ATTUNE 2545-00-976Recalls & reported problems

Depuy (Ireland) · ATTUNE CONV FB PS ARTICULATION SURFACE SIZE 6 · Class 1

All models of ATTUNE (1475)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy ATTUNE knee implant trial insert is used to try different sizes and fit during knee replacement surgery. Most of the 37 reports describe the plastic trial part breaking, cracking, chipping, or snapping during surgery, sometimes into two pieces that the surgeon had to remove; a few also mention a small wire spring missing from the instrument. There are no deaths and no injuries. One recall from June 2015 concerned that spring, which could come loose and enter the surgical site; it was terminated.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
37
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 10603295135531), received between Jul 12, 2017 and May 8, 2026 — classified as 37 malfunctions.

The most frequently reported specific device problem is “Break” (32 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 12, 2015Terminated

    Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.

    Z-2142-2015 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break32
  • Component Missing2
  • Crack2
  • Illegible Information2
  • Material Deformation2
  • Bent1
  • Detachment of Device or Device Component1
  • Material Twisted/Bent1

Patient problems

  • No Consequences Or Impact To Patient17
  • No Clinical Signs, Symptoms or Conditions14
  • No Information4
  • Insufficient Information1
  • No Code Available1
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 8, 2026Break

    IT WAS REPORTED THAT THE ATTUNE KNEE ARTICULATING SURFACE WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25125684 ↗
  • MalfunctionJan 13, 2026Break

    IT WAS REPORTED THAT DURING A TOTAL KNEE ARTHROPLASTY (TKA) PROCEDURE, THE POLYETHYLENE TRIAL SNAPPED IN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24052793 ↗
  • MalfunctionJul 7, 2025Break; Break

    IT WAS REPORTED THAT UPON REMOVING THE SIZE 6 FB INSERT TRIAL, THE SURGEON NOTICED THAT THE INSERT CRACKED INTO 2 PIECES. SURGEON REMOVED ALL PARTS FROM THE KNEE. ADDED NO TIME TO THE PROCEDURE AND NO HARM TO THE PATIENT. ALL PIECES WERE REMOVED AND CONFIRMED BROKEN PART WAS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22412457 ↗
  • MalfunctionDec 26, 2024Break; Break

    IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2024, THE TRIAL WAS BROKEN WHILE BEING REMOVED.

    Patient problems: Insufficient Information

    MAUDE report 21019660 ↗
  • MalfunctionFeb 18, 2024Break; Crack; Break; Crack

    IT WAS REPORTED THAT THE TRIALS IS CHIPPED/DAMAGED. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18728546 ↗
  • MalfunctionJan 11, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT SPRING OF THE INSTRUMENT IS MISSING FROM THE ARTICULATION SURFACE TRIAL. PATIENT STATUS/ OUTCOME / CONSEQUENCES, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN, IS THE PATIENT PART OF A CLINICAL STUDY UNKNOWN, IP-01939388 DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST. TRUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18493859 ↗
  • MalfunctionNov 17, 2023Break; Break

    IT WAS REPORTED THAT FB PS TIBIAL INSERT TRIAL SNAPPED UPON 2 FEMUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18158662 ↗
  • MalfunctionOct 24, 2023Break; Break

    IT WAS REPORTED THAT THE ATTUNE ARTICULATING SURFACE TRIAL BROKE IN TWO PIECES UPON IMPACTION DURING SURGERY. THERE WAS NO DELAY TO SURGERY CAUSED BY THIS, AND NO HARM TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17996266 ↗
  • MalfunctionMar 21, 2023Break; Break

    IT WAS REPORTED THAT ARTICULATE SURFACE CRACKED INTO TWO PIECES WHILE TRIALING. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16581924 ↗
  • MalfunctionMar 8, 2023Break; Break

    IT WAS REPORTED THAT A TIBIAL IMPACTOR WAS CRACKED DURING THE IMPACTION OF THE IMPLANTS. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16504106 ↗
  • MalfunctionFeb 16, 2023Break; Break

    IT WAS REPORTED THAT ALL PIECES WERE BROKEN DURING STERILIZATION. TWO FB ARTICULATION SURFACES WERE BROKEN IN HALF. ONE ARTICULATION SURFACE BROKE OFF AND A PIECE IS STUCK IN THE SHIM WHICH IS ALSO BEING SENT BACK. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16389384 ↗
  • MalfunctionJan 20, 2023Break; Break

