QUICKSET 2440-00-546Recalls & reported problems

Depuy (Ireland) · QUICKSET GRATER HEAD 46mm · Class 1

All models of QUICKSET (66)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy QuickSet grater head reamer has 755 reports, with no deaths and no injuries. Most reports describe the reamer failing to cut, and some say it was too dull or chipped to use when staff cleaned it. Reports rose to 26 in the past 12 months from 17 in the 12 months before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
754
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 754 adverse event reports that list this exact device (DI 10603295123934), received between Aug 16, 2017 and Apr 6, 2026 — classified as 754 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (523 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 8 reportsDec 2023: 5 reportsJan 2024: 9 reportsFeb 2024: 32 reportsMar 2024: 13 reportsApr 2024: 20 reportsMay 2024: 31 reportsJun 2024: 7 reportsJul 2024: 11 reportsAug 2024: 3 reportsSep 2024: 11 reportsOct 2024: 5 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 10 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 8 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 15 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202632
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Cut523
  • Dull, Blunt122
  • Naturally Worn82
  • Corroded47
  • No Apparent Adverse Event36
  • Break27
  • Material Twisted/Bent12
  • Material Deformation11

Patient problems

  • No Clinical Signs, Symptoms or Conditions554
  • No Consequences Or Impact To Patient145
  • Insufficient Information47
  • No Information22
  • No Code Available8
  • Not Applicable5
  • No Known Impact Or Consequence To Patient2
  • Unspecified Tissue Injury1

Recent adverse event reports

The 25 most recent of 754 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24793692 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24792432 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24792186 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24792184 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24792177 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791949 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791948 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791934 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791929 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791927 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791916 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791912 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791889 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791871 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING THE STAFF NOTICED THAT THESE GRATERS ARE TOO DULL AND CHIPPED TO BE USED ON A PATIENT. THE SUMMIT BROACH HANDLE CLASPS NO LONGER TIGHTEN SUFFICIENTLY ON THE BROACHES. THE EXCEL T HANDLES ARE CRACKED ON THE HANDLE. THE CUP INSURERS ARE CROSS THREADED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791859 ↗
  • MalfunctionFeb 2, 2026Failure to Cut; No Apparent Adverse Event

    IT WAS REPORTED THAT SURGEON HAS COMPLAINED THAT THE ACETABULAR REAMERS WERE NOT SHARP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24233829 ↗
  • MalfunctionFeb 2, 2026Failure to Cut; No Apparent Adverse Event

    IT WAS REPORTED THAT SURGEON HAS COMPLAINED THAT THE ACETABULAR REAMERS WERE NOT SHARP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24233468 ↗
  • MalfunctionJan 29, 2026Failure to Cut; No Apparent Adverse Event

    IT WAS REPORTED THAT SURGEON HAS COMPLAINED ACETABULAR REAMERS WERE NOT SHARP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24207103 ↗
  • MalfunctionJan 29, 2026Failure to Cut; No Apparent Adverse Event

    IT WAS REPORTED THAT SURGEON HAS COMPLAINED ACETABULAR REAMERS WERE NOT SHARP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24207100 ↗
  • MalfunctionJan 29, 2026Failure to Cut; No Apparent Adverse Event

    IT WAS REPORTED THAT SURGEON HAS COMPLAINED ACETABULAR REAMERS WERE NOT SHARP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24207099 ↗
  • MalfunctionJan 28, 2026Failure to Cut

    IT WAS REPORTED THAT THE ACETABULAR GRATERS IN THE HOSPITALS SETS ARE IN NEED OF SHARPENING AND THEY ARE NO LONGER FUNCTIONING AS EXPECTED. NO KNOWN PATIENT HARM. THE FOLLOWING COMPLAINTS ARE LINKED TO CAPTURE ALL IMPACTED PRODUCTS: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24202193 ↗
  • MalfunctionJan 28, 2026Failure to Cut

    IT WAS REPORTED THAT THE ACETABULAR GRATERS IN THE HOSPITALS SETS ARE IN NEED OF SHARPENING AND THEY ARE NO LONGER FUNCTIONING AS EXPECTED. NO KNOWN PATIENT HARM. THE FOLLOWING COMPLAINTS ARE LINKED TO CAPTURE ALL IMPACTED PRODUCTS: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24198375 ↗
  • MalfunctionJan 28, 2026Failure to Cut

    IT WAS REPORTED THAT THE ACETABULAR GRATERS IN THE HOSPITALS SETS ARE IN NEED OF SHARPENING AND THEY ARE NO LONGER FUNCTIONING AS EXPECTED. NO KNOWN PATIENT HARM. THE FOLLOWING COMPLAINTS ARE LINKED TO CAPTURE ALL IMPACTED PRODUCTS: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24198370 ↗
  • MalfunctionJan 28, 2026Failure to Cut

    IT WAS REPORTED THAT THE ACETABULAR GRATERS IN THE HOSPITALS SETS ARE IN NEED OF SHARPENING AND THEY ARE NO LONGER FUNCTIONING AS EXPECTED. NO KNOWN PATIENT HARM. THE FOLLOWING COMPLAINTS ARE LINKED TO CAPTURE ALL IMPACTED PRODUCTS: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24198368 ↗
  • MalfunctionJan 28, 2026Failure to Cut

    IT WAS REPORTED THAT THE ACETABULAR GRATERS IN THE HOSPITALS SETS ARE IN NEED OF SHARPENING AND THEY ARE NO LONGER FUNCTIONING AS EXPECTED. NO KNOWN PATIENT HARM THE FOLLOWING COMPLAINTS ARE LINKED TO CAPTURE ALL IMPACTED PRODUCTS: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24198362 ↗