QUICKSET 2440-00-562Recalls & reported problems

Depuy (Ireland) · QUICKSET GRATER HEAD 62mm · Class 1

All models of QUICKSET (66)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy QuickSet grater head appears in 268 adverse event reports, with no deaths and no injuries. The reports mostly describe graters that have become dull, fail to cut, or show rust and warping, often on older units, and say the problem did not happen during surgery. No reports were filed in the last 12 months, down from 26 in the year before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
268
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 268 adverse event reports that list this exact device (DI 10603295124092), received between Sep 18, 2017 and Jul 15, 2025 — classified as 268 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (177 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 7 reportsDec 2023: 1 reportsJan 2024: 8 reportsFeb 2024: 11 reportsMar 2024: 3 reportsApr 2024: 0 reportsMay 2024: 8 reportsJun 2024: 3 reportsJul 2024: 7 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 4 reportsFeb 2025: 1 reportsMar 2025: 7 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 3 reportsJul 2025: 8 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Cut177
  • Corroded60
  • No Apparent Adverse Event25
  • Dull, Blunt15
  • Naturally Worn13
  • Material Deformation7
  • Device-Device Incompatibility5
  • Adverse Event Without Identified Device or Use Problem4

Patient problems

  • No Clinical Signs, Symptoms or Conditions205
  • No Consequences Or Impact To Patient38
  • Insufficient Information22
  • No Information3
  • No Code Available2
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 268 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 15, 2025Corroded; Material Deformation; Corroded; Material Deformation

    IT WAS REPORTED THAT OLD PINNACLE GRATERS NEEDED TO BE REPLACED. DOCTORS COMPLAINED THAT THEY WERE VERY DULL AND STERILE PROCESSING COMPLAINED OF RUST AND WARPING. THESE ARE OVER 10 YEARS OLD. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22509638 ↗
  • MalfunctionJul 15, 2025Corroded; Material Deformation; Corroded; Material Deformation

    IT WAS REPORTED THAT OLD PINNACLE GRATERS NEEDED TO BE REPLACED. DOCTORS COMPLAINED THAT THEY WERE VERY DULL AND STERILE PROCESSING COMPLAINED OF RUST AND WARPING. THESE ARE OVER 10 YEARS OLD. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22509637 ↗
  • MalfunctionJul 15, 2025Corroded; Material Deformation; Corroded; Material Deformation

    IT WAS REPORTED THAT OLD PINNACLE GRATERS NEEDED TO BE REPLACED. DOCTORS COMPLAINED THAT THEY WERE VERY DULL AND STERILE PROCESSING COMPLAINED OF RUST AND WARPING. THESE ARE OVER 10 YEARS OLD. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22509635 ↗
  • MalfunctionJul 15, 2025Corroded; Material Deformation; Corroded; Material Deformation

    IT WAS REPORTED THAT OLD PINNACLE GRATERS NEEDED TO BE REPLACED. DOCTORS COMPLAINED THAT THEY WERE VERY DULL AND STERILE PROCESSING COMPLAINED OF RUST AND WARPING. THESE ARE OVER 10 YEARS OLD. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22509631 ↗
  • MalfunctionJul 15, 2025Corroded; Material Deformation; Corroded; Material Deformation

    IT WAS REPORTED THAT OLD PINNACLE GRATERS NEEDED TO BE REPLACED. DOCTORS COMPLAINED THAT THEY WERE VERY DULL AND STERILE PROCESSING COMPLAINED OF RUST AND WARPING. THESE ARE OVER 10 YEARS OLD. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22508101 ↗
  • MalfunctionJul 15, 2025Corroded; Material Deformation; Corroded; Material Deformation

