SUMMIT 2570-05-100Recalls & reported problems

Depuy (Ireland) · SUMMIT STANDARD IMPACTOR · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Summit standard impactor has 224 adverse event reports, all describing malfunctions, with no injuries and no deaths. Many reports describe worn, cracked, bent or blunted parts, and some describe the tool getting stuck inside the stem and having to be forced out. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
225
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 225 adverse event reports that list this exact device (DI 10603295143215), received between Sep 1, 2017 and Sep 15, 2026 — classified as 225 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (105 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 2 reportsFeb 2024: 1 reportsMar 2024: 3 reportsApr 2024: 4 reportsMay 2024: 0 reportsJun 2024: 4 reportsJul 2024: 3 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 3 reportsJun 2025: 4 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 1 reportsDec 2025: 10 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility105
  • Naturally Worn62
  • Material Deformation57
  • Break50
  • Material Twisted/Bent34
  • Material Integrity Problem8
  • Appropriate Device Problem Term/Code Not Available5
  • Entrapment of Device4

Patient problems

  • No Clinical Signs, Symptoms or Conditions145
  • No Consequences Or Impact To Patient58
  • Insufficient Information11
  • No Code Available8
  • No Information6
  • Limb Fracture3
  • Not Applicable3
  • Foreign Body In Patient1

Recent adverse event reports

The 25 most recent of 225 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50779052 ↗
  • MalfunctionMay 4, 2026Material Deformation

    IT WAS REPORTED THAT THE T-HANDLES HAVE A LARGE CRACK IN THE MIDDLE AND CANNOT BE STERILIZED. THE SUMMIT STANDARD STEM INSERTER HAS A DEFORMED TIP AND CANNOT BE USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25067590 ↗
  • MalfunctionFeb 20, 2026Break; Device-Device Incompatibility

    IT WAS REPORTED THE INSTRUMENT HAS A BLUNTED TIP; GETS STUCK INSIDE STEM AND MUST BE FORCIBLY REMOVED. NO PATIENT OUTCOMES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24400385 ↗
  • MalfunctionDec 26, 2025Device-Device Incompatibility; Device-Device Incompatibility

    ADDITIONAL INFORMATION WAS RECEIVED AND STATES THAT THERE WAS NO ADVERSE EVENTS, LOT NUMBERS NOT VISIBLE. THE ITEMS HAVE BEEN DISCARDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23899472 ↗
  • MalfunctionDec 26, 2025Device-Device Incompatibility; Device-Device Incompatibility

    ADDITIONAL INFORMATION WAS RECEIVED AND STATES THAT: NO ADVERSE EVENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23899381 ↗
  • MalfunctionDec 26, 2025Device-Device Incompatibility; Device-Device Incompatibility

    ADDITIONAL INFORMATION WAS RECEIVED AND STATES THAT NO ADVERSE EVENTS, LOT NUMBERS NOT VISIBLE. THE ITEMS HAVE BEEN DISCARDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23899333 ↗
  • MalfunctionDec 26, 2025Device-Device Incompatibility; Device-Device Incompatibility

    ADDITIONAL INFORMATION WAS RECEIVED AND STATES THAT NO ADVERSE EVENTS, LOT NUMBERS NOT VISIBLE. THE ITEMS HAVE BEEN DISCARDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23899279 ↗
  • MalfunctionDec 26, 2025Device-Device Incompatibility; Device-Device Incompatibility

    THE METALL IS WORN, SO THE FEMURCOMPONENT CAN NOT BE STABILE INSERTER TO THE FEMURCANAL WITH THE INSERTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23897698 ↗
  • MalfunctionDec 22, 2025Break; Break

    IT WAS REPORTED THAT (3) BROKEN IMPACTOR INSERTERS. THE DEVICES WERE BLUNT AND BROKEN. THERE WAS NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23864462 ↗
  • MalfunctionDec 22, 2025Break; Break

    IT WAS REPORTED THAT (3) BROKEN IMPACTOR INSERTERS. THE DEVICES WERE BLUNT AND BROKEN. THERE WAS NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23864450 ↗
  • MalfunctionDec 22, 2025Break; Break

    IT WAS REPORTED THAT (3) BROKEN IMPACTOR INSERTERS. THE DEVICES WERE BLUNT AND BROKEN. THERE WAS NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23864449 ↗
  • MalfunctionDec 16, 2025Device-Device Incompatibility; Material Deformation; Material Integrity Problem; Device-Device Incompatibility; Material Deformation; Material Integrity Problem

    ADDITIONAL INFORMATION RECEIVED: A. CAN YOU PLEASE CLARIFY WHAT DO YOU MEAN BY "DAMAGED"? WAS IT WORN, CRACKED, BROKEN INTO PIECES, BENT, STRIPPED, CROSS THREADED OR ANY DEVICE INTERACTION? RESPONSE: IT WAS WORN/DEFORMED AT THE TIP MEANING IT DOES NOT ENGAGE PROPERLY INTO THE IMPLANT. IT IS GENERAL WEAR AND TEAR AND HAPPEN FREQUENTLY WITH THIS INSTRUMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23821110 ↗
  • MalfunctionDec 11, 2025Break; Break

