ATTUNE 2545-00-978Recalls & reported problems

Depuy (Ireland) · ATTUNE CONV FB PS ARTICULATION SURFACE SIZE 8 · Class 1

All models of ATTUNE (1475)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Attune knee trial insert is a reusable instrument used to test fit during knee replacement surgery. The 24 reports mostly describe the insert breaking during trialing or sterilization, sometimes into pieces that were retrieved, with no deaths and no injuries mentioned. It has 1 closed recall started in June 2015 about a small spring coil that could come off and enter the surgical site.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 10603295135555), received between Oct 19, 2017 and Aug 13, 2025 — classified as 24 malfunctions.

The most frequently reported specific device problem is “Break” (21 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 12, 2015Class IITerminated

    Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.

    Z-2142-2015 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break21
  • Component Missing1
  • Crack1
  • Detachment of Device or Device Component1
  • Material Deformation1
  • Material Twisted/Bent1
  • Naturally Worn1

Patient problems

  • No Consequences Or Impact To Patient12
  • No Clinical Signs, Symptoms or Conditions9
  • No Information3

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 13, 2025Break; Break

    IT WAS REPORTED THAT THE ARTICULATING SURFACE BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22797186 ↗
  • MalfunctionAug 13, 2025Break; Break

    IT WAS REPORTED THAT THE ARTICULATING SURFACE BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22797181 ↗
  • MalfunctionApr 4, 2023Break; Break

    IT WAS REPORTED THAT THE TRIAL INSERT BROKE WHILE TRIALING. IT BROKE INTO TWO PIECES. ALL PIECES WERE RETRIEVED FROM THE KNEE. THIS DID NOT ADD ANYMORE TIME TO THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16673726 ↗
  • MalfunctionFeb 16, 2023Break; Break

    IT WAS REPORTED THAT ALL PIECES WERE BROKEN DURING STERILIZATION. TWO FB ARTICULATION SURFACES WERE BROKEN IN HALF. ONE ARTICULATION SURFACE BROKE OFF AND A PIECE IS STUCK IN THE SHIM WHICH IS ALSO BEING SENT BACK. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16389364 ↗
  • MalfunctionJan 4, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE ARTIC SURFACE PIECE WAS DEFECTIVE. ONE HAS MISSING SPRING ON PRONG, OTHER HAS WARPED PRONG WITH DESTROYED SPRING. NO SURGICAL DELAY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16098249 ↗
  • MalfunctionAug 16, 2022Break; Break

    IT WAS REPORTED THAT THE SZ 8 FB ARTIC SURFACE BROKE, ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15234018 ↗
  • MalfunctionNov 7, 2021Break; Break

    ADDITIONAL INFORMATION RECEIVED STATES THAT THE INSTRUMENT WAS BROKEN INTO MORE THAN 2 PIECES. THE PIECE IS AVAILABLE FOR RETURN. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12759462 ↗
  • MalfunctionOct 15, 2021Break; Break

    ATTUNE TRIAL ARTICULATING SURFACE BROKE DURING REDUCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12641021 ↗
  • MalfunctionJan 4, 2021Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT 2 ARTIC SURFACE PIECES THAT ARE CHIPPED, AND HAVE BENT PRONGS. THE DECONTAM DEPARTMENT BROUGHT THIS TO MY ATTENTION. NOBODY WAS INJURED, AND NO SURGICAL CASE WAS DELAYED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11109435 ↗
  • MalfunctionJul 10, 2020Break; Crack; Break; Crack

    IT WAS REPORTED THAT THE CENTRAL STERILE DEPARTMENT NOTIFIED OF A BROKEN ARTIC SURFACE SIZE 8. THIS WAS STILL ONE PIECE, BUT WAS BROKEN ALMOST HALF WAY THROUGH TRIAL. THIS DID NOT DELAY A CASE. NO ONE WAS INJURED BY ARTIC SURFACE PIECE. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10259127 ↗
  • MalfunctionMay 18, 2020Component Missing; Material Twisted/Bent; Component Missing; Material Twisted/Bent

    IT WAS REPORTED THAT THE ARTIC SURFACE HAS A BENT PRONGS AND MISSING A SPRING. CENTRAL STERILE HAS REQUESTED A NEW ONE. THIS WAS NOT USED IN A CASE SO DID NOT DELAY SURGERY. NO ONE WAS INJURED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10067629 ↗
  • MalfunctionFeb 18, 2020Break; Break

