Skip to content

ATTUNE 2545-00-975Recalls & reported problems

Depuy (Ireland) · ATTUNE CONV FB PS ARTICULATION SURFACE SIZE 5 · Class 1

All models of ATTUNE (1475)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This is a DePuy ATTUNE trial insert used to test knee implant fit during surgery. The 48 reports mostly describe parts cracking or breaking before or during surgery, bent or twisted metal, and small components such as a wire spring ring coming off or going missing; one report involved an injury. There were no deaths, and one recall, started in June 2015 and closed, because that small spring ring could come off and enter the surgical site.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
48
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 48 adverse event reports that list this exact device (DI 10603295135524), received between Jul 7, 2017 and Jun 6, 2025 — classified as 1 injury, 47 malfunctions.

The most frequently reported specific device problem is “Break” (38 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 12, 2015Class IITerminated

    Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.

    Z-2142-2015 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break38
  • Material Twisted/Bent5
  • Component Missing4
  • Crack3
  • Detachment of Device or Device Component2
  • Naturally Worn2
  • Material Deformation1
  • Scratched Material1

Patient problems

  • No Consequences Or Impact To Patient30
  • No Clinical Signs, Symptoms or Conditions12
  • No Information3
  • Not Applicable2
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 48 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 6, 2025Break; Break

    IT WAS NOTICED THAT THERE WAS A SMALL CHIP MISSING FOR AN ARTICULAR SURFACE BEFORE THE SURGERY STARTED. IT WAS NEEDED DURING THE SURGERY. THERE WAS NO DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22157995 ↗
  • MalfunctionJul 25, 2024Break; Break

    IT WAS REPORTED THAT THERE WAS HOLE IN POLY.

    Patient problems: Insufficient Information

    MAUDE report 19825740 ↗
  • MalfunctionJun 21, 2023Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation

    IT WAS REPORTED THAT THE PRONG WAS BENT AND DAMAGED. THE SPRING WAS MISSING AS WELL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17172879 ↗
  • MalfunctionMar 30, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE ATTUNE ARTICULATING SURFACE WHERE THE BALSEAL RING HAS BEEN LOST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16643946 ↗
  • MalfunctionFeb 13, 2023Break; Break

    IT WAS REPORTED THAT THE ARTICULATING TRIAL CRACKED IN HALF WHEN TRIED TO EXPLANT IT. IT BROKE IN ONLY TWO PIECES DOWN THE LEFT SIDE. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16364571 ↗
  • MalfunctionFeb 8, 2023Break; Break

    IT WAS REPORTED THAT TRIAL IS BROKEN. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16322616 ↗
  • MalfunctionFeb 8, 2023Break; Break

    IT WAS REPORTED THAT THE TRIAL IS BROKEN. NO SURGICAL DELAY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16322241 ↗
  • MalfunctionFeb 22, 2022Crack; Crack

    IT WAS REPORTED THAT THE BEARING SURFACE CRACKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13576438 ↗
  • MalfunctionSep 2, 2021Break; Break

    IT WAS REPORTED THAT THE ATTUNE SIZE 5 PS FB ARTIC SURF WAS BROKEN BY THE SURGEON DURING TRIAL REMOVAL. THE AFFECTED PORTION IS THE BACK LEFT CORNER. ALL PIECES WERE REMOVED FROM THE STERILE FIELD. SURGERY TIME WAS NOT EXTENDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12416044 ↗
  • MalfunctionApr 15, 2021Break; Break

    DURING A TOTAL KNEE ARTHROPLASTY, SURGEON WAS ATTEMPTING TO INSERT HIS SZ 5 FB TRIAL POLY INTO THE JOINT SPACE. AFTER IS FIRST UNSUCCESSFUL ATTEMPT DR. DECIDED TO USE THE HANDLE OF A MALLET, TO AID IN GETTING THE TRIAL BEARING TO SEAT. UNFORTUNATELY, THE METAL MALLET IS MUCH HARDER THEN A PLASTIC POLY TRIAL AND THE TRIAL BROKE. SURGERY DELAYED 1 MINUTE REMOVED BROKEN POLY TRIAL AND INSERTED ANOTHER POLY TRIAL, PROCEDURE SUCCESSFULLY COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11671446 ↗
  • MalfunctionFeb 3, 2021Use of Device Problem; Material Twisted/Bent; Use of Device Problem; Material Twisted/Bent

    IT WAS REPORTED BROKEN, WORN DOWN TRIALS AND MISSING BEARINGS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11275640 ↗
  • MalfunctionJan 29, 2021Break; Break

    IT WAS REPORTED THAT THE ATTUNE SZ 5 FB ARTICULATING SURFACE BROKE IN HALF DOWN THE MIDDLE WHEN THE SURGEON WAS REMOVING FROM THE JOINT. NO PATIENT INVOLVEMENT. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11251004 ↗
  • MalfunctionJan 5, 2021Break; Break

