NA 2490-49-000Recalls & reported problems
Depuy Orthopaedics, Inc. · EXTERNAL ALIGNMENT ANKLE CLAMP · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10603295125891), received between Sep 24, 2018 and Oct 10, 2025 — classified as 16 malfunctions.
The most frequently reported specific device problem is “Break” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break7
- Mechanical Jam6
- Device-Device Incompatibility4
- Material Twisted/Bent3
- Physical Resistance/Sticking3
- Material Deformation2
- Component Missing1
- Crack1
Patient problems
- No Consequences Or Impact To Patient9
- No Clinical Signs, Symptoms or Conditions5
- Not Applicable2
- No Code Available1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 10, 2025Break; Break
IT WAS REPORTED THAT ON THE LCS ALIGNMENT ANKLE CLAMP THE BLACK PLASTIC PIECE THAT CONNECTS WITH THE UPROD HAS BEEN SNAPPED. THERE WERE NO DELAYS TO THE SURGERY, AND THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23269485 ↗MalfunctionJul 9, 2025Break; Break
IT WAS REPORTED THAT THE EXTERNAL ALIGNMENT HANDLE ON THE LCS MILESTONE TIBIAL JIG IS BROKEN. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22463185 ↗MalfunctionMar 18, 2021Mechanical Jam; Mechanical Jam
BROKEN AND WORN. TIBIAL JIG LEG CLAMP WHEEL IS LOCKED IN PLACE POST STERILIZATION. BROKEN STEINMAN PIN WHICH SNAPPED AT THE END IN THE PIN DRIVER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11516396 ↗MalfunctionMar 15, 2021Mechanical Jam; Mechanical Jam
BROKEN AND WORN LCS MILESTONE TIBIAL JIG. THE SILVER LOCKING KNOB ON THE ANKLE CLAMP WAS JAMMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11487949 ↗MalfunctionFeb 11, 2021Material Twisted/Bent; Mechanical Jam; Material Twisted/Bent; Mechanical Jam
WORN INSTRUMENTS IDENTIFIED BY STERILIZATION DEPARTMENT, ITEMS ARE NEW CONSIGNMENT SENT TO THE HOSPITAL. THE LOCKING WHEEL IS COLD WELDED, SO THIS ITEM CAN NOT BE USED. DID NOT OCCUR DURING SURGERY SO NO SURGEON WAS INVOLVED. THIS HAS BEEN REPORTED BY ONE OF THE HEALTHCARE WORKERS AT THE HOSPITAL WHO IS THE CONTACT IN THIS FORM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11315742 ↗MalfunctionOct 7, 2020Break; Material Twisted/Bent; Break; Material Twisted/Bent
BROKEN AND WORN LCS MILESTONE TIBIAL JIG BLACK UPROD. PRODUCT CODE 2490-46. LOT AYO597F AND BLACK UPROD. THE SILVER BARREL DOES NOT ENGAGE PROPERLY ON THE THREADS OF THE UPROD.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 10642885 ↗MalfunctionMar 30, 2020Break; Device-Device Incompatibility; Material Deformation; Material Twisted/Bent; Break; Device-Device Incompatibility; Material Deformation; Material Twisted/Bent
BROKEN AND WORN LCS EXTERNAL ALIGNMENT CLAMP REPORTED BY CSSD STAFF.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9898318 ↗MalfunctionMar 12, 2020Mechanical Jam; Physical Resistance/Sticking; Mechanical Jam; Physical Resistance/Sticking
BROKEN AND WORN TIBIAL JIG LCS. WOULD NOT EXTEND DURING SURGERY, LOCKING MECHANISM WAS JAMMED, 10 MINUTE DELAY WHILST ALTERNATIVE KIT SOURCED. PIN PULLER THE SPRING HAS WORN AND A NEW ONE IS REQUIRED.
