Skip to content

QUICKSET 2440-00-567Recalls & reported problems

Depuy (Ireland) · QUICKSET GRATER HEAD 67mm · Class 1

All models of QUICKSET (66)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Quickset grater head, a bone reamer, has 120 adverse event reports, with no injuries and no deaths. Many describe the grater failing to cut, becoming dull or corroded over time. Reports fell from 21 in the prior 12 months to 0 in the last 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
120
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 120 adverse event reports that list this exact device (DI 10603295124146), received between Sep 20, 2017 and Aug 4, 2025 — classified as 120 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (63 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 6 reportsFeb 2024: 7 reportsMar 2024: 3 reportsApr 2024: 0 reportsMay 2024: 7 reportsJun 2024: 3 reportsJul 2024: 4 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 8 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 8 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Cut63
  • Corroded27
  • No Apparent Adverse Event21
  • Dull, Blunt7
  • Naturally Worn7
  • Adverse Event Without Identified Device or Use Problem5
  • Material Twisted/Bent4
  • Material Discolored2

Patient problems

  • No Clinical Signs, Symptoms or Conditions101
  • No Consequences Or Impact To Patient14
  • Insufficient Information3
  • No Information2
  • No Code Available1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 120 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY. (B)(4) ARE RELATED TO EACH OTHER. (B)(4) HAS BEEN CREATED FOR CAPTURED ADDITIONAL PRODUCTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22707890 ↗
  • MalfunctionAug 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY. (B)(4) ARE RELATED TO EACH OTHER. (B)(4) HAS BEEN CREATED FOR CAPTURED ADDITIONAL PRODUCTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22707349 ↗
  • MalfunctionAug 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY. (B)(4) ARE RELATED TO EACH OTHER. (B)(4) HAS BEEN CREATED FOR CAPTURED ADDITIONAL PRODUCTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22706950 ↗
  • MalfunctionAug 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY. (B)(4) ARE RELATED TO EACH OTHER. (B)(4) HAS BEEN CREATED FOR CAPTURED ADDITIONAL PRODUCTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22706691 ↗
  • MalfunctionAug 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY. (B)(4) AND (B)(4) ARE RELATED TO EACH OTHER. (B)(4) HAS BEEN CREATED FOR CAPTURED ADDITIONAL PRODUCTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22706044 ↗
  • MalfunctionAug 1, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22689982 ↗
  • MalfunctionAug 1, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22689980 ↗
  • MalfunctionAug 1, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22689978 ↗
  • MalfunctionJul 28, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT GRATERS HAVE BECOME DULL OVER TIME AND NEED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22631778 ↗
  • MalfunctionJul 28, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT GRATERS HAVE BECOME DULL OVER TIME AND NEED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22631666 ↗
  • MalfunctionJul 7, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE INSTRUMENTS HAD AN UNKNOWN ALLEGATION AND NEED REPLACED WITH CROSS BACKS. THE EVENT DID NOT HAPPEN IN SURGERY. DURING MANUFACTURER'S INVESTIGATION OF THE RETURNED DEVICE IT WAS FOUND THAT THERE WAS BLUNTING OF CUTTING EDGES. ADDITIONALLY, CORROSION PATTERNS WERE FOUND BOTH OUTSIDE AND INSIDE OF THE CUTTING SURFACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22414456 ↗
  • MalfunctionApr 21, 2025Corroded; Corroded

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THERE WERE RUSTY PINNACLE REAMERS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21874862 ↗
  • MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER TIME AND NEED REPLACEMENT. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21517943 ↗
  • MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER TIME AND NEED REPLACEMENT. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21517812 ↗
  • MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER TIME AND NEED REPLACEMENT. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21517702 ↗
  • MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER TIME AND NEED REPLACEMENT. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21517646 ↗
  • MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. THE ISSUE DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21517632 ↗
  • MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. THE ISSUE DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21517413 ↗
  • MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. THE ISSUE DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21517384 ↗
  • MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. THE ISSUE DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21516786 ↗
  • MalfunctionFeb 25, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE GRATERS ARE DULL. THE EVENT DID NOT HAPPEN DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21461842 ↗
  • MalfunctionOct 10, 2024Material Deformation; Naturally Worn; Material Deformation; Naturally Worn

    ADDITIONAL INFORMATION RECEIVED STATES THAT NO SURGICAL DELAY FOR ANY OF THESE REPORTED DEVICES, HOWEVER, TIME IS REQUIRED TO ORDER LOANER SETS WHILE WAITING FOR MISSING ITEMS TO BE RESTOCKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20422117 ↗
  • MalfunctionJul 24, 2024Failure to Cut; Device-Device Incompatibility; Failure to Cut; Device-Device Incompatibility

    IT WAS REPORTED THAT THE REAMERS WERE DULL. HANDLES DIDN¿T FIT.

    Patient problems: Insufficient Information

    MAUDE report 19816493 ↗
  • MalfunctionJul 22, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE INSTRUMENT WAS DULL.

    Patient problems: Insufficient Information

    MAUDE report 19798578 ↗
  • MalfunctionJul 22, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE PRODUCTS WERE DULL.

    Patient problems: Insufficient Information

    MAUDE report 19797889 ↗

Other REAMER devices

All REAMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page