QUICKSET 2440-00-574Recalls & reported problems

Depuy (Ireland) · QUICKSET GRATER HEAD 74mm · Class 1

All models of QUICKSET (66)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Quickset reamer is a surgical reamer head. The 25 reports mostly describe dull heads that did not cut as expected, including a mention of rust-like discoloration, with no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
25
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 10603295124214), received between Mar 26, 2018 and Jul 22, 2024 — classified as 25 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Cut19
  • Adverse Event Without Identified Device or Use Problem3
  • Contamination /Decontamination Problem1
  • Corroded1
  • Dull, Blunt1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions22
  • No Consequences Or Impact To Patient2
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 22, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE PRODUCTS WERE DULL.

    Patient problems: Insufficient Information

    MAUDE report 19800536 ↗
  • MalfunctionJul 18, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT REAMERS WERE DULL. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19772087 ↗
  • MalfunctionFeb 9, 2023Corroded; Corroded

    THE NURSE AT THE HOSPITAL SENT SOME PICTURES OF THE ACETABULAR FRESAS ALLEGING SIGNS OF OXIDATION. THERE WAS NO IMPACT ON SURGERY OR PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16335792 ↗
  • MalfunctionOct 10, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15570025 ↗
  • MalfunctionAug 23, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE SURGEON COMPLAINED GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15276040 ↗
  • MalfunctionJun 8, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS SIZE 36-80 WERE DULL AND NEEDED TO BE REPLACED WITH NEW ONES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14634219 ↗
  • MalfunctionMay 5, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT ALL GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14288081 ↗
  • MalfunctionMay 5, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT ALL GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14288070 ↗
  • MalfunctionMay 5, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT ALL GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14287406 ↗
  • MalfunctionNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE GRATERS ARE DULL BY THE SALES REP. UPON FOLLOW-UP, ONLY TWO GRATER HEADS WITHIN THE SET WERE IDENTIFIED AS DULL WITHIN THE HEALTHCARE SETTING BY THE SURGEON. THE REMAINING GRATERS WERE RETURNED FOR ROUTINE PRODUCT REPLENISHMENT ONLY AND A FAILURE WAS NOT IDENTIFIED AT THE TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12805416 ↗
  • MalfunctionNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE GRATERS ARE DULL BY THE SALES REP. UPON FOLLOW-UP, ONLY TWO GRATER HEADS WITHIN THE SET WERE IDENTIFIED AS DULL WITHIN THE HEALTHCARE SETTING BY THE SURGEON. THE REMAINING GRATERS WERE RETURNED FOR ROUTINE PRODUCT REPLENISHMENT ONLY AND A FAILURE WAS NOT IDENTIFIED AT THE TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12805399 ↗
  • MalfunctionNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE GRATERS ARE DULL BY THE SALES REP. UPON FOLLOW-UP, ONLY TWO GRATER HEADS WITHIN THE SET WERE IDENTIFIED AS DULL WITHIN THE HEALTHCARE SETTING BY THE SURGEON. THE REMAINING GRATERS WERE RETURNED FOR ROUTINE PRODUCT REPLENISHMENT ONLY AND A FAILURE WAS NOT IDENTIFIED AT THE TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12805312 ↗
  • MalfunctionJun 2, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11921442 ↗
  • MalfunctionMay 19, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS SIZE 36-80 ARE DULL AND NEED TO BE SHARPENED . NO PATIENT CONSEQUENCES OR SURGICAL DELAYS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11849562 ↗
  • MalfunctionApr 6, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11625254 ↗
  • MalfunctionApr 6, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11625251 ↗
  • MalfunctionApr 6, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11625241 ↗
  • MalfunctionMar 29, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11582720 ↗
  • MalfunctionMar 29, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11582712 ↗
  • MalfunctionMar 29, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11582708 ↗
  • MalfunctionFeb 19, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT REAMERS ARE DULL AND NEED REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11348447 ↗
  • MalfunctionFeb 19, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT REAMERS ARE DULL AND NEED REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11348438 ↗
  • MalfunctionFeb 19, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT REAMERS ARE DULL AND NEED REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11348437 ↗
  • MalfunctionSep 11, 2020Contamination /Decontamination Problem; Material Integrity Problem; Contamination /Decontamination Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE ACETABULAR REAMERS WERE DAMAGED AND WILL BE RETURNED TO BE ANALYZED. THE 74 MM REAMER HAS A STAIN MARKS ON THE INSIDE. THE HOSPITAL USED NEUTRAL PH.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10524843 ↗
  • MalfunctionMar 26, 2018Dull, Blunt; Dull, Blunt

    IT WAS REPORTED THAT THE SURGEON STRUGGLED AS THE REAMERS WERE BLUNT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7369427 ↗

Other REAMER devices

All REAMER devices →