Patient problems: Not Applicable
MAUDE report 9823908 ↗MalfunctionDec 11, 2019Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED IT WAS REVEALED AT THE TIME OF WASHING THE INSTRUMENTS POSTOPERATIVELY TKA ON (B)(6) 2019 THAT THE TIBIAL RESECTION GUIDE (P/N: 228701000) AND THE ANKLE CLAMP (P/N: 249049000) COULD NOT BE SEPARATED BECAUSE IT COULD NOT BE LOOSENED THE SCREW OF THE CONNECTION PART. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9451768 ↗MalfunctionDec 3, 2019Break; Break
IT WAS REPORTED THAT ON (B)(6) 2019, THE SCREW PART OF THE ALIGN ANKLE CLAMP (P:N/ 249049000) BROKEN DURING TKA SURGERY FOR OA ON (B)(6) 2019. WHEN THE SURGEON TRIED TO FIX THE BONE TO CUT PROXIMAL TIBIAL BONE, THE SCREW PART OF THE CLAMP BROKE. THERE WAS NO ADVERSE EVENT TO THE PATIENT AND NO SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9405689 ↗MalfunctionNov 26, 2019Device-Device Incompatibility; Mechanical Jam; Physical Resistance/Sticking; Device-Device Incompatibility; Mechanical Jam; Physical Resistance/Sticking
IT WAS REPORTED THAT THE TIBIAL UPROD WILL NOT SLIDE DOWN INTO THE CLIP. LOOKS LIKE METAL IMPEDING THE TUBE. A SURGICAL DELAY OF 5 MINUTES.
Patient problems: Not Applicable
MAUDE report 9376108 ↗MalfunctionJun 25, 2019Break; Break
TOTAL KNEE REPLACEMENT: (B)(6) HOSPITAL, PRIMARY IMPLANTED DATE: (B)(6) 2019. NOT A REVISION. CONSIGNMENT INSTRUMENT BROKEN NEEDING URGENT REPLACEMENT. INSTRUMENT IS IN EXCESS OF 25 YEARS OLD. INSTRUMENT WAS ABLE TO STILL WORK DESPITE BEING BROKEN. SPRING ATTACHED TO ANKLE CLAMP KEEPS FALLING OFF.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8730521 ↗MalfunctionApr 3, 2019Component Missing; Device-Device Incompatibility; Material Deformation; Physical Resistance/Sticking; Component Missing; Device-Device Incompatibility; Material Deformation; Physical Resistance/Sticking
ONE INSTRUMENT DID NOT PERFORM AS INTENDED DUE TO MISSING PART AND JAMMED SCREW. OTHER 2-PART INSTRUMENT DID NOT FIT TOGETHER. PATIENT STATUS/ OUTCOME / CONSEQUENCES: NO.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8480083 ↗MalfunctionMar 28, 2019Break; Break
TIBIAL JIG IS BROKEN. PATIENT STATUS/ OUTCOME / CONSEQUENCES: NO.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8461557 ↗MalfunctionNov 19, 2018Mechanical Jam; Mechanical Jam
THE LCS ANKLE CLAMP BLACK UP ROD, HAS THE SLEEVE STUCK IN PLACE. UNABLE TO INSERT AND SECURE THE PROXIMAL ROD THAT ATTACHES THE TIBIAL CUTTING BLOCK INTO THE ANKLE CLAMP.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8086118 ↗MalfunctionSep 24, 2018Crack; Crack
A CRACK HAS DEVELOPED IN AN LCS RP TRIAL INSERT FROM THE INSTRUMENT KIT. SPD DEPARTMENT IS UNABLE TO STERILIZE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7902936 ↗
Other CLAMP devices
- ATTUNE 2544-00-001Depuy (Ireland) · CLAMP144 reports
- NA 2178-63-134Depuy (Ireland) · CLAMP137 reports
- ATTUNE 2545-01-055Depuy (Ireland) · CLAMP128 reports
- SPECIALIST 86-5035Depuy (Ireland) · CLAMP101 reports
- SPECIALIST 96-6670Depuy Orthopaedics, Inc. · CLAMP92 reports
- ATRAUMATA FB469RAesculap AG · CLAMP1 recall0 reports
- Centurion I68080Centurion Medical Products Corporation · CLAMP1 recall0 reports
- ATRAUMATA FB459RAesculap AG · CLAMP1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page