Skip to content

DELTA XTEND 2307-91-001Recalls & reported problems

Depuy (Ireland) · DELTA Xtend SCREW DEPTH GAUGE · Class 1

All models of DELTA XTEND (194)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Delta Xtend depth gauge is a measuring tool used to check screw depth during orthopedic surgery. Reports mostly describe the plastic body snapping or coming apart, sometimes found bent or broken on arrival or after sterilization cycles, with pieces going missing. There are 54 reports, no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
54
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 10603295116554), received between Oct 9, 2017 and Dec 30, 2025 — classified as 54 malfunctions.

The most frequently reported specific device problem is “Break” (31 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 3 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break31
  • Detachment of Device or Device Component6
  • Material Deformation6
  • Bent3
  • Incorrect Measurement2
  • Naturally Worn2
  • Adverse Event Without Identified Device or Use Problem1
  • Component Missing1

Patient problems

  • No Clinical Signs, Symptoms or Conditions36
  • No Consequences Or Impact To Patient12
  • Insufficient Information4
  • No Information2

Recent adverse event reports

The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 30, 2025Break; Break

    IT WAS REPORTED THAT THE DEPTH GAUGE FELL APART, THERE WAS NO DELAY IN THE CASE. EVERY THING WAS FINE, NO PARTS WERE LOST. NO ISSUES WITH THE CASE NO ISSUES WITH THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23927037 ↗
  • MalfunctionDec 12, 2025Material Deformation; Material Deformation

    IT WAS REPORTED CSSD ALERTED THAT DEPTH GAUGE ON CONSIGNMENT IS BENT. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23791330 ↗
  • MalfunctionOct 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THERE WAS A MISSING SCREW ON A 4 PIECE SCREWDRIVER. DEVICE WAS FOUND IN STERILE PROCESSING. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23227603 ↗
  • MalfunctionJun 24, 2025Break; Break

    IT WAS REPORTED THAT THE DEPTH GAUGE ON THE SET WAS BROKEN UPON ARRIVAL. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22318010 ↗
  • MalfunctionMar 27, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025. WHEN OPENING THE SETS PRIOR TO SURGERY, IT WAS DETERMINED THAT THE DEPTH GAUGE AND VERSION ROD WERE BOTH BENT.

    Patient problems: Insufficient Information

    MAUDE report 21708054 ↗
  • MalfunctionNov 21, 2024Break; Break

    IT WAS REPORTED THE DEPTH GAUGE WAS BROKE; MAIN PLASTIC BODY OF DEPTH GAUGE SNAPPED. BREAKAGE DID NOT HAPPEN IN SURGERY. MAIN PLASTIC BODY OF DEVICE FRAGILE/BRITTLE POSSIBLY FROM LONG TERM EXPOSURE TO HEAT CYCLES FROM STERILIZATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20747030 ↗
  • MalfunctionNov 19, 2024Break; Break

    IT WAS REPORTED THAT DURING A PRIMARY REVERSE TOTAL SHOULDER ARTHROPLASTY ON (B)(6) 2024, THE SURGEON NOTED THAT THE METAL SLEEVE OF THE SCREW DEPTH GAUGE WAS MISSING. THE PIECE WAS NOT PRESENT DURING THE SURGERY. THERE WERE NO COMPLICATIONS OR BROKEN PIECES IN THE SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 20723435 ↗
  • MalfunctionOct 17, 2024Break; Break

    IT WAS REPORTED THAT DUE TO A BROKEN DEPTH GAGE THIS CAUSED ANOTHER HOSPITALS TRAY TO BE BORROWED SURGERY WAS NOT DELAYED DUE TO THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20470410 ↗
  • MalfunctionSep 20, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE INSTRUMENT IS MISSING THE BLACK CAP OFF THE END OF IT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20266047 ↗
  • MalfunctionJul 5, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    DEPTH GAUGE IS MISSING A COMPONENT. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT DATE AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19681426 ↗
  • MalfunctionJul 1, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE DEPTH GAUGE IS MISSING A COMPONENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19643136 ↗
  • MalfunctionJun 19, 2024Break; Break

    IT WAS REPORTED THAT THE GAUGE DOES NOT RUN SMOOTH AND HAS A BROKEN TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19566997 ↗
  • MalfunctionApr 22, 2024Break; Break

    IT WAS REPORTED THAT THE DEPTH GAUGE BROKE. NO PIECES WERE LEFT INSIDE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 19152549 ↗
  • MalfunctionApr 17, 2024Break; Break

    IT WAS REPORTED THAT THE DEPTH GAUGE BROKE. NO PIECES WERE LEFT INSIDE PATIENT. UNKNOWN SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19127236 ↗
  • MalfunctionApr 16, 2024Incorrect Measurement; Incorrect Measurement

    IT WAS REPORTED THAT THE DEPTH GAUGE WAS NOT READING ACCURATE. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19116452 ↗
  • MalfunctionMar 28, 2024Material Deformation; Material Deformation

    ADDITIONAL INFORMATION WAS RECEIVED AND IT WAS STATED THAT THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18995248 ↗
  • MalfunctionJan 31, 2024Break; Break

    IT WAS REPORTED THAT THE DEPTH GAUGE BROKE DURING USE. NO DELAY REPORTED. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? UNKNOWN, WAS PROCEDURE SUCCESSFULLY COMPLETED? UNKNOWN, WERE FRAGMENTS GENERATED? UNKNOWN, IF YES, WERE THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION? UNKNOWN, PATIENT STATUS/ OUTCOME / CONSEQUENCES NO, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN, IS THE PATIENT PART OF A CLINICAL STUDY UNKNOWN, (B)(4) DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST. TRUE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18614382 ↗
  • MalfunctionDec 8, 2023Break; Break

    IT WAS REPORTED THAT THE DEPTH GAUGE BROKE DURING SURGERY. ALL PIECES WERE RETRIEVED FROM PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 18289994 ↗
  • MalfunctionOct 23, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE GAUGED IS MISSING A COMPONENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17984799 ↗
  • MalfunctionAug 28, 2023No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE (B)(6) TRAY HAD BEEN DELIVERED TO THE HOSPITAL MISSING AN ITEM (230784003) AND WITH ONE ITEM INCOMPLETE (230791001).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17631813 ↗
  • MalfunctionAug 8, 2023Break; Break

    IT WAS REPORTED THAT THE BLACK TIP BROKE AND DOES NOT FIT IN THE DEPTH GAUGE ANYMORE. NO PIECES WERE LEFT IN PATIENT. IT WAS UNKNOWN IF THERE WAS ANY SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17479193 ↗
  • MalfunctionJul 12, 2023Break; Break

    IT WAS REPORTED THAT THE DEPTH GAUGE WAS BROKEN IN SPD IN THE CENTER WHERE SCREW LENGTH IS READ.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17313345 ↗
  • MalfunctionJun 20, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE BLACK KNOB WAS MISSING FROM DEPTH GAUGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17164008 ↗
  • MalfunctionMay 5, 2023Material Deformation; Naturally Worn; Material Deformation; Naturally Worn

    IT WAS REPORTED THAT DEPTH GAUGE WAS BENT AND THE BLACK PIECE ON BOTTOM WAS FRAYED AND CROSS THREADED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16876362 ↗
  • MalfunctionMar 24, 2023Break; Break

    IT WAS REPORTED THAT DEPTH GAUGE WAS MISSING GAUGE BLACK TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16612305 ↗

Other GAUGE, DEPTH devices

All GAUGE, DEPTH devices →