QUICKSET 2440-00-572Recalls & reported problems
Depuy (Ireland) · QUICKSET GRATER HEAD 72mm · Class 1
All models of QUICKSET (66)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy QuickSet grater head is a surgical reamer used to prepare bone. The 27 reports mostly describe blades that would not cut or had gone dull and needed replacing, plus a few describing rust or corrosion, with no harm to any patient reported. No deaths or injuries were described, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 27
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 10603295124191), received between Mar 29, 2018 and Jul 22, 2024 — classified as 27 malfunctions.
The most frequently reported specific device problem is “Failure to Cut” (20 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Cut20
- Adverse Event Without Identified Device or Use Problem3
- Corroded3
- Dull, Blunt1
Patient problems
- No Clinical Signs, Symptoms or Conditions25
- Insufficient Information1
- No Consequences Or Impact To Patient1
- Not Applicable1
Recent adverse event reports
The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 22, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE PRODUCTS WERE DULL.
Patient problems: Insufficient Information
MAUDE report 19800931 ↗MalfunctionJul 18, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT REAMERS WERE DULL. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19772152 ↗MalfunctionMay 10, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT REAMERS WERE DULL. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19285603 ↗MalfunctionFeb 9, 2023Corroded; Corroded
THE NURSE AT THE HOSPITAL SENT SOME PICTURES OF THE ACETABULAR FRESAS ALLEGING SIGNS OF OXIDATION. THERE WAS NO IMPACT ON SURGERY OR PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16335791 ↗MalfunctionOct 10, 2022Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15573851 ↗MalfunctionAug 23, 2022Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE SURGEON COMPLAINED GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15275758 ↗MalfunctionJun 8, 2022Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS SIZE 36-80 WERE DULL AND NEEDED TO BE REPLACED WITH NEW ONES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14634203 ↗MalfunctionJun 7, 2022Corroded; Corroded
IT WAS REPORTED THAT THE GRATERS ARE RUSTY. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14622636 ↗MalfunctionJun 3, 2022Corroded; Corroded
IT WAS REPORTED THAT GRATERS ARE RUSTY. THERE WAS NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14598231 ↗MalfunctionMay 5, 2022Failure to Cut; Failure to Cut
IT WAS REPORTED THAT ALL GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14287846 ↗MalfunctionMay 5, 2022Failure to Cut; Failure to Cut
IT WAS REPORTED THAT ALL GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14287845 ↗MalfunctionMay 5, 2022Failure to Cut; Failure to Cut
IT WAS REPORTED THAT ALL GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14287361 ↗MalfunctionNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE GRATERS ARE DULL BY THE SALES REP. UPON FOLLOW-UP, ONLY TWO GRATER HEADS WITHIN THE SET WERE IDENTIFIED AS DULL WITHIN THE HEALTHCARE SETTING BY THE SURGEON. THE REMAINING GRATERS WERE RETURNED FOR ROUTINE PRODUCT REPLENISHMENT ONLY AND A FAILURE WAS NOT IDENTIFIED AT THE TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12805425 ↗MalfunctionNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE GRATERS ARE DULL BY THE SALES REP. UPON FOLLOW-UP, ONLY TWO GRATER HEADS WITHIN THE SET WERE IDENTIFIED AS DULL WITHIN THE HEALTHCARE SETTING BY THE SURGEON. THE REMAINING GRATERS WERE RETURNED FOR ROUTINE PRODUCT REPLENISHMENT ONLY AND A FAILURE WAS NOT IDENTIFIED AT THE TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12805361 ↗MalfunctionNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE GRATERS ARE DULL BY THE SALES REP. UPON FOLLOW-UP, ONLY TWO GRATER HEADS WITHIN THE SET WERE IDENTIFIED AS DULL WITHIN THE HEALTHCARE SETTING BY THE SURGEON. THE REMAINING GRATERS WERE RETURNED FOR ROUTINE PRODUCT REPLENISHMENT ONLY AND A FAILURE WAS NOT IDENTIFIED AT THE TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12805321 ↗MalfunctionJun 2, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS ARE DULL .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11921411 ↗MalfunctionMay 19, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS SIZE 36-80 ARE DULL AND NEED TO BE SHARPENED. NO PATIENT CONSEQUENCES OR SURGICAL DELAYS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11849554 ↗MalfunctionApr 6, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11625234 ↗MalfunctionApr 6, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11620638 ↗MalfunctionApr 6, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11620611 ↗MalfunctionMar 29, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11582636 ↗MalfunctionMar 29, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11581916 ↗MalfunctionMar 29, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11581897 ↗MalfunctionFeb 19, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT REAMERS ARE DULL AND NEED REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11348410 ↗MalfunctionFeb 19, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT REAMERS ARE DULL AND NEED REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11348349 ↗
Other REAMER devices
- QUICKSET 2440-00-551Depuy (Ireland) · REAMER1,150 reports
- QUICKSET 2440-00-549Depuy (Ireland) · REAMER1,113 reports
- QUICKSET 2440-00-553Depuy (Ireland) · REAMER1,105 reports
- QUICKSET 2440-00-547Depuy (Ireland) · REAMER1,073 reports
- QUICKSET 2440-00-555Depuy (Ireland) · REAMER1,065 reports
- QUICKSET 2440-00-557Depuy (Ireland) · REAMER1,031 reports
- QUICKSET 2440-00-545Depuy (Ireland) · REAMER947 reports
- QUICKSET 2440-00-552Depuy (Ireland) · REAMER888 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page