ATTUNE 2545-00-973Recalls & reported problems

Depuy (Ireland) · ATTUNE CONV FB PS ARTICULATION SURFACE SIZE 3 · Class 1

All models of ATTUNE (1475)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10603295135500), received between Sep 22, 2017 and Sep 10, 2026 — classified as 16 malfunctions.

The most frequently reported specific device problem is “Break” (12 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 12, 2015Class IITerminated

    Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.

    Z-2142-2015 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break12
  • Material Deformation4
  • Component Missing2
  • Crack1
  • Detachment of Device or Device Component1
  • Device-Device Incompatibility1
  • Disassembly1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient7
  • No Code Available2
  • No Information1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50724090 ↗
  • MalfunctionMay 12, 2026Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT ON B)(6) 2026, THE PATIENT UNDERWENT TKA FOR KNEE OA ON THE RIGHT SIDE. AT WOUND CLOSURE, WHEN A NURSE REMOVED THE SHIM ASSEMBLED TO THE DEVICE, THE METAL SPRING RING USED TO SECURE THE SHIM BECAME DISLODGED (SIZE #4 PSRP). THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. LOOSENING OF THE SAME SPRING RING WAS CONFIRMED IN THE SIZE #3 PSFB AND SIZE #5 PERP INSERT TRIALS. THE SURGEON CONFIRMED BY X-RAY THAT NO BROKEN FRAGMENTS OR COMPONENTS REMAINED IN THE PATIENT¿S BODY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25152327 ↗
  • MalfunctionNov 3, 2023Break; Break

    IT WAS REPORTED THAT THE ATTUNE CONV FB PS TB TRL SZ3 BROKE AT THE SPRING AROUND THE SURFACE FACE OFF, DURING KNEE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18064730 ↗
  • MalfunctionOct 31, 2022Crack; Crack

    IT WAS REPORTED THAT SEVERAL ITEMS HAVE CRACKS AND NEED TO BE REPLACED. IT WAS NOTIFIED BY CENTRAL STERILE, THIS DID NOT INVOLVE A SURGERY. RETURNING ALL COMPONENTS FOR INSPECTION. THIS DID NOT DELAY A SURGERY. NO ONE WAS INJURED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15700246 ↗
  • MalfunctionFeb 9, 2022Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE X1 SPRING IS STRETCHED SO DOESN'T FIT THE SHIMS ANYMORE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13497183 ↗
  • MalfunctionNov 19, 2021Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED AS "CENTRAL STERILE HAS 2 ARTIC SURF THAT ARE BROKEN, AND NEED REPLACED. 1 IS IN 2 PIECES, AM RETURNING BOTH PIECES. THE SECOND HAS A BENT PRONG AND SPRING. I AM RETURNING BOTH FOR INSPECTION. THIS DID NOT DELAY A SURGERY, NO ONE WAS INJURED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12847097 ↗
  • MalfunctionJan 29, 2021Break; Break

    IT WAS REPORTED THAT THE POLY TRIAL BROKE WHILE PLACING FINAL IMPLANTS. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11253672 ↗
  • MalfunctionJan 6, 2021Break; Break

    IT WAS REPORTED THAT THE INSERT TRIAL BROKE AFTER EXTRACTION. NO PIECES WERE LEFT IN THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11123565 ↗
  • MalfunctionSep 22, 2020Use of Device Problem; Component Missing; Material Deformation; Material Twisted/Bent; Use of Device Problem; Component Missing; Material Deformation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CENTRAL STERILE DEPARTMENT HAS INDICATED THEY HAVE SEVERAL BROKEN DEPUY ITEMS. A SPACER BLOCK THAT ON THE EXTENSION END HAS ONE BENT/BROKEN PRONG, MISSING A SPRING, WHILE THE OTHER SIDE HAS A WARPED SPRING. A 3 FB ARTIC SURF THAT IS MISSING A SPRING ON THE PRONG, AND HAS A WARPED PRONG ON OPPOSITE SIDE. A SIZE 6 FB ARTIC SURF WITH A BENT PRONG, AND MISSING SPRING. A 8 ARTIC SURFACE THAT IS MISSING A SPRING. NONE OF THESE ITEMS CAUSED A DELAY IN ANY SURGERY. NOBODY WAS INJURED BY THESE MISSING PIECES, OR BROKEN PIECES.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 10569776 ↗
  • MalfunctionSep 15, 2020Break; Break

    IT WAS REPORTED THAT THE SPRING OF AN ATTUNE CONV FB PS TB TRL SZ3 BROKE WHEN TRYING TO REMOVE THE TRIAL INSERT FROM FINAL COMPONENTS. ALL PARTS REMOVED FROM PATIENT.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 10536399 ↗
  • MalfunctionSep 6, 2018Component Missing; Component Missing

    CSD REPORTED A MISSING SPRING ON THE TRIAL. PATIENT CONSEQUENCE? NO.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7850587 ↗
  • MalfunctionAug 9, 2018Break; Break

    IT WAS REPORTED THAT THE ARTICULATING SURFACE CRACKED; ALL PIECES RETRIEVED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7767384 ↗
  • MalfunctionJun 7, 2018Break; Break

    RUPTURE OF A KNEE ATTUNE IMPACTOR. PLEASE FIND PICTURE ATTACHED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7577553 ↗
  • MalfunctionMay 23, 2018Break; Disassembly; Material Deformation; Break; Disassembly; Material Deformation

    IT WAS REPORTED THAT THE SPRING CAME OFF SIZE 3 ARTICULATING SURFACE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7537603 ↗
  • MalfunctionDec 27, 2017Break; Break

    ARTICULATING SURFACE FB CRACKED AND BROKE AND OTHER SIZE 3 ARTICULATING SURFACE FB HAS BROKEN SPRING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7147825 ↗
  • MalfunctionSep 22, 2017Break; Break

    IT WAS REPORTED THAT THE SIZE 3 PS INSERT TRIAL BROKE INTO 3 PIECES WHEN REMOVING THE TRIAL FROM THE KNEE.

    Patient problems: No Information

    MAUDE report 6886198 ↗

Other TEMPLATE devices