ATTUNE recallZ-2142-2015 · DePuy Orthopaedics, Inc. · initiated Jun 12, 2015
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from DePuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.
FDA root cause category: Device Design
What the company is doing
URGENT INFORMATION: MEDICAL DEVICE CORRECTION was issued for all lots of the affected devices. The device correction is to inform US Distributors and Medical Professionals about the issue, how to prevent the issue, and any actions to take. US Distributors will provide Hospitals/User Facilities with the Medical Professional Notice and Acknowledgment Cards. US Distributors will also work with the sales consultants to ensure all consultants complete sales training regarding this issue.
Recall details
- Recall number
- Z-2142-2015
- Status
- Terminated
- Initiated by the company
- Jun 12, 2015
- Posted by the FDA
- Jul 20, 2015
- Recalling firm
- DePuy Orthopaedics, Inc.
- Quantity
- 36607 units
- Product code
- MBH
- FDA submissions
- K140881
- Distribution
- Worldwide Distribution - US including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY and Internationally to CANADA , AUSTRALIA, AUSTRIA, BELGIUM, ENGLAND/ UK, FRANCE, GERMANY, HONG KONG, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, JAPAN, MALAYSIA, NETHERLANDS, NEW ZEALAND, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN & IBERIA, SWEDEN, SWITZERLAND, THAILAND, UAE, CHILE, COLOMBIA, MEXICO, PERU, and PUERTO RICO.
Product description
ATTUNE CONV FB PS ARTICULATION SURFACE SZ1-10 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a total knee replacement. A Balseal is attached to each of the two post features on the Articulation Surface to provide a connection force between the Tibial Articulation Surface and Shim to ensure secure engagement between the components.
Code information (models, lots, serial numbers)
Product Codes:254500971, 254500972, 254500973, 254500974, 254500975, 254500976, 254500977, 254500978, 254500979, and 254500980. GTIN: 10603295135487, 10603295135494, 10603295135500, 10603295135517, 10603295135524, 10603295135531, 10603295135548, 10603295135555, 10603295135562, and 10603295135579.
Devices named in this recall
10 devices in the FDA device registry whose identifier appears in the recall's code information.
- ATTUNE 2545-00-975Depuy (Ireland) · TEMPLATE1 recall48 reports
- ATTUNE 2545-00-977Depuy (Ireland) · TEMPLATE1 recall48 reports
- ATTUNE 2545-00-974Depuy (Ireland) · TEMPLATE1 recall37 reports
- ATTUNE 2545-00-976Depuy (Ireland) · TEMPLATE1 recall37 reports
- ATTUNE 2545-00-978Depuy (Ireland) · TEMPLATE1 recall24 reports
- ATTUNE 2545-00-979Depuy (Ireland) · TEMPLATE1 recall19 reports
- ATTUNE 2545-00-973Depuy (Ireland) · TEMPLATE1 recall16 reports
- ATTUNE 2545-00-980Depuy (Ireland) · TEMPLATE1 recall2 reports
- ATTUNE 2545-00-971Depuy (Ireland) · TEMPLATE1 recall0 reports
- ATTUNE 2545-00-972Depuy (Ireland) · TEMPLATE1 recall0 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K140881 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ATTUNE 2999-65-599Depuy Orthopaedics, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 71421).
- Z-2137-2015 · TerminatedATTUNE FB PS ARTICULATION SURFACE SIZES 3-8 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a tot…
- Z-2138-2015 · TerminatedATTUNE FB CR ARTICULATION SURFACE SIZE 3-8 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a tota…
- Z-2139-2015 · TerminatedATTUNE RP PS ARTICULATION SURFACE SIZE 3-8 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a tota…
- Z-2140-2015 · TerminatedATTUNE CONV RP CR ARTICULATION SURFACE SZ1, 2, 9, and 10 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Tria…
- Z-2141-2015 · TerminatedATTUNE RP CR ARTICULATION SURFACE SIZE 3-8 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a total…
- Z-2143-2015 · TerminatedATTUNE CONV RP PS ARTICULATION SURFACE SZ1-9 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a to…
- Z-2144-2015 · TerminatedATTUNE CONV RP PS TB TRL SZ10 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a total knee repla…
- Z-2145-2015 · TerminatedATTUNE CONV FB CR ARTICULATION SURFACE SZ1 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a tot…
- Z-2146-2015 · TerminatedATTUNE CONV FB CR TB TRL SZ3 - 10 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a total knee re…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page