QUICKSET 2440-00-538Recalls & reported problems

Depuy (Ireland) · QUICKSET GRATER HEAD 38mm · Class 1

All models of QUICKSET (66)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy QuickSet grater head is a surgical reamer attachment. Many reports describe graters that became dull or failed to cut, and some mention corrosion, usually noticed outside of surgery. The 157 reports include no deaths and no injuries. No recalls name this device, and reports fell from 13 in the prior 12 months to none in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
157
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 157 adverse event reports that list this exact device (DI 10603295123859), received between Sep 20, 2017 and Jul 28, 2025 — classified as 157 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (85 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 5 reportsFeb 2024: 5 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 3 reportsJun 2024: 3 reportsJul 2024: 4 reportsAug 2024: 0 reportsSep 2024: 4 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 8 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 4 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Cut85
  • Corroded29
  • Dull, Blunt18
  • No Apparent Adverse Event16
  • Naturally Worn8
  • Crack4
  • Material Twisted/Bent4
  • Adverse Event Without Identified Device or Use Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions119
  • No Consequences Or Impact To Patient31
  • Insufficient Information8
  • No Information2
  • No Code Available1
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 157 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 28, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS ARE DULL AND NEED REPLACED (WITH CROSSBACK GRATERS). IT DIDN'T HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22631473 ↗
  • MalfunctionJul 24, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. THIS EVENT DID NOT HAPPEN IN SURGERY. (B)(4) ARE RELATED TO EACH OTHER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22587789 ↗
  • MalfunctionJul 24, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. THIS EVENT DID NOT HAPPEN IN SURGERY. (B)(4) ARE RELATED TO EACH OTHER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22587609 ↗
  • MalfunctionJul 7, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT ITEMS NEED REPLACED WITH CROSS BACKS. A PRODUCT HAS BEEN REPORTED BUT THE REASON FOR REPORTING IS UNKNOWN. THIS EVENT DID NOT HAPPEN IN SURGERY. DURING MANUFACTURER'S INVESTIGATION OF THE RETURNED DEVICE IT WAS FOUND THAT THE CUTTING EDGES OF THE DEVICE ARE BLUNTING/ROUNDING. ADDITIONALLY, SIGNS OF CORROSION WERE OBSERVED AROUND THE CUTTING SURFACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22413996 ↗
  • MalfunctionApr 17, 2025Break; Crack; Material Deformation; Break; Crack; Material Deformation

    IT WAS REPORTED THAT THE TECH NOTICED ALL 3 GRATERS HAD THE GRATER CUT OUT BENT INWARD AND WERE DULL, THE OTHER LOOKED TO HAVE A SMALL CRACK ON THE CROSS SECTION ALONG THE BACKSIDE THAT CONNECTS TO HANDLE. THE GRATERS WERE PULLED FROM THAT TRAY. NO ISSUES WITH DELAYS OR ISSUES WITH THE GRATERS USED FOR THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21855131 ↗
  • MalfunctionMar 24, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER REPEATED USE AND TIME. THE EVENT DID NOT OCCUR DURING A SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21677710 ↗
  • MalfunctionMar 24, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER REPEATED USE AND TIME. THE EVENT DID NOT OCCUR DURING A SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21677254 ↗
  • MalfunctionMar 7, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT GRATERS HAVE BECOME DULL OVER TIME WITH REPEATED USE. THE ISSUE DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21545785 ↗
  • MalfunctionMar 7, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT GRATERS HAVE BECOME DULL OVER TIME WITH REPEATED USE. THE ISSUE DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21545602 ↗
  • MalfunctionMar 6, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT GRATERS HAVE BECOME DULL OVER TIME WITH REPEATED USE. THE ISSUE DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21537482 ↗
  • MalfunctionMar 6, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT GRATERS HAVE BECOME DULL OVER TIME WITH REPEATED USE. THE ISSUE DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21537031 ↗
  • MalfunctionMar 5, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER TIME AND NEEDS TO BE REPLACED. THE EVENT DID NOT HAPPENED DURING A SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21525060 ↗
  • MalfunctionMar 5, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER TIME AND NEEDS TO BE REPLACED. THE EVENT DID NOT HAPPENED DURING A SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21524980 ↗
  • MalfunctionSep 6, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE REAMERS DULL - NEED REPLACEMENTS.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20156540 ↗
  • MalfunctionSep 4, 2024Corroded; Corroded

    IT WAS REPORTED THAT THE GRATERS ARE RUSTY SO SPD WILL NOT STERILIZE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20134374 ↗
  • MalfunctionSep 3, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL AND NEED REPLACEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20125980 ↗
  • MalfunctionSep 3, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL AND NEED REPLACEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20125975 ↗
  • MalfunctionJul 19, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE PRODUCTS WERE DULL.

    Patient problems: Insufficient Information

    MAUDE report 19781532 ↗
  • MalfunctionJul 19, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE INSTRUMENTS WERE DULL.

    Patient problems: Insufficient Information

    MAUDE report 19781498 ↗
  • MalfunctionJul 18, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE REAMERS WERE DULL. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19775514 ↗
  • MalfunctionJul 9, 2024Failure to Cut; Naturally Worn; Failure to Cut; Naturally Worn

    STAFF HAVE INDICATED THAT THE GRADERS REQUIRE A REPLACEMENT DUE TO WEAR AND TEAR AND BEING DULL OVERTIME. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19702935 ↗
  • MalfunctionJun 10, 2024No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE CROSSBACK GRATER WAS REPORTED FOR UNKNOWN REASON. DID NOT HAPPEN IN SURGERY. ADDITIONAL EVENT INFORMATION: VISUAL ANALYSIS OF THE DEVICE REVEALED THAT THE CUTTING EDGED PRESENTS SIGNS OF WEAR CONSISTENT WITH REPEATED USE AND SERVICING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19499991 ↗
  • MalfunctionJun 10, 2024No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE CROSSBACK GRATER WAS REPORTED FOR UNKNOWN REASON. DID NOT HAPPEN IN SURGERY. ADDITIONAL EVENT INFORMATION: VISUAL ANALYSIS OF THE DEVICE REVEALED THAT THE CUTTING EDGED PRESENTS SIGNS OF WEAR CONSISTENT WITH REPEATED USE AND SERVICING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19499976 ↗
  • MalfunctionJun 10, 2024No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE CROSSBACK GRATER WAS REPORTED FOR UNKNOWN REASON. DID NOT HAPPEN IN SURGERY. ADDITIONAL EVENT INFORMATION: VISUAL ANALYSIS OF THE DEVICE REVEALED THAT THE CUTTING EDGED PRESENTS SIGNS OF WEAR CONSISTENT WITH REPEATED USE AND SERVICING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19499956 ↗
  • MalfunctionMay 10, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT REAMERS WERE DULL. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19286723 ↗

Other REAMER devices

All REAMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page