Skip to content

ATTUNE 2544-01-005Recalls & reported problems

Depuy (Ireland) · ATTUNE FEMORAL INTRODUCER · Class 1

All models of ATTUNE (1475)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Attune femoral introducer is a surgical instrument used during knee replacement. Most of the 1,368 reports describe parts of the instrument breaking, such as a handle, knob or clip breaking off, sometimes while implanting. The reports include no deaths, 1 injury and 1,367 malfunctions. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,383
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,383 adverse event reports that list this exact device (DI 10603295130215), received between Jun 21, 2017 and Sep 23, 2026 — classified as 1 injury, 1,382 malfunctions.

The most frequently reported specific device problem is “Break” (1,297 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 11 reportsDec 2023: 16 reportsJan 2024: 22 reportsFeb 2024: 21 reportsMar 2024: 14 reportsApr 2024: 22 reportsMay 2024: 13 reportsJun 2024: 25 reportsJul 2024: 8 reportsAug 2024: 7 reportsSep 2024: 23 reportsOct 2024: 25 reportsNov 2024: 16 reportsDec 2024: 7 reportsJan 2025: 28 reportsFeb 2025: 4 reportsMar 2025: 17 reportsApr 2025: 12 reportsMay 2025: 23 reportsJun 2025: 14 reportsJul 2025: 18 reportsAug 2025: 20 reportsSep 2025: 8 reportsOct 2025: 24 reportsNov 2025: 33 reportsDec 2025: 25 reportsJan 2026: 28 reportsFeb 2026: 31 reportsMar 2026: 18 reportsApr 2026: 21 reportsMay 2026: 15 reportsJun 2026: 10 reportsJul 2026: 2 reportsAug 2026: 12 reportsSep 2026: 20 reportsOct 2026: 0 reports20242025202633
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1,297
  • Crack32
  • No Apparent Adverse Event23
  • Device-Device Incompatibility21
  • Fracture10
  • Naturally Worn10
  • Mechanical Jam9
  • Detachment of Device or Device Component7

Patient problems

  • No Clinical Signs, Symptoms or Conditions904
  • No Consequences Or Impact To Patient319
  • Insufficient Information141
  • No Information24
  • No Code Available21
  • Not Applicable5

Recent adverse event reports

The 25 most recent of 1,383 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026No Apparent Adverse Event

    It was reported that during a total knee arthroplasty surgery, the device had an unspecified malfunction. During evaluation, it was determined that the wing knob screw component on the ATTUNE FEMORAL INTRODUCER device was broken.

    Patient problems: Insufficient Information

    MAUDE report 50871915 ↗
  • MalfunctionSep 23, 2026Break

    It was reported that femoral initiator broken, red outer knob broken off.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50869017 ↗
  • MalfunctionSep 22, 2026Break

    It was reported that the femoral Initiator red handle knob was fallen off. It was reported that all fragments were retrieved.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50852314 ↗
  • MalfunctionSep 21, 2026Break

    It was reported that adjusting knob broke off during surgery. All pieces recovered. No surgical delay. Instrument removed from tray.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50841532 ↗
  • MalfunctionSep 21, 2026Break

    It was reported that the lateral knob to tighten/loosen has broken off. There was no patient harm or surgical delay.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50840877 ↗
  • MalfunctionSep 18, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50822704 ↗
  • MalfunctionSep 17, 2026Break

    It was reported that both femoral holders broke on impacting the femoral prosthesis, no damage to the patient or implant was reported. Another Instrument was used from another set.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50804682 ↗
  • MalfunctionSep 17, 2026Break

    It was reported that both femoral holders broke on impacting the femoral prosthesis, no damage to the patient or implant was reported. Another Instrument was used from another set.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50804673 ↗
  • MalfunctionSep 16, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50794858 ↗
  • MalfunctionSep 14, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50764554 ↗
  • MalfunctionSep 14, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50764511 ↗
  • MalfunctionSep 14, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50764504 ↗
  • MalfunctionSep 14, 2026Leak/Splash

    It was reported that the VELYS SAW HANDPIECE device was leaking an oily substance where the blade inserts. It was not reported if the device was used in surgery. There was no delay to the procedure reported. There were no reports of injuries, medical intervention or prolonged hospitalization. All available information has been disclosed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50755089 ↗
  • MalfunctionSep 11, 2026Break

    It was reported that the attune femoral introducer broke on impacting.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50737312 ↗
  • MalfunctionSep 11, 2026Break

    It was reported that femoral introducer side tightening handle broke off at the threads. Device broke into two pieces during cleaning. The event did not happen in surgery.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50734367 ↗
  • MalfunctionSep 9, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50698578 ↗
  • MalfunctionSep 9, 2026Device-Device Incompatibility

    It was reported that during an unspecified surgical procedure it was observed that the femoral introducer on attune primary tray allows too much movement of the implant on the handle when the implant is being inserted. The surgeon stated that this makes it difficult to insert the femoral implant accurately. There was no patient harm, or surgical delay. No further information was provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50692571 ↗
  • MalfunctionSep 8, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50677057 ↗
  • MalfunctionSep 8, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50677055 ↗
  • MalfunctionSep 3, 2026Break

    It was reported that the femoral introducer was broken during case. The red tightening knob fell off. All pieces were removed. No more information is available.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50587395 ↗
  • MalfunctionAug 31, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50531506 ↗
  • MalfunctionAug 28, 2026Break

    It was reported that the femoral introducer handle was broke. Both pieces were recovered. No additional surgery time was required. Backup instruments were on hand.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50496764 ↗
  • MalfunctionAug 27, 2026Break

    It was reported that the red piece on the femoral introducer broke off.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50483649 ↗
  • MalfunctionAug 27, 2026Break

    It was reported that during a Total Knee Arthroplasty (TKA) surgery the tightening knob on the femoral introducer broke off. The incident occurred after the implants had been secured in place. The surgery was successfully completed. There were no delays to the procedure. There were no adverse events.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50476689 ↗
  • MalfunctionAug 27, 2026Break

    It was reported that the Impactor handle broke while implanting the femur. It did not impact the case

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50476600 ↗

Other DRIVER, PROSTHESIS devices

All DRIVER, PROSTHESIS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page