    TRIAL FB PS TIBIAL INSERT BROKE IN 2 WHILE REMOVING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11118638 ↗
  • MalfunctionOct 23, 2020Break; Break

    IT WAS REPORTED THAT THE FEMORAL IMPACTOR WAS BROKEN WHILE IMPLANTING FEMUR, AND ARTICULAR SURFACE WAS BROKEN WHILE TRYING TO REMOVE IT WHEN TRYING. ALL PIECES WERE RETRIEVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10729975 ↗
  • MalfunctionOct 16, 2020Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE PS INSERT TRIAL FOUND WITH HOLE WORN INTO PLASTIC PIECE WAS NOT USED DURING PROCEDURE. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10689463 ↗
  • MalfunctionAug 31, 2020Break; Break

    IT WAS REPORTED THAT THE ARTICULATING SURFACE BROKE WHILE REMOVING TRIAL INSERT. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10474430 ↗
  • MalfunctionJul 22, 2020Break; Component Missing; Material Twisted/Bent; Naturally Worn; Break; Component Missing; Material Twisted/Bent; Naturally Worn

    IT WAS REPORTED THAT THEY HAVE SEVERAL ITEMS THAT THE CENTRAL STERILE DEPARTMENT WANTS REPLACED, DUE TO BROKEN PIECES. I HAVE 2 ARTIC SURF PIECES MISSING SPRINGS. I HAVE 1 ARTIC SURF MISSING AN ENTIRE PRONG, THEY DO NOT HAVE MISSING PRONG, THERE WAS JUST A HOLE WHERE IT USED TO BE. THEY HAVE IMPACTION HANDLE THAT THE RED TRIGGERS AS BROKEN OFF, THEY HAVE ALL 3 PIECES, HANDPIECE, TRIGGER, AND SPRING. NONE OF THESE ITEMS WERE BROKEN IN SURGERY, NONE OF THESE ITEMS DELAYED A SURGERY, NO ONE WAS INJURED BY THESE ITEMS. THEY ARE BROKEN, MISSING PIECES FROM NORMAL WEAR AND TEAR.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10309860 ↗
  • MalfunctionApr 8, 2020Break; Break

    IT WAS REPORTED THAT THE TRIAL BROKE IN 2 PIECES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9934511 ↗
  • MalfunctionJan 29, 2020Component Missing; Material Twisted/Bent; Scratched Material; Component Missing; Material Twisted/Bent; Scratched Material

    DURING REPROCESSING, THE STERILIZATION TECHNICIAN FOUND THAT ONE OF THE BALSEAL SPRINGS ON AN ATTUNE SIZE 5 TRIAL ARTICULATING SURFACE WAS BENT. THE TECHNICIAN ALSO FOUND THERE TO BE DEEP SCRATCHES ON THE FRONT SURFACE OF THE TRIAL. WHEN I WAS GIVEN THE ITEM, THE ONE OF THE 2 BALSEAL SPRINGS WAS MISSING. IT IS NOT AVAILABLE FOR RETURN. AS FAR AS I HAVE BEEN TOLD, THE SPRING WAS DISPOSED OF. THERE WAS NO PATIENT CONSEQUENCE TO MY KNOWLEDGE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9643807 ↗
  • MalfunctionJan 17, 2020Break; Break

    IT WAS REPORTED THAT THE ATTUNE ARTICULAR SURFACE BEARING POST WAS BROKE OFF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9603626 ↗
  • MalfunctionJan 9, 2020Break; Break

    IT WAS REPORTED THAT THE SIZE 5 FB ARTICULATING SURFACE WAS BROKEN. ALL BROKEN PARTS WERE ACCOUNTED FOR. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9570818 ↗
  • MalfunctionNov 1, 2019Break; Break

    IT WAS REPORTED THAT AFTER THE CASE, WHILE TAKING SHIMS OFF OF INSERT TRIALS, THE TRIAL WAS BROKEN. THE BROKEN PIECE WAS THROWN AWAY BUT THE INSERT WAS BEING RETURNED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9267639 ↗
  • MalfunctionOct 31, 2019Component Missing; Component Missing

    UPON INSPECTION NOTICED THE SPRING MISSING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9262306 ↗
  • MalfunctionOct 11, 2019Break; Material Twisted/Bent; Break; Material Twisted/Bent

    IT WAS REPORTED THAT SPRING OF THE ATTUNE CONV FB PS TB TRL SZ5 WAS BROKEN. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9179717 ↗
  • MalfunctionOct 3, 2019Component Missing; Component Missing

    SPRING CAME OFF OF AN INSTRUMENT AND WAS LOST IN THEATRE. STAFF REPORT THAT THE INSTRUMENTS WERE CHECKED AT THE BEGINNING OF THE SURGERY AND ALL WAS OK. X-RAY WAS USED TO TRY AND LOCATE THE SPRING. NOT IDENTIFIED ON X-RAY AND NOT FOUND IN THEATRE. NOT RESOLVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9150517 ↗