Skip to content

PROSTALAC 2541-88-000Recalls & reported problems

Depuy Orthopaedics, Inc. · PROSTALAC LONG STEM MOLD - RIGHT SIZE 3 240mm · Class 2 · Implant

All models of PROSTALAC (21)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10603295129653), received between Oct 23, 2023 and Apr 2, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Device-Device Incompatibility1
  • Difficult to Open or Close1
  • Mechanical Jam1
  • Mechanical Problem1
  • Naturally Worn1
  • Scratched Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 2, 2026Break; Scratched Material

    IT WAS REPORTED THAT THE STERILE PROCESSING DEPARTMENT REFUSED TO RESTERILIZE THE INSTRUMENTS ANY LONGER DUE THE WEAR AND TEAR OF THE METAL. THERE WERE MANY CHIPS AND SCRATCHES ON THEM, WHICH COMPROMISED PATIENT SAFETY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24767996 ↗
  • MalfunctionJan 28, 2026Break

    IT WAS REPORTED THAT THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24199439 ↗
  • MalfunctionJan 20, 2026Break; Mechanical Jam

    IT WAS REPORTED THAT THE PROSTALAC MOLD BOLT CAME OFF AND THE POSITIONERS FOR CENTERING WERE NOT ON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24117623 ↗
  • MalfunctionJul 8, 2024Mechanical Problem; Naturally Worn; Mechanical Problem; Naturally Worn

    IT WAS REPORTED THAT THE EVENT DID NOT HAPPEN IN SURGERY. THE MOLDS WERE WORN AND THE THREADS WERE STRIPPED, SO THE NUTS WOULD NOT TIGHTEN ON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19691207 ↗
  • MalfunctionOct 23, 2023Device-Device Incompatibility; Difficult to Open or Close; Device-Device Incompatibility; Difficult to Open or Close

    IT WAS REPORTED THAT THE PROSTALAC MOLDS DON'T FIT TOGETHER PROPERLY AND THE PROSTALAC NUTS ARE STRIPPED. THE EVENT DID NOT HAPPEN DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17989196 ↗

Other BONE CEMENT, ANTIBIOTIC devices

All BONE CEMENT, ANTIBIOTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page