE IS BROKEN. DID NOT OCCUR DURING SURGERY SO NO SURGICAL DELAY OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13289000 ↗
  • MalfunctionJan 6, 2022Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn

    IT WAS REPORTED THAT THE THREADED PIECE OF DEPTH GAUGE WAS CROSS THREADED AND WOULD NOT SCREW DOWN. SURGICAL DELAY 2MINS.

    Patient problems: Insufficient Information

    MAUDE report 13174638 ↗
  • MalfunctionApr 29, 2021Incorrect Measurement; Incorrect Measurement

    IT WAS REPORTED THAT THE RIBBED COLLAR ON THE DEPTH GAUGE DOES NOT ¿CATCH¿ OR STOP AT THE PORTION THAT IS INSERTED INTO THE SCREW HOLE. NO DELAY TO SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11743851 ↗
  • MalfunctionMar 29, 2021Break; Break

    IT WAS REPORTED THAT THE DEPTH GAUGE SPRING BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11585648 ↗
  • MalfunctionAug 25, 2020Break; Break

    IT WAS REPORTED THAT THE TIP OF THE DEPTH GAUGE HAS BROKEN OFF. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10449434 ↗
  • MalfunctionAug 3, 2020Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT UPON REASSEMBLY AFTER SURGERY, DEPTH GAUGE BENT AND UNUSABLE. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10360212 ↗
  • MalfunctionOct 25, 2019Break; Break

    TIP BROKEN OFF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9240193 ↗
  • MalfunctionOct 7, 2019Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT THE DEPTH GAUGE BENT AND BALL BEARING WON'T HOLD IN PLACE. FOUND WHEN SET RETURNED TO THE OFFICE FOR INSPECTION. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9161887 ↗
  • MalfunctionAug 21, 2019Break; Material Twisted/Bent; Break; Material Twisted/Bent

    IT WAS REPORTED THAT THE DEPTH GAUGE WAS BROKEN WHEN TRAY WAS OPENED AT THE BEGINNING OF CASE. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8915287 ↗
  • MalfunctionJul 17, 2019Break; Break

    IT WAS REPORTED THAT THE DEPTH GAUGE BROKE INTO TWO PIECES AS IT WAS BEING PULLED OUT OF THE ACETABULUM AFTER IT HAD BEEN USED TO GAUGE THE SIZE OF THE FIRST SCREW. ALL PIECES (TWO TOTAL PIECES) HAVE BEEN RETRIEVED AND NO ADVERSE AFFECTS HAD HAPPENED AS A RESULT. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8799685 ↗

Other GAUGE, DEPTH devices

All GAUGE, DEPTH devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page