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE CORAIL STEM COULD NOT FIT ONTO THE HANDLE. IT WAS REPORTED THAT AT THE POINT OF CONTACT, THE METAL SITE HAD CHANGED SHAPE. IT WAS REPORTED THAT THERE WAS NO ALLEGATION AGAINST THE CORAIL STEM DEVICE. IT WAS FURTHER OBSERVED THAT THE METAL RING ON TOP OF THE PINNACLE INSERTER HANDLE HAD SNAPPED OFF. IT WAS REPORTED THAT A NEW DEVICE WAS OPENED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT, OR SURGICAL DELAY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23778172 ↗
  • MalfunctionNov 12, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE TIP OF THE INSTRUMENT IS DAMAGED. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23533822 ↗
  • MalfunctionOct 27, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE IMPACTOR TIP WAS DAMAGED DUE TO WEAR AND TEAR, GETS STUCK IN IMPLANT WHEN FINISHED IMPACTING. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23390308 ↗
  • MalfunctionOct 15, 2025Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT A SINGLE PIECE CORAIL STEM INSERTER INSTRUMENT WAS FAULTY. UNABLE TO THREAD INTO STEM IMPLANT. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THERE WAS NO PATIENT CONSEQUENCE. UPON VISUAL INSPECTION, A SECOND INSERTER WAS BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23294615 ↗
  • MalfunctionSep 9, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT A CORAIL STEM INTRODUCER NEEDS REPLACING AS WHERE IT ENGAGES WITH THE STEM IT LOCKS THE STEM IN PLACE AND IT IS HARD TO REMOVE. THERE WAS NO SURGICAL DELAY AND NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23008515 ↗
  • MalfunctionJul 30, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE END OF THE INSERTER IS BENT. THERE WAS NO HARM TO PATIENT AND NO DELAY TO SURGERY OCCURRED AS A RESULT. THE ISSUE WAS FOUND DURING ROUTINE INSPECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22664170 ↗
  • MalfunctionJun 26, 2025Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE ITEM HAD A BENT TIP AND WAS STUCK IN THE IMPLANT. THERE WAS NO PATIENT CONSEQUENCE. THE INSTRUMENT WAS SUCCESSFULLY RETRIEVED FROM THE IMPLANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22335258 ↗
  • MalfunctionJun 23, 2025Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT A SUMMIT STEM INSERTER HAS A DAMAGED/WARPED TIP AND GETS STUCK IN IMPLANT WHEN IMPACTING. T HANDLE HAS A CRACK IN IT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22309103 ↗
  • MalfunctionJun 19, 2025Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE SINGLE PIECE CORAIL STEM INSERTER INSTRUMENT IS FAULTY. IT IS UNABLE TO THREAD INTO STEM IMPLANT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22293535 ↗
  • MalfunctionJun 11, 2025Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE INSTRUMENT IS BURRED AND DOES NOT ENGAGE WITH PROSTHESIS PROPERLY. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22186292 ↗
  • MalfunctionMay 20, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING A PROCEDURE HE INSTRUMENT CORAIL AMT NECK SEG 135D STD AND BROACH CORAIL SIZE 14, GOT STUCK TOGETHER WHEN TRIALING AND CANNOT BE SEPARATED. THE INSTRUMENTS SUMMIT STANDARD IMP. INSERTER AND QUICKSET TAPER HEX SCDR RIGID HAS A WORN OUT THREADS/TIP WHICH TENDS FOR THE IMPLANT TO GET STUCK AS WELL. ANOTHER SET WAS OPENED TO USE. THERE WAS NO PATIENT CONSEQUENCE. THERE WAS NO ADDITIONAL TIME REQUIRED ONLY WAS WHEN OPENING FOR ANOTHER SET.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22054865 ↗
  • MalfunctionMay 1, 2025Device-Device Incompatibility; Unintended Movement; Device-Device Incompatibility; Unintended Movement

    DURING ROUTINE MAINTENANCE IT WAS DISCOVERED THAT TWO STANDARD IMPACTORS WERE DAMAGED DUE TO REGULAR USE. THE TIP WHICH CONNECTS TO THE STEM IS WORN AND FEELS LOOSE WHEN THE STEM IS BEING INSERTED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21942678 ↗
  • MalfunctionMay 1, 2025Device-Device Incompatibility; Unintended Movement; Device-Device Incompatibility; Unintended Movement

    IT WAS REPORTED THAT DURING ROUTINE MAINTENANCE IT WAS DISCOVERED THAT TWO STANDARD IMPACTORS WERE DAMAGED DUE TO REGULAR USE. THE TIP WHICH CONNECTS TO THE STEM IS WORN AND FEELS LOOSE WHEN THE STEM IS BEING INSERTED. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21942657 ↗

Other DRIVER, PROSTHESIS devices

All DRIVER, PROSTHESIS devices →