    IT WAS REPORTED THAT THE TIBIAL TRAY TRIAL WAS BROKEN. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16218082 ↗
  • MalfunctionOct 24, 2022Break; Break

    IT WAS REPORTED THAT THE PEG ON BOTTOM OF INSERT BROKE OFF DURING SURGERY. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15659719 ↗
  • MalfunctionOct 7, 2022Break; Break

    IT WAS REPORTED THAT THE PS FB SIZE 6 TRIAL INSERT BROKE DURING TRIALLING. ALL PIECES REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15561690 ↗
  • MalfunctionFeb 21, 2022Break; Break

    SPRING BROKE OFF OF TRIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13572383 ↗
  • MalfunctionNov 9, 2020Break; Break

    ADDITIONAL INFORMATION RECEIVED STATED THAT THE WHOLE TRIAL BROKE IN TWO PIECES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10810415 ↗
  • MalfunctionSep 22, 2020Use of Device Problem; Component Missing; Material Deformation; Material Twisted/Bent; Use of Device Problem; Component Missing; Material Deformation; Material Twisted/Bent

    IT WAS REPORTED THAT THE CENTRAL STERILE DEPARTMENT HAS INDICATED THEY HAVE SEVERAL BROKEN DEPUY ITEMS. A SPACER BLOCK THAT ON THE EXTENSION END HAS ONE BENT/BROKEN PRONG, MISSING A SPRING, WHILE THE OTHER SIDE HAS A WARPED SPRING. A 3 FB ARTIC SURF THAT IS MISSING A SPRING ON THE PRONG, AND HAS A WARPED PRONG ON OPPOSITE SIDE. A SIZE 6 FB ARTIC SURF WITH A BENT PRONG, AND MISSING SPRING. A 8 ARTIC SURFACE THAT IS MISSING A SPRING. NONE OF THESE ITEMS CAUSED A DELAY IN ANY SURGERY. NOBODY WAS INJURED BY THESE MISSING PIECES, OR BROKEN PIECES.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 10569803 ↗
  • MalfunctionApr 15, 2020Break; Illegible Information; Break; Illegible Information

    IT WAS REPORTED THAT THE INSTRUMENTS ARE BROKEN OR SIZES CAN¿T BE READ ANYMORE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9964460 ↗
  • MalfunctionApr 15, 2020Break; Illegible Information; Break; Illegible Information

    IT WAS REPORTED THAT THE INSTRUMENTS ARE BROKEN OR SIZES CAN¿T BE READ ANYMORE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9964457 ↗
  • MalfunctionMar 23, 2020Break; Break

    IT WAS REPORTED THAT THE TRIAL WAS BROKEN WHEN TRYING TO REMOVE FROM KNEE. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9868230 ↗
  • MalfunctionDec 27, 2019Break; Break

    IT WAS REPORTED THAT THE PIECE OF PLASTIC BROKE OFF THE PEG THAT HOLDS THE SPRING IN PLACE. ALL PIECES WERE RETRIEVED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9525864 ↗
  • MalfunctionDec 26, 2019Break; Break

    ATTUNE PS ARTICULATION SURFACE SZ6 CRACKED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9520742 ↗
  • MalfunctionOct 11, 2019Break; Break

    IT WAS REPORTED THAT WHEN REMOVING THE SIZE 6 ATTUNE FB PS TRIAL, THE POSTERIOR CORNER BROKE OFF INTO 2 PIECES. ALL PIECES WERE RETRIEVED. 1 MINUTE SURGICAL DELAY.

    Patient problems: Not Applicable

    MAUDE report 9180899 ↗
  • MalfunctionSep 30, 2019Break; Break

    ATTUNE INSERT TRIALS SPRINGS ARE BROKEN. PATIENT STATUS/ OUTCOME / CONSEQUENCES: NO.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9135444 ↗
  • MalfunctionMay 23, 2019Break; Break

    I WAS REPORTED THAT THE NIPPLE OF THE POLY CONNECTION BROKE OFF. ALL PIECES WERE RETRIEVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8637908 ↗