Skip to content

QUICKSET 2440-00-575Recalls & reported problems

Depuy (Ireland) · QUICKSET GRATER HEAD 75mm · Class 1

All models of QUICKSET (66)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
23
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 10603295124221), received between Mar 26, 2018 and Jul 22, 2024 — classified as 23 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Cut19
  • Adverse Event Without Identified Device or Use Problem3
  • Dull, Blunt1

Patient problems

  • No Clinical Signs, Symptoms or Conditions21
  • Insufficient Information1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 22, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE PRODUCTS WERE DULL.

    Patient problems: Insufficient Information

    MAUDE report 19796471 ↗
  • MalfunctionJul 18, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT REAMERS WERE DULL. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19772085 ↗
  • MalfunctionOct 11, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15574772 ↗
  • MalfunctionAug 23, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE SURGEON COMPLAINED GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15275895 ↗
  • MalfunctionJun 8, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS SIZE 36-80 WERE DULL AND NEEDED TO BE REPLACED WITH NEW ONES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14634223 ↗
  • MalfunctionMay 5, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT ALL GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14288055 ↗
  • MalfunctionMay 5, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT ALL GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14288037 ↗
  • MalfunctionMay 5, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT ALL GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14288003 ↗
  • MalfunctionNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE GRATERS ARE DULL BY THE SALES REP. UPON FOLLOW-UP, ONLY TWO GRATER HEADS WITHIN THE SET WERE IDENTIFIED AS DULL WITHIN THE HEALTHCARE SETTING BY THE SURGEON. THE REMAINING GRATERS WERE RETURNED FOR ROUTINE PRODUCT REPLENISHMENT ONLY AND A FAILURE WAS NOT IDENTIFIED AT THE TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12805409 ↗
  • MalfunctionNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE GRATERS ARE DULL BY THE SALES REP. UPON FOLLOW-UP, ONLY TWO GRATER HEADS WITHIN THE SET WERE IDENTIFIED AS DULL WITHIN THE HEALTHCARE SETTING BY THE SURGEON. THE REMAINING GRATERS WERE RETURNED FOR ROUTINE PRODUCT REPLENISHMENT ONLY AND A FAILURE WAS NOT IDENTIFIED AT THE TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12805325 ↗
  • MalfunctionNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE GRATERS ARE DULL BY THE SALES REP. UPON FOLLOW-UP, ONLY TWO GRATER HEADS WITHIN THE SET WERE IDENTIFIED AS DULL WITHIN THE HEALTHCARE SETTING BY THE SURGEON. THE REMAINING GRATERS WERE RETURNED FOR ROUTINE PRODUCT REPLENISHMENT ONLY AND A FAILURE WAS NOT IDENTIFIED AT THE TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12805304 ↗
  • MalfunctionJun 2, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11921423 ↗
  • MalfunctionMay 19, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS SIZE 36-80 ARE DULL AND NEED TO BE SHARPENED . NO PATIENT CONSEQUENCES OR SURGICAL DELAYS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11849559 ↗
  • MalfunctionApr 6, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11625263 ↗
  • MalfunctionApr 6, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11625261 ↗
  • MalfunctionApr 6, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11625257 ↗
  • MalfunctionMar 29, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11582749 ↗
  • MalfunctionMar 29, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11582737 ↗
  • MalfunctionMar 29, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE GRATERS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11582730 ↗
  • MalfunctionFeb 19, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT REAMERS ARE DULL AND NEED REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11348465 ↗
  • MalfunctionFeb 19, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT REAMERS ARE DULL AND NEED REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11348464 ↗
  • MalfunctionFeb 19, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT REAMERS ARE DULL AND NEED REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11348451 ↗
  • MalfunctionMar 26, 2018Dull, Blunt; Dull, Blunt

    IT WAS REPORTED THAT THE SURGEON STRUGGLED AS THE REAMERS WERE BLUNT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7369664 ↗

Other REAMER devices

All REAMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page