DELTA XTEND 2307-82-003Recalls & reported problems
Depuy (Ireland) · DELTA Xtend DELTA CTA HEAD REAMER · Class 1
All models of DELTA XTEND (194)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10603295116356), received between Aug 18, 2017 and May 20, 2022 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Naturally Worn” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Naturally Worn4
- Device-Device Incompatibility3
- Adverse Event Without Identified Device or Use Problem1
- Connection Problem1
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Bone Fracture(s)1
- No Consequences Or Impact To Patient1
- No Information1
- Not Applicable1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 20, 2022Naturally Worn; Naturally Worn
IT WAS REPORTED THAT INSTRUMENTS WERE HIGHLY WORN AND NEEDED TO BE REPLACED. THERE WAS NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14462793 ↗MalfunctionMay 20, 2022Naturally Worn; Naturally Worn
IT WAS REPORTED THAT INSTRUMENTS WERE HIGHLY WORN AND NEEDED TO BE REPLACED. THERE WAS NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14462639 ↗MalfunctionJun 28, 2021Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE CTA REAMER WAS WELDED ONTO THE REAMER GUIDE. SURGERY WAS NOT EXTENDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12080834 ↗MalfunctionApr 8, 2021Device-Device Incompatibility; Device-Device Incompatibility
SURGEON HAD CONVERTED TO EXTEND CTA DUE TO POOR GLENOID BONE QUALITY. WHEN HE ATTEMPTED TO PUT DELTA EXTEND CTA HEAD REAMER INTO THE REAMER GUIDE IT WOULD NOT FIT IN ALL THE WAY. SURGEON ATTEMPTED TO BEGIN REAMING AND IT STUCK AND ROTATED THE WHOLE CONSTRUCT, STEM, HEAD REAMER GUIDE AN REAMER. SURGICAL DELAY 10 MINUTES. SURGEON COMPLETED BONE REMOVAL WITH RONGEUR, PROCEDURE SUCCESSFULLY COMPLETED. NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11633357 ↗MalfunctionOct 9, 2019Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE LINER TRIAL WAS CRACKED AND WON¿T TRIAL THE HEAD CORRECTLY DUE TO CRACKED. THE CTA GUIDE AND THE REAMER HAVE BURRS INTERNALLY WHERE THEY MEET TOGETHER AND THEY BIND UP WHEN PUT TOGETHER. TOLERANCES ARE NOT CORRECT. NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9171980 ↗InjuryJan 16, 2019Device-Device Incompatibility; Naturally Worn; Adverse Event Without Identified Device or Use Problem; Device-Device Incompatibility; Naturally Worn; Adverse Event Without Identified Device or Use Problem
DURING PREPARATION OF THE PROXIMAL HUMERUS FOR THE CTA HEAD THE REAMER GUIDE WAS MOUNTED ON TO THE HUMERAL STEM. WHILE REAMING WITH THE CTA-HEAD-REAMER, THE REAMER SUDDENLY BLOCKED IN THE REAMER-GUIDE. THIS TURNED PRESSURE ON THE HUMERAL STEM AND GENERATED A FRACTURE OF THE PROXIMAL HUMERUS. THIS FRACTURE HAD TO BE FIXED WITH SUTURES. THE OPERATION COULD BE COMPLETED.
Patient problems: Bone Fracture(s); Not Applicable
MAUDE report 8252956 ↗MalfunctionAug 18, 2017Connection Problem; Mechanical Jam; Connection Problem; Mechanical Jam
IT WAS REPORTED THAT THE SURGEON ATTEMPTED TO IMPLANT AN XTEND CTA HEAD. WHEN REAMING THE LATERAL PROXIMAL HUMERUS, THE REAMER GOT STUCK ON THE GUIDE AND COULD NOT BE REMOVED.
Patient problems: No Information
MAUDE report 6805001 ↗
Other REAMER devices
- QUICKSET 2440-00-551Depuy (Ireland) · REAMER1,150 reports
- QUICKSET 2440-00-549Depuy (Ireland) · REAMER1,113 reports
- QUICKSET 2440-00-553Depuy (Ireland) · REAMER1,105 reports
- QUICKSET 2440-00-547Depuy (Ireland) · REAMER1,073 reports
- QUICKSET 2440-00-555Depuy (Ireland) · REAMER1,065 reports
- QUICKSET 2440-00-557Depuy (Ireland) · REAMER1,031 reports
- QUICKSET 2440-00-545Depuy (Ireland) · REAMER947 reports
- QUICKSET 2440-00-552Depuy (Ireland) · REAMER888 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page