QUICKSET 2440-00-571Recalls & reported problems

Depuy (Ireland) · QUICKSET GRATER HEAD 71mm · Class 1

All models of QUICKSET (66)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy QuickSet grater head is a reamer used to shape bone during hip surgery. The 28 reports mostly describe blades that would not cut or had gone dull and needed replacing, along with some corrosion and one broken reamer, with most reports noting no effect on the patient. No deaths or injuries were reported, and no reports came in during the past year. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
28
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 10603295124184), received between Mar 26, 2018 and Jul 22, 2024 — classified as 28 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (20 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Cut20
  • Adverse Event Without Identified Device or Use Problem3
  • Corroded3
  • Break1
  • Dull, Blunt1

Patient problems

  • No Clinical Signs, Symptoms or Conditions26
  • Insufficient Information1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 22, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE PRODUCTS WERE DULL.

    Patient problems: Insufficient Information

    MAUDE report 19796203 ↗
  • MalfunctionJul 18, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT REAMERS WERE DULL. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19772074 ↗
  • MalfunctionMay 10, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT REAMERS WERE DULL. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19285599 ↗
  • MalfunctionJul 28, 2023Break; Break

    IT WAS REPORTED THAT THE ACETABULAR REAMER WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17413514 ↗
  • MalfunctionFeb 9, 2023Corroded; Corroded

    THE NURSE AT THE HOSPITAL SENT SOME PICTURES OF THE ACETABULAR FRESAS ALLEGING SIGNS OF OXIDATION. THERE WAS NO IMPACT ON SURGERY OR PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16335788 ↗
  • MalfunctionOct 11, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15575072 ↗
  • MalfunctionAug 23, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE SURGEON COMPLAINED GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15275842 ↗
  • MalfunctionJun 8, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS SIZE 36-80 WERE DULL AND NEEDED TO BE REPLACED WITH NEW ONES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14634390 ↗
  • MalfunctionJun 7, 2022Corroded; Corroded

    IT WAS REPORTED THAT THE GRATERS ARE RUSTY. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14622807 ↗
  • MalfunctionJun 3, 2022Corroded; Corroded

    IT WAS REPORTED THAT GRATERS ARE RUSTY. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14598241 ↗
  • MalfunctionMay 5, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT ALL GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14287432 ↗
  • MalfunctionMay 5, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT ALL GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14287404 ↗
  • MalfunctionMay 5, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT ALL GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14287364 ↗
  • MalfunctionNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE GRATERS ARE DULL BY THE SALES REP. UPON FOLLOW-UP, ONLY TWO GRATER HEADS WITHIN THE SET WERE IDENTIFIED AS DULL WITHIN THE HEALTHCARE SETTING BY THE SURGEON. THE REMAINING GRATERS WERE RETURNED FOR ROUTINE PRODUCT REPLENISHMENT ONLY AND A FAILURE WAS NOT IDENTIFIED AT THE TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12805435 ↗
  • MalfunctionNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE GRATERS ARE DULL BY THE SALES REP. UPON FOLLOW-UP, ONLY TWO GRATER HEADS WITHIN THE SET WERE IDENTIFIED AS DULL WITHIN THE HEALTHCARE SETTING BY THE SURGEON. THE REMAINING GRATERS WERE RETURNED FOR ROUTINE PRODUCT REPLENISHMENT ONLY AND A FAILURE WAS NOT IDENTIFIED AT THE TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12805421 ↗
  • MalfunctionNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE GRATERS ARE DULL BY THE SALES REP. UPON FOLLOW-UP, ONLY TWO GRATER HEADS WITHIN THE SET WERE IDENTIFIED AS DULL WITHIN THE HEALTHCARE SETTING BY THE SURGEON. THE REMAINING GRATERS WERE RETURNED FOR ROUTINE PRODUCT REPLENISHMENT ONLY AND A FAILURE WAS NOT IDENTIFIED AT THE TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12805351 ↗
  • MalfunctionJun 2, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11921374 ↗
  • MalfunctionMay 19, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS SIZE 36-80 ARE DULL AND NEED TO BE SHARPENED . NO PATIENT CONSEQUENCES OR SURGICAL DELAYS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11849552 ↗
  • MalfunctionApr 6, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11620578 ↗
  • MalfunctionApr 6, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11620496 ↗
  • MalfunctionApr 6, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11620439 ↗
  • MalfunctionMar 29, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11581891 ↗
  • MalfunctionMar 29, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11581876 ↗
  • MalfunctionMar 29, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11581867 ↗
  • MalfunctionFeb 19, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT REAMERS ARE DULL AND NEED REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions