QUICKSET 2440-00-576Recalls & reported problems
Depuy (Ireland) · QUICKSET GRATER HEAD 76mm · Class 1
All models of QUICKSET (66)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 23
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 10603295124238), received between Mar 23, 2018 and Jul 18, 2024 — classified as 23 malfunctions.
The most frequently reported specific device problem is “Failure to Cut” (19 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Cut19
- Adverse Event Without Identified Device or Use Problem3
- Dull, Blunt1
Patient problems
- No Clinical Signs, Symptoms or Conditions22
- No Consequences Or Impact To Patient1
- Not Applicable1
Recent adverse event reports
The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 18, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE REAMERS WERE DULL. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19771139 ↗MalfunctionJul 18, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE REAMERS WERE DULL. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19771125 ↗MalfunctionOct 11, 2022Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15576959 ↗MalfunctionAug 23, 2022Failure to Cut; Failure to Cut
IT WAS REPORTED THAT SURGEON COMPLAINED GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15275442 ↗MalfunctionJun 8, 2022Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS SIZE 36-80 WERE DULL AND NEEDED TO BE REPLACED WITH NEW ONES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14634196 ↗MalfunctionMay 7, 2022Failure to Cut; Failure to Cut
THE GRATERS ARE DULL. SURGICAL DELAY IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14315646 ↗MalfunctionMay 6, 2022Failure to Cut; Failure to Cut
THE GRATERS ARE DULL. SURGICAL DELAY IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14295725 ↗MalfunctionMay 6, 2022Failure to Cut; Failure to Cut
THE GRATERS ARE DULL. SURGICAL DELAY IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14295332 ↗MalfunctionNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE GRATERS ARE DULL BY THE SALES REP. UPON FOLLOW-UP, ONLY TWO GRATER HEADS WITHIN THE SET WERE IDENTIFIED AS DULL WITHIN THE HEALTHCARE SETTING BY THE SURGEON. THE REMAINING GRATERS WERE RETURNED FOR ROUTINE PRODUCT REPLENISHMENT ONLY AND A FAILURE WAS NOT IDENTIFIED AT THE TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12805209 ↗MalfunctionNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE GRATERS ARE DULL BY THE SALES REP. UPON FOLLOW-UP, ONLY TWO GRATER HEADS WITHIN THE SET WERE IDENTIFIED AS DULL WITHIN THE HEALTHCARE SETTING BY THE SURGEON. THE REMAINING GRATERS WERE RETURNED FOR ROUTINE PRODUCT REPLENISHMENT ONLY AND A FAILURE WAS NOT IDENTIFIED AT THE TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12805149 ↗MalfunctionNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE GRATERS ARE DULL BY THE SALES REP. UPON FOLLOW-UP, ONLY TWO GRATER HEADS WITHIN THE SET WERE IDENTIFIED AS DULL WITHIN THE HEALTHCARE SETTING BY THE SURGEON. THE REMAINING GRATERS WERE RETURNED FOR ROUTINE PRODUCT REPLENISHMENT ONLY AND A FAILURE WAS NOT IDENTIFIED AT THE TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12805099 ↗MalfunctionJun 2, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11921350 ↗MalfunctionMay 19, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS SIZE 36-80 ARE DULL AND NEED TO BE SHARPENED. NO PATIENT CONSEQUENCES OR SURGICAL DELAYS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11849571 ↗MalfunctionApr 6, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11625384 ↗MalfunctionApr 6, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11625380 ↗MalfunctionApr 6, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11625374 ↗MalfunctionMar 30, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11588733 ↗MalfunctionMar 30, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11588697 ↗MalfunctionMar 30, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE GRATERS ARE DULL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11588683 ↗MalfunctionFeb 19, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT REAMERS ARE DULL AND NEED REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11348490 ↗MalfunctionFeb 19, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT REAMERS ARE DULL AND NEED REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11348486 ↗MalfunctionFeb 19, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT REAMERS ARE DULL AND NEED REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11348477 ↗MalfunctionMar 23, 2018Dull, Blunt; Dull, Blunt
SURGEON STRUGGLED AS THE REAMERS WERE BLUNT.
Patient problems: No Consequences Or Impact To Patient; Not Applicable
MAUDE report 7365219 ↗
Other REAMER devices
- QUICKSET 2440-00-551Depuy (Ireland) · REAMER1,150 reports
- QUICKSET 2440-00-549Depuy (Ireland) · REAMER1,113 reports
- QUICKSET 2440-00-553Depuy (Ireland) · REAMER1,105 reports
- QUICKSET 2440-00-547Depuy (Ireland) · REAMER1,073 reports
- QUICKSET 2440-00-555Depuy (Ireland) · REAMER1,065 reports
- QUICKSET 2440-00-557Depuy (Ireland) · REAMER1,031 reports
- QUICKSET 2440-00-545Depuy (Ireland) · REAMER947 reports
- QUICKSET 2440-00-552Depuy (Ireland) · REAMER888 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page