Skip to content

NA 2421-04-000Recalls & reported problems

Depuy (Ireland) · THIN OSTEOTOME 12.5mm · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10603295122746), received between Oct 13, 2017 and Oct 21, 2025 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Crack2
  • Bent1
  • Dull, Blunt1
  • Fracture1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient4
  • No Information2

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 21, 2025Break; Break

    IT WAS REPORTED THAT PATIENT UNDERWENT TOTAL KNEE ARTHROPLASTY (TKA). THE CEMENT OSTEOTOME HAD BROKEN DURING THE KNEE EXTRACTION. ALL BROKEN PIECES WERE IDENTIFIED AND ACCOUNTED FOR. IT WAS STATED THAT THIS WAS "THE FIRST USE OF THIS INSTRUMENT." THE PRODUCT WAS DISPOSED OF IN THE SHARPS CONTAINER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23349125 ↗
  • MalfunctionMay 2, 2025Break; Crack; Break; Crack

    IT WAS REPORTED THAT THE DEVICES WERE CHIPPED AND CRACKED. IT DIDN'T HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21954192 ↗
  • MalfunctionMar 21, 2023Break; Break

    IT WAS REPORTED THAT THE TIP BROKE OFF OF THE CHISEL. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16581880 ↗
  • MalfunctionFeb 28, 2022Break; Break

    IT WAS REPORTED THAT THE FEMORAL INTRODUCER RED KNOB BROKE OFF. TIP OF CHISEL BROKE OFF. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13620225 ↗
  • MalfunctionFeb 10, 2020Crack; Crack

    PHENOLIC HANDLE CRACKED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9689235 ↗
  • MalfunctionDec 7, 2018Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE OSTEOTOME WAS CHIPPED DURING CEMENT REMOVAL. UPON INSPECTION OF TORQUE WRENCH, THE BLACK PIECE CAME LOOSE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8142720 ↗
  • MalfunctionJul 2, 2018Bent; Dull, Blunt; Bent; Dull, Blunt

    IT WAS REPORTED THAT THE INSTRUMENT IS BENT AND DULL. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7655047 ↗
  • MalfunctionJun 29, 2018Break; Break

    IT WAS REPORTED THAT THE TIPS OF THE INSTRUMENTS ARE BROKEN OFF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7650299 ↗
  • MalfunctionNov 14, 2017Fracture; Fracture

    IT WAS REPORTED THAT THE KNEE EXTRACTION INSTRUMENTS WERE EITHER CHIPPED OR DULL AND THE SURGEON ASKED FOR THEM TO BE REPLACED.

    Patient problems: No Information

    MAUDE report 7030484 ↗
  • MalfunctionOct 13, 2017Break; Break

    IT WAS REPORTED THAT THE MOORELAND HANDLE WAS BROKE DURING EXPLANTATION.

    Patient problems: No Information

    MAUDE report 6947988 ↗

Other OSTEOTOME devices

All OSTEOTOME devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page