Skip to content

MORELAND 2422-14-000Recalls & reported problems

Depuy Orthopaedics, Inc. · MORELAND REV INST CUP CHISEL 64mm 80 DEGREES · Class 1

All models of MORELAND (22)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10603295123118), received between Dec 14, 2017 and Dec 14, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 14, 2017Break; Break

    IT WAS REPORTED THAT BOTH ACETABULAR CUP OSTEOTOMES BROKE DURING USE AND TIPS AND STEM IMPACTORS WERE ALSO BROKEN DURING IMPACTION. NO SURGICAL DELAY. ALL PIECES WERE RETRIEVED FROM THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7118149 ↗

Other OSTEOTOME devices

All OSTEOTOME devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page