    IT WAS REPORTED THAT THE PLASTIC TAB BROKE OFF DURING TRIALING. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9722769 ↗
  • MalfunctionJan 8, 2020Break; Break

    IT WAS REPORTED THAT THE SPRING AROUND THE COIL BROKE DURING EXTRACTION OF TRIAL ARTICULATE SURFACE. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9565559 ↗
  • MalfunctionNov 8, 2019Break; Break

    IT WAS REPORTED THAT THE SPRING WAS BROKEN. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9296047 ↗
  • MalfunctionNov 4, 2019Break; Break

    IT WAS REPORTED THAT DURING THE TKA PROCEDURE, DOCTOR WAS REMOVING ATTUNE PS FB SHIM, SIZE 8. THE SHIM BROKE IN HALF. BOTH PIECES WERE RETRIEVED FROM THE PATIENT. THE PRODUCT CODE WAS LEGIBLE AND THE PRODUCT WAS DAMAGED TO THE BROKEN SHIM. NOT SURE IF THE LOT NUMBER IS CORRECT. ONLY NUMBERS ON TRIAL SHIM. NO ADDITIONAL INFORMATION AVAILABLE. PLEASE SEND REPLACEMENT SHIM TO WNC TEAM. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9274772 ↗
  • MalfunctionMar 21, 2019Break; Break

    BROKEN INSTRUMENTS DURING AN UNSPECIFIED SURGERY. NO DELAY IN SURGERY. SURGERY COMPLETED SUCCESSFULLY. NO FRAGMENTS GENERATED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8439392 ↗
  • MalfunctionJun 1, 2018Naturally Worn; Naturally Worn

    INSTRUMENTS WERE REPORTED FOR UNKNOWN REASON.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7560183 ↗
  • MalfunctionMay 31, 2018Break; Break

    IT WAS REPORTED THAT SIZE 8 FB ARTIC SURF WAS FOUND TO HAVE A CHIP MISSING FROM THE FRONT OF IT DURING THE DECONTAMINATION PROCESS. NO PATIENT WAS INVOLVED IN THIS INCIDENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7557600 ↗
  • MalfunctionFeb 26, 2018Break; Break

    DER STATED THAT THE ATTUNE SIZE 8 FB ARTICULATING SURFACE TRAY BROKE INTO TWO SEPARATE PIECES WHILE THE DOCTOR WAS REMOVING IT AFTER TRIALING. NO SURGICAL DELAY. ALL PIECES WERE RETRIEVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7297306 ↗
  • MalfunctionJan 31, 2018Break; Break

    IT WAS REPORTED THAT SIZE 8 SPRING BROKEN ON ARTICULATING SURFACE AND SIZE 7 NUB BROKE OFF NEAR SPRING ARTICULATING SURFACE. NO SURGICAL DELAY. NO PIECES OF BROKEN INSTRUMENTATION LEFT IN PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7230486 ↗
  • MalfunctionDec 20, 2017Break; Break

    IT WAS REPORTED THAT SPRING BROKE OFF OF TRIAL ARTICULATION SURFACE UPON REMOVAL. NO SURGICAL DELAY. NO PIECES OF BROKEN INSTRUMENTATION LEFT IN PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7135021 ↗
  • MalfunctionNov 17, 2017Break; Break

    IT WAS REPORTED THAT THE TRIAL HAS BROKEN INTO THREE PIECES.

    Patient problems: No Information

    MAUDE report 7041116 ↗
  • MalfunctionNov 13, 2017Break; Break

    IT WAS REPORTED THAT THE TIBIAL TRIAL WAS BROKEN WHEN TRIALING. THE RING WAS BUSTED ON BOTTOM SIDE OF ARTICULATING SURFACE.

    Patient problems: No Information

    MAUDE report 7026343 ↗
  • MalfunctionOct 19, 2017Break; Break

    DER STATES THAT THE ATTUNE CONV FB PS TB TRL SZ8 WAS IN SPD, WITH THE BROKEN PIECE INCLUDED, AFTER BEING REMOVED FROM THE WASH.

    Patient problems: No Information

    MAUDE report 6960915 ↗