QUICKSET 2440-00-568Recalls & reported problems
Depuy (Ireland) · QUICKSET GRATER HEAD 68mm · Class 1
All models of QUICKSET (66)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy QuickSet grater head, a surgical reaming tool, has 110 adverse event reports, with no deaths and no injuries. Many reports describe the grater failing to cut, going dull over time, or corroding. Reports fell to 0 in the last 12 months from 17 in the 12 months before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 110
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 110 adverse event reports that list this exact device (DI 10603295124153), received between Sep 21, 2017 and Aug 1, 2025 — classified as 110 malfunctions.
The most frequently reported specific device problem is “Failure to Cut” (58 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Cut58
- Corroded25
- No Apparent Adverse Event18
- Dull, Blunt7
- Adverse Event Without Identified Device or Use Problem6
- Material Twisted/Bent3
- Naturally Worn3
- Material Discolored2
Patient problems
- No Clinical Signs, Symptoms or Conditions95
- No Consequences Or Impact To Patient13
- Insufficient Information2
- No Information1
- Not Applicable1
Recent adverse event reports
The 25 most recent of 110 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 1, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT THE GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22689981 ↗MalfunctionAug 1, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT THE GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22689977 ↗MalfunctionAug 1, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT THE GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22689976 ↗MalfunctionJul 28, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT GRATERS HAVE BECOME DULL OVER TIME AND NEED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22631945 ↗MalfunctionJul 28, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT GRATERS HAVE BECOME DULL OVER TIME AND NEED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22631808 ↗MalfunctionJul 8, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER TIME AND NEEDED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22438556 ↗MalfunctionJul 8, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER TIME AND NEEDED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22438552 ↗MalfunctionJul 8, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT THE GRATER SETS HAVE BECOME DULL OVER TIME AND NEED REPLACEMENT. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22435978 ↗MalfunctionJul 8, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT THE GRATER SETS HAVE BECOME DULL OVER TIME AND NEED REPLACEMENT. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22435976 ↗MalfunctionJul 7, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT THE INSTRUMENTS HAD AN UNKNOWN ALLEGATION AND NEED REPLACED WITH CROSS BACKS. THE EVENT DID NOT HAPPEN IN SURGERY. DURING MANUFACTURER'S INVESTIGATION OF THE RETURNED DEVICE IT WAS FOUND THAT THERE WAS BLUNTING OF CUTTING EDGES. ADDITIONALLY, CORROSION PATTERNS WERE FOUND BOTH OUTSIDE AND INSIDE OF THE CUTTING SURFACE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22414452 ↗MalfunctionMar 19, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT THE GRATERS HAD BECOME DULL OVER TIME WITH REPEATED USE AND NEED TO BE REPLACED. THE EVENT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21647363 ↗MalfunctionMar 19, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT THE GRATERS HAD BECOME DULL OVER TIME WITH REPEATED USE AND NEED TO BE REPLACED. THE EVENT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21647359 ↗MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER TIME AND NEED REPLACEMENT. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21518015 ↗MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. THE ISSUE DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21517869 ↗MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT THE GRATERS HAVE BECOME DULL OVER TIME AND NEED REPLACEMENT. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21517672 ↗MalfunctionMar 4, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT GRATERS HAD BECOME DULL OVER TIME AND NEED TO BE REPLACED. THE ISSUE DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21517271 ↗MalfunctionFeb 25, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT THE GRATERS ARE DULL. THE EVENT DID NOT HAPPEN DURING SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21461924 ↗MalfunctionOct 10, 2024Material Deformation; Naturally Worn; Material Deformation; Naturally Worn
ADDITIONAL INFORMATION RECEIVED STATES THAT NO SURGICAL DELAY FOR ANY OF THESE REPORTED DEVICES, HOWEVER, TIME IS REQUIRED TO ORDER LOANER SETS WHILE WAITING FOR MISSING ITEMS TO BE RESTOCKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20423180 ↗MalfunctionSep 9, 2024Corroded; Corroded
IT WAS REPORTED THAT THE CONSIGNMENT GRATER HEADS ARE RUSTED. ADDITIONAL INFORMATION RECEIVED STATES THAT THIS IS A PRE-OPERATIVE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20171025 ↗MalfunctionJul 22, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED, THAT THE INSTRUMENT WAS DULL.
Patient problems: Insufficient Information
MAUDE report 19798822 ↗MalfunctionJul 22, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE PRODUCTS WERE DULL.
Patient problems: Insufficient Information
MAUDE report 19797081 ↗MalfunctionJul 18, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT REAMERS WERE DULL. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19772062 ↗MalfunctionJun 11, 2024No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT THE INSTRUMENT HAS AN UNKNOWN ALLEGATION. THE EVENT DID NOT HAPPEN DURING SURGERY. ADDITIONAL EVENT INFORMATION: VISUAL ANALYSIS OF THE RETURNED DEVICE FOUND SIGNS OF WORN OUT AT THE SURFACE OF THE QUICKSET ACE GRATER HEAD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19509734 ↗MalfunctionJun 5, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT GRATERS WERE DULL AND NEED TO BE SHARPENED. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19472274 ↗MalfunctionJun 5, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS WERE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19472252 ↗
Other REAMER devices
- QUICKSET 2440-00-551Depuy (Ireland) · REAMER1,150 reports
- QUICKSET 2440-00-549Depuy (Ireland) · REAMER1,113 reports
- QUICKSET 2440-00-553Depuy (Ireland) · REAMER1,105 reports
- QUICKSET 2440-00-547Depuy (Ireland) · REAMER1,073 reports
- QUICKSET 2440-00-555Depuy (Ireland) · REAMER1,065 reports
- QUICKSET 2440-00-557Depuy (Ireland) · REAMER1,031 reports
- QUICKSET 2440-00-545Depuy (Ireland) · REAMER947 reports
- QUICKSET 2440-00-552Depuy (Ireland) · REAMER888 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page