    IT WAS REPORTED THAT OLD PINNACLE GRATERS NEEDED TO BE REPLACED. DOCTORS COMPLAINED THAT THEY WERE VERY DULL AND STERILE PROCESSING COMPLAINED OF RUST AND WARPING. THESE ARE OVER 10 YEARS OLD. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22508095 ↗
  • MalfunctionJul 8, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAD BECOME DULL AND NEED TO BE REPLACED. THIS EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22438453 ↗
  • MalfunctionJul 8, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT GRATERS HAD BECOME DULL AND NEED TO BE REPLACED. THIS EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22437668 ↗
  • MalfunctionJun 11, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS WERE DULL. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22184266 ↗
  • MalfunctionJun 11, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS WERE DULL. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22183214 ↗
  • MalfunctionJun 3, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE INSTRUMENTS HAD AN UNKNOWN ALLEGATION. THE EVENT DID NOT HAPPEN IN SURGERY. UPON MANUFACTURERS INVESTIGATION IT WAS NOTED THAT THE INSTRUMENT HAS CORROSION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22131299 ↗
  • MalfunctionMar 24, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER REPEATED USE AND TIME. THE EVENT DID NOT OCCUR DURING A SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21678953 ↗
  • MalfunctionMar 24, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER REPEATED USE AND TIME. THE EVENT DID NOT OCCUR DURING A SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21678407 ↗
  • MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER TIME AND NEED REPLACEMENT. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21517331 ↗
  • MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER TIME AND NEED REPLACEMENT. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21514977 ↗
  • MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER TIME AND NEED REPLACEMENT. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21514660 ↗
  • MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER TIME AND NEED REPLACEMENT. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21514595 ↗
  • MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER TIME AND NEED REPLACEMENT. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21514285 ↗
  • MalfunctionFeb 24, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT GRATERS WERE DULL. THE ISSUE DID NOT HAPPEN IN SURGERY. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED FOR REPORTING. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21453803 ↗
  • MalfunctionJan 14, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE INSTRUMENT WAS REPORTED FOR AN UNKNOWN REASON. IT DID NOT HAPPEN IN SURGERY. DURING MANUFACTURER'S INVESTIGATION OF THE RETURNED DEVICE IT WAS IDENTIFIED THAT THE DEVICE HAS ROUNDING OF CUTTING EDGES AND CORROSION PATTERNS ON BOTH INSIDE AND OUTSIDE OF THE CUTTING SURFACE.

    Patient problems: Insufficient Information

    MAUDE report 21152354 ↗
  • MalfunctionJan 14, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE INSTRUMENT WAS REPORTED FOR AN UNKNOWN REASON. IT DID NOT HAPPEN IN SURGERY. DURING MANUFACTURER'S INVESTIGATION OF THE RETURNED DEVICE IT WAS IDENTIFIED THAT THE DEVICE HAS ROUNDING OF CUTTING EDGES AND CORROSION PATTERNS ON BOTH INSIDE AND OUTSIDE OF THE CUTTING SURFACE.

    Patient problems: Insufficient Information

    MAUDE report 21152339 ↗
  • MalfunctionJan 14, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE INSTRUMENT WAS REPORTED FOR AN UNKNOWN REASON. IT DID NOT HAPPEN IN SURGERY. DURING MANUFACTURER'S INVESTIGATION OF THE RETURNED DEVICE IT WAS IDENTIFIED THAT THE DEVICE HAS ROUNDING OF CUTTING EDGES AND CORROSION PATTERNS ON BOTH INSIDE AND OUTSIDE OF THE CUTTING SURFACE.

    Patient problems: Insufficient Information

    MAUDE report 21152326 ↗
  • MalfunctionJan 14, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE INSTRUMENT WAS REPORTED FOR AN UNKNOWN REASON. IT DID NOT HAPPEN IN SURGERY. DURING MANUFACTURER'S INVESTIGATION OF THE RETURNED DEVICE IT WAS IDENTIFIED THAT THE DEVICE HAS ROUNDING OF CUTTING EDGES AND CORROSION PATTERNS ON BOTH INSIDE AND OUTSIDE OF THE CUTTING SURFACE.

    Patient problems: Insufficient Information

    MAUDE report 21152317 ↗
  • MalfunctionNov 13, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THERE WERE DULL ACETABULAR GRATERS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO PATIENT CONSEQUENCE. THERE WAS NO SURGICAL DELAY DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20677883 ↗
  • MalfunctionNov 13, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THERE WERE DULL ACETABULAR GRATERS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO PATIENT CONSEQUENCE. THERE WAS NO SURGICAL DELAY DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20677772 ↗